      "status": "Terminated",
      "city": "Waukesha",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80484",
      "recalling_firm": "GE Healthcare, LLC",
      "address_1": "3000 N Grandview Blvd",
      "address_2": "N/A",
      "postal_code": "53188-1615",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide distribution to AL, CA, CO, FL, IL, IN, MA, MD, MI, MS, NC, NJ, NY,OH, OK, SC, TN, TX, VA.",
      "recall_number": "Z-2579-2018",
      "product_description": "Discovery MI Digital Ready",
      "product_quantity": "42 units total",
      "reason_for_recall": "DIGITAL ReadyCT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible  Power Supply).",
      "recall_initiation_date": "20180524",
      "center_classification_date": "20180731",
      "termination_date": "20190208",
      "report_date": "20180808",
      "code_info": "Model Number 5986030.   a. Serial Number PTDMI1800003PT, System ID Number RADNETPT1839, UDI Number 01008406821209751118010421PTDMI1800003PT;    b. Serial Number PTDMI1800005PT, System ID Number RADNETPT1838, UDI Number 01008406821209751118010421PTDMI1800003PT;    c.  Serial Number PTDMI1800009PT, System ID Number 813745DMIDR2, UDI Number 01008406821209751118021921PTDMI1800009PT"
    }
  ]
}