  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Crumlin (North)",
      "state": "N/A",
      "country": "Ireland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80475",
      "recalling_firm": "Randox Laboratories Ltd.",
      "address_1": "55 Diamond Road",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide Distribution - US in the state of WV. Foreign distribution to Australia, Chile, China, Germany, Hungary, Ireland, Italy, Moldova, Netherlands, Peru, Poland, Saudi Arabia, Slovakia, Spain, Sultanate of Oman, UAE, UK, and Vietnam",
      "recall_number": "Z-2575-2018",
      "product_description": "Liquid Cardiac Control - Level 3, cat. no. CQ5053    Product Usage:  This product is intended for in vitro diagnostic use, in the quality control of Cardiac Markers on clinical chemistry and Immunoassay systems.",
      "product_quantity": "1 kit (US), 602 kits (OUS)",
      "reason_for_recall": "The device is not suitable for the control of the Troponin T assay due to unacceptable variation between vials. The deviation in precision observed is up to 28%. IQC that is reported as out of range could lead to a delay in reporting Troponin T results. If the failure occurs the operator may observe the quality control result for Troponin T falling outside range and/or increased imprecision for Quality Control monitoring.",
      "recall_initiation_date": "20180608",
      "center_classification_date": "20180729",
      "termination_date": "20181105",
      "report_date": "20180808",
      "code_info": "Lot 4245CK."
    }
  ]
}