  "results": [
    {
      "status": "Ongoing",
      "city": "Espoo",
      "state": "N/A",
      "country": "Finland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98918",
      "recalling_firm": "SURGIFY MEDICAL OY",
      "address_1": "Otakaari 5 I",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Nationwide distribution in the states of MN, MA, IN, LA, CA, NY, FL.",
      "recall_number": "Z-2567-2026",
      "product_description": "Surgify Halo, 5.4 mm, Short, Model/Catalog Number: 54.070.NVG.U1; drills, burrs, trephines & accessories (simple, powered)",
      "product_quantity": "N/A",
      "reason_for_recall": "Potential for burr breakage, occurring during bi-portal endoscopic spinal surgery (BESS).",
      "recall_initiation_date": "20260430",
      "center_classification_date": "20260624",
      "report_date": "20260701",
      "code_info": "UDI: 06429811532298 ; All lots until the IFU update has been implemented"
    }
  ]
}