      "status": "Terminated",
      "city": "Palo Alto",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68627",
      "recalling_firm": "Varian Medical Systems, Inc.",
      "address_1": "911 Hansen Way",
      "address_2": "N/A",
      "postal_code": "94304-1028",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the countries of  Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Byelarus, Belgium, Bosniak, Brazil;, Bulgaria, Burma, Canada, Chile, China, Columbia, Costa Rica, Czech Republic, Denmark, Dominican Republic, Ecuador, Estonia, Finland, France, Germany, hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Luxembourg, Macau, Macedonia, Malaysia, Mauritania, Mexico, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia,, Singapore, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Venezuela.",
      "recall_number": "Z-2567-2014",
      "product_description": "C-Series High Energy Linear Accelerator - Models: Novalis Tx, Trilogy, Trilogy Tx, Clinac iX, Clinac 2100 C/D, Clinac 2300C/D, Clinac 21EX, Clinac 23EX with versions 7, 8, 9 and 6 MV configuration and maximum dose rate of at least 6000 MU/min only.    Varian High Energy Linear Accelerator is intended to provide sterotactic radiosurgery and precision radio-therapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.",
      "product_quantity": "3416",
      "reason_for_recall": "Trend in reports of unexpected decrease in beam output in C-series High Energy Linear Accelerators for 6MV photon treatment mode.",
      "recall_initiation_date": "20140617",
      "center_classification_date": "20140912",
      "termination_date": "20180920",
      "report_date": "20140924",
      "code_info": "Model numbers HCX, H14, H27, H29.  Serial numbers: H140001  H140152  H140724  H140791  H140857  H140924  H140985  H271272  H271729  H272009  H272180  H272353  H272819  H140011  H140201  H140725  H140792  H140858  H140925  H140986  H271276  H271755  H272014  H272181  H272354  H272820  H140018  H140202  H140727  H140793  H140859  H140926  H140987  H271293  H271759  H272015  H272187  H272359  H272822  H140025  H140203  H140728  H140795  H140860  H140927  H140988  H271308  H271762  H272016  H272188  H272362  H272824  H140026  H140204  H140729  H140796  H140861  H140928  H140989  H271330  H271763  H272018  H272195  H272365  H272825  H140027  H140206  H140730  H140797  H140863  H140930  H140990  H271334  H271773  H272022  H272197  H272370  H272826  H140029  H140207  H140732  H140799  H140864  H140931  H140991  H271385  H271783  H272023  H272198  H272377  H272828  H140036  H140208  H140733  H140800  H140865  H140932  H140992  H271400  H271785  H272024  H272201  H272382  H272834  H140038  H140209  H140734  H140801  H140866  H140933  H140993  H271407  H271787  H272027  H272202  H272388  H272837  H140039  H140210  H140735  H140802  H140867  H140934  H140994  H271413  H271802  H272029  H272204  H272391  H272841  H140045  H140211  H140736  H140803  H140868  H140935  H140995  H271414  H271809  H272032  H272209  H272394  H272846  H140046  H140213  H140737  H140804  H140869  H140936  H140996  H271420  H271815  H272035  H272212  H272397  H272847  H140047  H140214  H140738  H140805  H140870  H140937  H140997  H271438  H271825  H272038  H272216  H272401  H272848  H140048  H140215  H140742  H140806  H140871  H140938  H140998  H271439  H271828  H272043  H272219  H272405  H272849  H140058  H140216  H140743  H140807  H140872  H140939  H140999  H271443  H271833  H272045  H272220  H272408  H272851  H140062  H140217  H140744  H140808  H140874  H140940  H141000  H271452  H271834  H272049  H272223  H272410  H272852  H140067  H140218  H140745  H140809  H140875  H140941  H141002  H271459  H271835  H272050  H272226  H272413  H272854  H140070  H140219  H140747  H140810  H140876  H140942  H141003  H271490  H271838  H272051  H272231  H272415  H272856  H140071  H140220  H140748  H140811  H140877  H140943  H141004  H271497  H271843  H272052  H272235  H272418  H272857  H140072  H140221  H140749  H140812  H140878  H140944  H141005  H271500  H271845  H272062  H272236  H272421  H272858  H140076  