  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Woburn",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88485",
      "recalling_firm": "Lymol Medical Corporation",
      "address_1": "4 Plympton St",
      "address_2": "N/A",
      "postal_code": "01801-2917",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-2562-2021",
      "product_description": "LYMOL Medical Elite X Class Rigid Bronchoscope System, Part No. BD2410, BB2701, BB2702, BB2401, BB2402, BA2700, BT2000, BT2101, BT2103, BT2105, BT2106, BT2210, BT2201, BT2203, BT2205, BR2200, BP2200, BI2200, BR2000, BP2000, BI2999, BR2103, BP2103, BI2103, BR2101, BP2101, BI2101, BI2001, BDL2420",
      "product_quantity": "309",
      "reason_for_recall": "Reprocessing instructions were revised November 15, 2020, to reflect new processes and equipment used at consignee user facilities. New instructions were not sent to all consignees of the devices at the time of the revision.",
      "recall_initiation_date": "20210729",
      "center_classification_date": "20210924",
      "termination_date": "20221004",
      "report_date": "20211006",
      "code_info": "All distributed lots"
    }
  ]
}