{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Cambridge",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99126",
      "recalling_firm": "Philips North America Llc",
      "address_1": "222 Jacobs St",
      "address_2": "N/A",
      "postal_code": "02141-2289",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of AZ, CA, CO, IN, MA, MI, MN, PA, SD, TN.",
      "recall_number": "Z-2579-2026",
      "product_description": "Philips Telemetry Monitor 5500 1.4 GHz. Model Number: 867232. Includes the following system descriptions: TELE Monitor 5500, 1.4 Ghz, ECG only, ex; Telemetry Monitor 5500 1.4 GHz FAST EX.",
      "product_quantity": "909 units",
      "reason_for_recall": "Potential for device reset to default \"NEW_DEVICE\" state, which may lead to a to loss of device configuration and the equipment label, resulting in a failure to reconnect to the PIC iX central monitoring system.",
      "recall_initiation_date": "20260605",
      "center_classification_date": "20260624",
      "report_date": "20260701",
      "code_info": "Model Number: 867232. UDI-DI: (01)00884838102842. Software Version Number: 4.0.0.0. Serial Numbers: US001A002F, US001A00AW, US001A00C9, US001A00EW, US001A00F0, US001A00GM, US001A00N4, US001A00PB, US001A00PZ, US001A00R1, US001A00UE, US001A00VU, US001A00Z4, US001A013T, US001A018A, US001A018R."
    }
  ]
}