H140222  H140750  H140813  H140879  H140945  H141007  H271502  H271853  H272064  H272237  H272422  H272859  H140077  H140223  H140751  H140816  H140880  H140946  H141008  H271507  H271856  H272069  H272239  H272423  H272862  H140078  H140224  H140752  H140817  H140881  H140947  H141009  H271512  H271858  H272077  H272240  H272425  H272864  H140079  H140225  H140753  H140818  H140882  H140948  H141010  H271515  H271862  H272078  H272241  H272426  H272865  H140086  H140226  H140754  H140819  H140883  H140949  H141011  H271516  H271866  H272079  H272250  H272427  H272867  H140090  H140227  H140755  H140820  H140884  H140950  H141012  H271519  H271867  H272081  H272259  H272428  H272872  H140092  H140228  H140756  H140821  H140885  H140951  H141013  H271526  H271870  H272085  H272260  H272432  H272873  H140096  H140229  H140757  H140822  H140888  H140952  H141015  H271531  H271872  H272086  H272271  H272438  H272880  H140097  H140230  H140758  H140823  H140889  H140953  H141016  H271534  H271879  H272089  H272283  H272440  H272884  H140098  H140231  H140760  H140824  H140890  H140954  H141017  H271550  H271881  H272094  H272284  H272441  H272885  H140101  H140232  H140761  H140826  H140891  H140955  H141018  H271569  H271893  H272096  H272290  H272444  H272888  H140103  H140233  H140762  H140827  H140892  H140956  H141019  H271571  H271896  H272111  H272291  H272445  H272889  H140106  H140234  H140763  H140829  H140893  H140957  H141020  H271575  H271903  H272115  H272293  H272446  H272890  H140109  H140235  H140765  H140830  H140895  H140958  H141021  H271588  H271905  H272117  H272299  H272454  H272893  H140111  H140236  H140766  H140832  H140896  H140959  H141022  H271593  H271906  H272118  H272300  H272456  H272896  H140113  H140237  H140767  H140833  H140897  H140960  H141024  H271594  H271918  H272125  H272301  H272460  H272899  H140114  H140238  H140768  H140834  H140901  H140961  H141025  H271600  H271921  H272127  H272302  H272462  H272908  H140115  H140239  H140769  H140835  H140902  H140962  H141026  H271602  H271926  H272129  H272306  H272463  H272909  H140117  H140240  H140770  H140836  H140903  H140963  H141027  H271607  H271931  H272132  H272307  H272464  H272910  H140121  H140701  H140771  H140837  H140904  H140966  H141028  H271614  H271935  H272134  H272309  H272466  H272911  H140122  H140702  H140772  H140838  H140905  H140967  H141029  H271618  H271938  H272136  H272310  H272469  H272922  H140124  H140703  H140773  H140839  H140906  H140968  H141032  H271624  H271941  H272138  H272311  H272470  H272926  H140125  H140705  H140775  H140840  H140907  H140969  H141034  H271626  H271942  H272143  H272313  H272471  H272932  H140127  H140706  H140776  H140841  H140908  H140970  H141035  H271628  H271951  H272153  H272318  H272473  H272933  H140128  H140708  H140777  H140843  H140909  H140971  H141036  H271635  H271954  H272155  H272321  H272476  H272936  H140132  H140709  H140778  H140844  H140910  H140972  H141037  H271662  H271958  H272156  H272322  H272480  H272938  H140133  H140710  H140779  H140845  H140911  H140973  H141039  H271663  H271959  H272157  H272324  H272484  H272941  H140134  H140712  H140780  H140846  H140912  H140974  H141041  H271664  H271964  H272158  H272326  H272488  H272943  H140135  H140713  H140781  H140847  H140913  H140975  H141044  H271666  H271966  H272159  H272327  H272491  H272946  H140136  H140714  H140782  H140848  H140914  H140976  H270767  H271669  H271972  H272162  H272333  H272493  H272953  H140137  H140715  H140783  H140849  H140915  H140977  H271031  H271678  H271973  H272163  H272336  H272497  H272956  H140138  H140716  H140784  H140850  H140917  H140978  H271105  H271681  H271974  H272168  H272339  H272500  H272960  H140139  H140718  H140785  H140851  H140918  H140979  H271157  H271693  H271975  H272169  H272341  H272801  H272961  H140141  H140719  H140786  H140852  H140919  H140980  H271158  H271712  H271987  H272172  H272343  H272803  H272963  H140143  H140720  H140787  H140853  H140920  H140981  H271167  H271713  H271993  H272173  H272345  H272804  H272964  H140145  