  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Snoqualmie",
      "address_1": "35301 SE Center St",
      "reason_for_recall": "The transmitter display and the Telemetry Central Station display may show a SpO2 value when the sensor is not connected to the patient or in certain sensor failed conditions.    This is an expansion of recall Z-0871-2014 because additional serial numbers were identified as being part of this recall action as part of the firm's retrospective analysis of all recalls since 2011.",
      "address_2": "",
      "product_quantity": "8 units",
      "code_info": "Serial numbers:  6281-002500, 6281-002501, 6281-002659, 6281-002661, 6281-002662, 6281-002663, 6281-002664, and 6281-003056.    Model 96281, with SpO2 Option C.",
      "center_classification_date": "20140904",
      "distribution_pattern": "US Distribution to TX.",
      "state": "WA",
      "product_description": "ARIATELE TELEMETRY TRANSMITTER, Model 96281, with SpO2 Option C.  Physiological patient monitor.  Used in a hospital environment.",
      "report_date": "20140910",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Spacelabs Healthcare Inc",
      "recall_number": "Z-2586-2014",
      "initial_firm_notification": "E-Mail",
      "product_type": "Devices",
      "event_id": "69125",
      "termination_date": "20141016",
      "more_code_info": "",
      "recall_initiation_date": "20140702",
      "postal_code": "98065-9216",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}