  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Freiburg Im Breisgau",
      "state": "",
      "country": "Germany",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92879",
      "recalling_firm": "PTW-FREIBURG",
      "address_1": "Lorracher Str. 7",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-2577-2023",
      "product_description": "BeamAdjust software Version 2.3.3  Ref: S080032   The following devices are affected when a relative calibration is performed using ArrayCal (module from BeamAdjust S080032, UDI-DI: EPTWS0800320\t):  \"\tOCTAVIUS Detector 1600 SRS \t(T10056) (UDI-DI: EPTWT100560)  \"\tOCTAVIUS Detector 1600 MR \t(T10057) (UDI-DI: EPTWT100570)  \"\tOCTAVIUS Detector 1600 XDR \t(T10058) (UDI-DI: EPTWT100580)  \"\tOCTAVIUS Detector 1000 SRS \t(T10036) (UDI-DI: EPTWT100360)",
      "product_quantity": "42 US",
      "reason_for_recall": "Generated calibration file will not be correct in case of a relative calibration of an OCTAVIUS Detector 1000 or OCTAVIUS Detector 1600, if a Tiff file or a DICOM data set is used as reference matrix. Therefore, the measurement result will be incorrect by using such a calibration file.",
      "recall_initiation_date": "20230607",
      "center_classification_date": "20230911",
      "report_date": "20230920",
      "code_info": "UDI-DI: EPTWS0800320 BeamAdjust All software Versions",
      "more_code_info": ""
    }
  ]
}