H140721  H140788  H140854  H140921  H140982  H271214  H271714  H271994  H272174  H272346  H272806  H272966  H140147  H140722  H140789  H140855  H140922  H140983  H271227  H271720  H272001  H272178  H272349  H272811  H272968  H140151  H140723  H140790  H140856  H140923  H140984  H271242  H271724  H272003  H272179  H272352  H272816  H272971  H272973  H273137  H290282  H290559  H291067  H291143  H291212  H293307  H293383  H293450  H293528  H293602  H293674  H272978  H273139  H290287  H290560  H291069  H291144  H291213  H293309  H293384  H293453  H293529  H293603  H293676  H272981  H273140  H290294  H290561  H291070  H291146  H291214  H293310  H293385  H293455  H293530  H293604  H293677  H272982  H273145  H290295  H290600  H291071  H291147  H291215  H293311  H293386  H293456  H293531  H293605  H293678  H272985  H273147  H290304  H291001  H291072  H291148  H291216  H293313  H293387  H293457  H293532  H293606  H293679  H272990  H273148  H290308  H291002  H291073  H291149  H291217  H293314  H293388  H293458  H293534  H293607  H293681  H272996  H273149  H290317  H291003  H291074  H291150  H291218  H293318  H293389  H293461  H293535  H293608  H293682  H272998  H273150  H290321  H291004  H291075  H291151  H291219  H293319  H293390  H293462  H293536  H293609  H293683  H272999  H273151  H290326  H291005  H291077  H291152  H291220  H293320  H293391  H293464  H293537  H293610  H293684  H273003  H273152  H290341  H291006  H291079  H291154  H291221  H293321  H293392  H293465  H293538  H293611  H293685  H273004  H273153  H290352  H291007  H291080  H291155  H291223  H293322  H293393  H293466  H293539  H293612  H293686  H273006  H273155  H290358  H291008  H291081  H291156  H291224  H293323  H293394  H293467  H293540  H293613  H293688  H273011  H273157  H290362  H291009  H291082  H291157  H291225  H293324  H293395  H293470  H293541  H293614  H293689  H273016  H273159  H290371  H291010  H291083  H291158  H291226  H293325  H293396  H293472  H293542  H293617  H293690  H273026  H273163  H290372  H291011  H291084  H291159  H291227  H293326  H293397  H293473  H293543  H293618  H293691  H273027  H273164  H290376  H291012  H291085  H291160  H291228  H293327  H293398  H293475  H293544  H293619  H293693  H273029  H273165  H290380  H291013  H291086  H291161  H291229  H293328  H293399  H293476  H293545  H293620  H293695  H273030  H273166  H290386  H291014  H291087  H291162  H291230  H293329  H293401  H293477  H293546  H293621  H293696  H273031  H273169  H290392  H291015  H291088  H291163  H291231  H293330  H293402  H293478  H293548  H293622  H293697  H273040  H273170  H290397  H291017  H291089  H291164  H291232  H293331  H293403  H293479  H293550  H293623  H293698  H273041  H273172  H290399  H291018  H291090  H291165  H291233  H293335  H293404  H293480  H293551  H293626  H293699  H273042  H273173  H290400  H291019  H291091  H291166  H291234  H293336  H293405  H293481  H293553  H293627  H293700  H273049  H273174  H290401  H291020  H291092  H291167  H291237  H293337  H293406  H293482  H293554  H293628  H293701  H273051  H273175  H290402  H291021  H291093  H291168  H291238  H293338  H293407  H293483  H293555  H293630  H293702  H273052  H273176  H290403  H291022  H291094  H291169  H291239  H293339  H293408  H293484  H293556  H293631  H293703  H273053  H273177  H290405  H291023  H291095  H291170  H291240  H293340  H293409  H293485  H293558  H293632  H293704  H273054  H273178  H290414  H291024  H291096  H291171  H291241  H293341  H293411  H293486  H293559  H293634  H293705  H273055  H273179  H290422  H291025  H291097  H291172  H291242  H293342  H293412  H293488  H293560  H293635  H293708  H273056  H273180  H290423  H291027  H291099  H291173  H291243  H293343  H293414  H293489  H293561  H293636  H293709  H273057  H273181  H290500  H291028  H291102  H291174  H291245  H293344  H293415  H293490  H293562  H293637  H293710  H273058  H273182  H290501  H291029  H291103  H291175  H291247  H293346  H293416  H293491  H293563  H293638  H293712  H273063  H273183  H290503  H291030  H291105  H291176  H291248  H293347  H293417  H293492  H293564  H293639  H293713  H273065  H273184  H290504  H291033  H291107  H291177  H291249  H293348  H293418  H293493  H293565  H293640  H293714  H273066  H273186  H290505  H291034  H291108  H291178  H291250  H293351  H293419  H293494  H293567  H293642  H293715  H273072  H273187  H290508  H291035  H291109  H291179  H291251  H293352  H293420  H293495  H293568  H293643  H293716  H273074  H273190  H290509  H291036  H291111  H291180  H291252  H293353  H293421  H293496  H293569  H293645  H293717  H273075  H273191  H290510  H291037  H291112  H291181  H291253  H293354  H293423  H293497  H293570  H293647  H293718  H273077  H273193  H290512  H291039  H291113  H291182  H291254  H293355  H293424  H293498  H293571  H293648  H293722  H273080  H273194  H290513  H291040  H291114  H291185  H291255  H293356  H293425  H293499  H293572  H293649  H293723  H273081  H273195  H290517  H291041  H291115  H291186  H291256  H293357  H293426  H293500  H293573  H293650  H293724  H273082  H273198  H290519  H291043  H291116  H291187  H291258  H293358  H293427  H293501  H293574  H293652  H293725  H273083  H273199  H290520  H291044  H291117  H291189  H291259  H293360  H293428  H293502  H293575  H293653  H293726  H273084  H273200  H290525  H291045  H291118  H291190  H291260  H293361  H293429  H293503  H293577  H293654  H293727  H273088  H273201  H290529  H291046  H291119  H291191  H291261  H293363  H293430  H293505  H293579  H293655  H293728  H273092  H273202  H290530  H291047  H291120  H291192  H291262  H293364  H293431  H293506  H293584  H293656  H293729  H273100  H273203  H290531  H291048  H291121  H291193  H291264  H293365  H293432  H293507  H293585  H293657  H293731  H273101  H273206  H290533  H291049  H291123  H291194  H291265  H293366  H293433  H293508  H293586  H293658  H293732  H273102  H273208  H290535  H291050  H291124  H291195  H291267  H293367  H293434  H293509  H293587  H293659  H293733  H273103  H273209  H290540  H291051  H291125  H291196  H291268  H293368  H293435  H293510  H293588  H293660  H293735  H273105  H273210  H290541  H291052  H291126  H291198  H291269  H293369  H293436  H293511  H293589  H293661  H293736  H273106  H290186  H290547  H291053  H291127  H291199  H291270  H293370  H293437  H293512  H293590  H293662  H293737  H273107  H290192  H290548  H291054  H291128  H291200  H291271  H293371  H293438  H293513  H293591  H293663  H293738  H273108  H290208  H290549  H291055  H291129  H291201  H291272  H293372  H293440  H293515  H293592  H293664  H293739  H273109  H290222  H290550  H291056  H291130  H291202  H293009  H293374  H293441  H293516  H293593  H293665  H293740  H273112  H290224  H290551  H291058  H291131  H291203  H293207  H293375  H293442  H293517  H293594  H293666  H293741  H273113  H290241  H290552  H291059  H291132  H291204  H293300  H293376  H293443  H293518  H293595  H293667  H293742  H273114  H290243  H290553  H291060  H291133  H291205  H293301  H293377  H293444  H293519  H293596  H293668  H293743  H273117  H290248  H290554  H291061  H291135  H291206  H293302  H293378  H293445  H293520  H293597  H293669  H293745  H273125  H290256  H290555  H291062  H291137  H291207  H293303  H293379  H293446  H293521  H293598  H293670  H293746  H273128  H290272  H290556  H291063  H291138  H291208  H293304  H293380  H293447  H293525  H293599  H293671  H293747  H273134  H290274  H290557  H291064  H291140  H291209  H293305  H293381  H293448  H293526  H293600  H293672  H293748  H273136  H290278  H290558  H291065  H291141  H291211  H293306  H293382  H293449  H293527  H293601  H293673  H293749  H293750  H293823  H293894  H293964  H294035  H294104  H294175  H294248  H294324  H294388  H294464  H294535  H294602  H293751  H293825  H293895  H293966  H294036  H294105  H294176  H294249  H294325  H294389  H294465  H294536  H294603  H293752  H293826  H293897  H293967  H294037  H294106  H294177  H294250  H294326  H294390  H294466  H294537  H294604  H293753  H293827  H293898  H293968  H294038  H294108  H294178  H294252  H294327  H294391  H294467  H294538  H294605  H293754  H293828  H293899  H293970  H294039  H294109  H294179  H294253  H294328  H294392  H294468  H294539  H294606  H293756  H293829  H293900  H293971  H294041  H294110  H294180  H294254  H294329  H294393  H294469  H294540  H294607  H293758  H293830  H293901  H293972  H294042  H294111  H294182  H294256  H294330  H294394  H294470  H294541  H294608  H293759  H293831  H293902  H293974  H294043  H294112  H294183  H294257  H294331  H294395  H294471  H294542  H294611  H293760  H293832  H293903  H293975  H294044  H294113  H294184  H294259  H294332  H294397  H294472  H294543  H294612  H293761  H293833  H293904  H293976  H294045  H294114  H294185  H294260  H294333  H294398  H294473  H294544  H294613  H293762  H293834  H293905  H293977  H294046  H294116  H294186  H294261  H294334  H294399  H294474  H294545  H294614  H293763  H293835  H293906  H293978  H294047  H294117  H294187  H294263  H294335  H294401  H294475  H294546  H294615  H293764  H293836  H293907  H293979  H294048  H294118  H294188  H294264  H294336  H294402  H294476  H294547  H294616  H293765  H293837  H293908  H293980  H294049  H294119  H294189  H294265  H294337  H294403  H294477  H294548  H294617  H293766  H293838  H293909  H293981  H294050  H294120  H294190  H294266  H294338  H294404  H294478  H294549  H294618  H293767  H293839  H293910  H293982  H294051  H294121  H294191  H294267  H294339  H294405  H294479  H294550  H294620  H293768  H293840  H293911  H293983  H294052  H294122  H294194  H294268  H294340  H294406  H294480  H294551  H294621  H293769  H293841  H293912  H293984  H294053  H294123  H294196  H294269  H294341  H294408  H294481  H294552  H294622  H293770  H293842  H293913  H293985  H294054  H294125  H294197  H294270  H294342  H294409  H294482  H294553  H294623  H293771  H293843  H293914  H293986  H294055  H294126  H294199  H294272  H294343  H294410  H294483  H294554  H294624  H293772  H293844  H293915  H293987  H294056  H294127  H294200  H294273  H294345  H294411  H294484  H294555  H294625  H293773  H293845  H293916  H293988  H294057  H294128  H294202  H294274  H294346  H294412  H294485  H294556  H294626  H293774  H293846  H293917  H293989  H294059  H294129  H294203  H294275  H294347  H294413  H294486  H294557  H294628  H293775  H293847  H293918  H293990  H294061  H294130  H294204  H294277  H294348  H294414  H294488  H294558  H294629  H293776  H293848  H293919  H293991  H294062  H294131  H294206  H294279  H294349  H294415  H294490  H294559  H294630  H293777  H293849  H293920  H293993  H294063  H294132  H294207  H294281  H294350  H294416  H294491  H294560  H294631  H293778  H293850  H293921  H293994  H294064  H294133  H294208  H294282  H294351  H294417  H294493  H294564  H294632  H293780  H293851  H293923  H293995  H294065  H294134  H294209  H294284  H294352  H294418  H294494  H294565  H294633  H293781  H293852  H293924  H293996  H294066  H294135  H294210  H294285  H294353  H294420  H294495  H294567  H294634  H293783  H293854  H293925  H293997  H294067  H294136  H294211  H294286  H294354  H294421  H294496  H294568  H294636  H293784  H293855  H293926  H293998  H294068  H294137  H294212  H294287  H294355  H294422  H294497  H294569  H294637  H293785  H293856  H293928  H293999  H294069  H294139  H294213  H294288  H294356  H294423  H294498  H294570  H294638  H293786  H293858  H293929  H294000  H294070  H294140  H294214  H294289  H294357  H294424  H294499  H294571  H294639  H293787  H293859  H293930  H294001  H294072  H294141  H294215  H294290  H294358  H294426  H294500  H294572  H294640  H293789  H293860  H293931  H294002  H294073  H294142  H294216  H294291  H294359  H294427  H294501  H294573  H294641  H293790  H293862  H293933  H294003  H294074  H294143  H294217  H294292  H294360  H294428  H294504  H294574  H294642  H293791  H293863  H293934  H294004  H294075  H294146  H294218  H294294  H294361  H294429  H294505  H294575  H294643  H293792  H293864  H293935  H294007  H294076  H294147  H294219  H294295  H294362  H294430  H294507  H294576  H294644  H293793  H293866  H293936  H294008  H294077  H294148  H294220  H294296  H294363  H294431  H294508  H294577  H294645  H293794  H293867  H293937  H294009  H294078  H294149  H294221  H294297  H294364  H294432  H294509  H294578  H294646  H293796  H293868  H293938  H294011  H294079  H294150  H294222  H294298  H294365  H294433  H294510  H294579  H294647  H293798  H293869  H293940  H294012  H294080  H294151  H294224  H294299  H294366  H294434  H294511  H294580  H294648  H293801  H293870  H293941  H294013  H294081  H294153  H294225  H294300  H294367  H294435  H294513  H294581  H294650  H293802  H293871  H293942  H294014  H294083  H294154  H294227  H294301  H294369  H294436  H294514  H294582  H294652  H293803  H293872  H293944  H294015  H294084  H294155  H294228  H294302  H294370  H294437  H294515  H294583  H294653  H293804  H293873  H293945  H294016  H294085  H294156  H294229  H294303  H294371  H294438  H294516  H294584  H294654  H293805  H293875  H293946  H294017  H294086  H294157  H294230  H294304  H294372  H294439  H294517  H294585  H294655  H293806  H293878  H293947  H294018  H294087  H294158  H294231  H294305  H294373  H294440  H294518  H294586  H294656  H293807  H293880  H293948  H294020  H294088  H294160  H294232  H294306  H294374  H294442  H294519  H294587  H294657  H293808  H293881  H293950  H294021  H294089  H294161  H294233  H294308  H294375  H294443  H294520  H294588  H294658  H293810  H293882  H293951  H294022  H294090  H294162  H294235  H294310  H294376  H294444  H294521  H294589  H294660  H293811  H293883  H293952  H294023  H294091  H294163  H294236  H294311  H294377  H294445  H294522  H294591  H294661  H293812  H293884  H293953  H294024  H294092  H294164  H294237  H294313  H294378  H294446  H294523  H294592  H294664  H293814  H293885  H293954  H294025  H294093  H294165  H294238  H294314  H294379  H294448  H294524  H294593  H294665  H293815  H293886  H293955  H294026  H294094  H294166  H294239  H294315  H294380  H294449  H294525  H294594  H294666  H293816  H293887  H293956  H294028  H294095  H294167  H294240  H294316  H294381  H294450  H294526  H294595  H294667  H293817  H293888  H293957  H294029  H294098  H294168  H294241  H294317  H294382  H294458  H294528  H294596  H294668  H293818  H293889  H293958  H294030  H294099  H294169  H294242  H294319  H294383  H294459  H294530  H294597  H294669  H293819  H293890  H293959  H294031  H294100  H294170  H294243  H294320  H294384  H294460  H294531  H294598  H294670  H293820  H293891  H293960  H294032  H294101  H294171  H294244  H294321  H294385  H294461  H294532  H294599  H294671  H293821  H293892  H293961  H294033  H294102  H294172  H294245  H294322  H294386  H294462  H294533  H294600  H294674  H293822  H293893  H293962  H294034  H294103  H294174  H294246  H294323  H294387  H294463  H294534  H294601  H294675  H294677  H294747  H294820  H294886  H294954  H295023  H295091  H295160  H295239  H295312  H295398  H295471  H295547  H294678  H294748  H294821  H294887  H294955  H295024  H295092  H295161  H295240  H295313  H295399  H295472  H295548  H294679  H294749  H294822  H294888  H294956  H295026  H295093  H295162  H295241  H295314  H295401  H295474  H295549  H294680  H294750  H294823  H294889  H294957  H295027  H295094  H295163  H295242  H295316  H295402  H295475  H295550  H294681  H294751  H294824  H294890  H294958  H295028  H295095  H295165  H295244  H295317  H295403  H295476  H295551  H294682  H294752  H294826  H294891  H294959  H295029  H295096  H295166  H295245  H295319  H295405  H295479  H295552  H294683  H294753  H294827  H294892  H294960  H295030  H295097  H295167  H295246  H295320  H295406  H295481  H295553  H294684  H294754  H294828  H294893  H294961  H295031  H295098  H295168  H295247  H295321  H295407  H295482  H295554  H294685  H294755  H294829  H294894  H294962  H295032  H295099  H295170  H295248  H295322  H295410  H295484  H295557  H294686  H294756  H294830  H294895  H294963  H295033  H295100  H295172  H295249  H295323  H295412  H295485  H295558  H294687  H294757  H294831  H294897  H294964  H295034  H295101  H295173  H295250  H295324  H295414  H295486  H295559  H294688  H294758  H294832  H294899  H294965  H295035  H295102  H295174  H295251  H295325  H295415  H295487  H295560  H294690  H294759  H294833  H294900  H294967  H295036  H295103  H295175  H295252  H295326  H295416  H295488  H295561  H294691  H294760  H294834  H294901  H294968  H295037  H295105  H295176  H295253  H295327  H295417  H295489  H295563  H294692  H294761  H294835  H294902  H294969  H295038  H295106  H295177  H295254  H295328  H295419  H295490  H295565  H294694  H294762  H294836  H294903  H294970  H295039  H295107  H295178  H295255  H295330  H295420  H295491  H295566  H294695  H294763  H294837  H294904  H294972  H295040  H295108  H295179  H295256  H295331  H295421  H295492  H295568  H294696  H294766  H294838  H294905  H294974  H295041  H295109  H295180  H295262  H295332  H295422  H295493  H295570  H294697  H294767  H294839  H294906  H294975  H295042  H295110  H295181  H295263  H295333  H295423  H295494  H295571  H294698  H294768  H294840  H294907  H294976  H295043  H295111  H295182  H295264  H295336  H295424  H295495  H295572  H294701  H294769  H294841  H294908  H294977  H295044  H295112  H295184  H295265  H295337  H295425  H295496  H295573  H294702  H294771  H294842  H294910  H294978  H295045  H295114  H295187  H295266  H295339  H295426  H295498  H295575  H294704  H294772  H294843  H294911  H294979  H295046  H295115  H295188  H295268  H295342  H295427  H295499  H295576  H294705  H294773  H294844  H294912  H294980  H295047  H295116  H295192  H295269  H295343  H295428  H295500  H295577  H294706  H294774  H294845  H294914  H294981  H295048  H295118  H295193  H295270  H295345  H295429  H295501  H295579  H294707  H294775  H294846  H294915  H294982  H295049  H295119  H295195  H295271  H295346  H295431  H295502  H295581  H294708  H294776  H294847  H294916  H294983  H295050  H295120  H295196  H295272  H295350  H295433  H295503  H295582  H294710  H294777  H294848  H294917  H294984  H295051  H295121  H295198  H295273  H295352  H295434  H295504  H295584  H294711  H294778  H294850  H294918  H294986  H295053  H295122  H295199  H295275  H295353  H295435  H295505  H295585  H294712  H294779  H294851  H294919  H294987  H295054  H295123  H295200  H295276  H295354  H295436  H295506  H295586  H294713  H294780  H294853  H294920  H294988  H295056  H295124  H295201  H295277  H295355  H295437  H295507  H295587  H294714  H294781  H294854  H294921  H294989  H295057  H295125  H295202  H295278  H295356  H295438  H295508  H295588  H294715  H294782  H294855  H294923  H294990  H295058  H295126  H295203  H295279  H295357  H295439  H295510  H295589  H294716  H294783  H294856  H294924  H294991  H295059  H295127  H295204  H295280  H295358  H295440  H295511  H295590  H294717  H294785  H294857  H294925  H294993  H295060  H295128  H295205  H295281  H295360  H295441  H295512  H295591  H294718  H294786  H294858  H294926  H294995  H295061  H295130  H295206  H295282  H295362  H295442  H295514  H295592  H294719  H294788  H294859  H294927  H294996  H295063  H295131  H295207  H295284  H295363  H295443  H295515  H295593  H294720  H294789  H294860  H294928  H294997  H295064  H295132  H295208  H295285  H295366  H295444  H295517  H295594  H294721  H294792  H294861  H294929  H294998  H295065  H295133  H295210  H295286  H295367  H295445  H295519  H295596  H294722  H294793  H294862  H294930  H294999  H295066  H295134  H295211  H295287  H295368  H295446  H295521  H295600  H294723  H294794  H294863  H294931  H295000  H295067  H295136  H295214  H295288  H295369  H295447  H295522  H295601  H294724  H294795  H294864  H294932  H295001  H295068  H295138  H295215  H295289  H295370  H295448  H295523  H295602  H294726  H294796  H294865  H294933  H295003  H295069  H295139  H295216  H295290  H295371  H295449  H295524  H295603  H294727  H294797  H294866  H294934  H295004  H295070  H295140  H295219  H295291  H295372  H295450  H295525  H295604  H294728  H294798  H294867  H294935  H295005  H295071  H295141  H295220  H295293  H295375  H295451  H295527  H295605  H294729  H294799  H294868  H294936  H295006  H295072  H295142  H295221  H295295  H295377  H295452  H295528  H295606  H294730  H294801  H294869  H294937  H295007  H295073  H295143  H295222  H295296  H295378  H295453  H295529  H295607  H294732  H294802  H294870  H294938  H295008  H295074  H295144  H295223  H295297  H295381  H295455  H295530  H295608  H294733  H294803  H294871  H294939  H295009  H295075  H295145  H295224  H295298  H295382  H295456  H295531  H295609  H294734  H294804  H294872  H294940  H295010  H295076  H295147  H295225  H295299  H295384  H295457  H295532  H295610  H294735  H294806  H294873  H294941  H295011  H295077  H295148  H295226  H295300  H295385  H295458  H295533  H295614  H294736  H294807  H294874  H294942  H295012  H295078  H295149  H295227  H295301  H295386  H295459  H295534  H295615  H294737  H294808  H294875  H294944  H295013  H295079  H295150  H295228  H295302  H295387  H295460  H295535  H295616  H294738  H294809  H294876  H294945  H295014  H295080  H295151  H295229  H295303  H295388  H295461  H295536  H295617  H294739  H294810  H294877  H294946  H295015  H295081  H295152  H295230  H295304  H295389  H295462  H295537  H295618  H294740  H294811  H294879  H294947  H295016  H295082  H295153  H295231  H295305  H295390  H295463  H295538  H295619  H294741  H294812  H294880  H294948  H295017  H295084  H295154  H295232  H295306  H295391  H295464  H295539  H295626  H294742  H294813  H294881  H294949  H295018  H295085  H295155  H295233  H295307  H295392  H295465  H295540  H295627  H294743  H294814  H294882  H294950  H295019  H295086  H295156  H295234  H295308  H295393  H295467  H295541  H295628  H294744  H294815  H294883  H294951  H295020  H295087  H295157  H295236  H295309  H295394  H295468  H295542  H295629  H294745  H294816  H294884  H294952  H295021  H295089  H295158  H295237  H295310  H295395  H295469  H295543  H295630  H294746  H294818  H294885  H294953  H295022  H295090  H295159  H295238  H295311  H295397  H295470  H295545  H295633  H295634  H295714  H295790  H295870  H295659  H295738  H295813  H295912  H295685  H295760  H295839  Hcx5349  H295658  H295635  H295715  H295793  H295871  H295660  H295739  H295814  H295913  H295687  H295761  H295840  Hcx5376  H295736  H295638  H295716  H295794  H295873  H295661  H295740  H295815  H295922  H295688  H295763  H295841  Hcx5380  H295812  H295639  H295717  H295795  H295877  H295662  H295741  H295816  H295925  H295689  H295764  H295843  Hcx5418  H295911  H295641  H295718  H295796  H295878  H295663  H295742  H295818  H295926  H295690  H295765  H295844  Hcx5480  H295684  H295642  H295719  H295798  H295880  H295665  H295743  H295821  Hcx4662  H295692  H295767  H295846  Hcx5483  H295759  H295643  H295720  H295799  H295885  H295666  H295744  H295822  Hcx4852  H295693  H295769  H295847  Hcx5544  H295837  H295644  H295721  H295800  H295887  H295667  H295745  H295823  Hcx4913  H295694  H295770  H295848  Hcx5562  Hcx5340  H295646  H295722  H295801  H295891  H295668  H295746  H295824  Hcx4973  H295696  H295773  H295849  Hcx5625  H295711  H295647  H295723  H295802  H295892  H295669  H295747  H295825  Hcx4994  H295697  H295775  H295850  Hcx5705  H295786  H295648  H295724  H295803  H295893  H295670  H295749  H295826  Hcx5055  H295699  H295776  H295851  Hcx5762  H295867  H295649  H295725  H295804  H295895  H295672  H295750  H295827  Hcx5088  H295701  H295777  H295853  Hcx5771  H295650  H295726  H295805  H295896  H295673  H295751  H295828  Hcx5189  H295702  H295778  H295854  Hcx5781  H295651  H295727  H295806  H295901  H295674  H295752  H295830  Hcx5194  H295703  H295779  H295856  H295712  H295652  H295728  H295807  H295905  H295675  H295753  H295831  Hcx5243  H295704  H295780  H295858  H295713  H295653  H295731  H295808  H295906  H295677  H295754  H295832  Hcx5274  H295707  H295782  H295859  H295787  H295654  H295732  H295809  H295908  H295678  H295755  H295833  Hcx5283  H295708  H295783  H295861  H295789  H295655  H295733  H295810  H295909  H295679  H295757  H295834  Hcx5315  H295709  H295784  H295862  H295868  H295656  H295734  H295811  H295910  H295683  H295758  H295836  Hcx5329  H295710  H295785  H295863  H295869.  Update 11-4-14-16 new installations and serial numbers added.  H294985  H295779  H295949  H141042  H295715  H295925  H295958  H141026  H295716  H295859  H295966  HCX5762  H295747  H295877  H295974  HCX5771  H294558   H296014  H294559   H296016  H295712   H296017  H295991   H296031.",
      "more_code_info": ""
    }
  ]
}