  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Batesville",
      "address_1": "1069 State Road 46 E",
      "reason_for_recall": "The attachment latch mechanism can become bent depending on user handling.  Latch mechanism bending could lead to improper engagement of the foot section and this could result in injury to the user due to a fall.",
      "address_2": "",
      "product_quantity": "13,000 units",
      "code_info": "Model #'s: P3700B, P3700C, P3700D and P3700E",
      "center_classification_date": "20140901",
      "distribution_pattern": "Worldwide Distribution - US Nationwide and the following countries: Argentina , Aruba, Australia, Austria, Bahamas, Bahrain, Belgium,  Belize, Bermuda, Brazil, Brunei, Bulgaria, Cayman Islands, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, Equatorial Guinea, Finland, France, Germany, Greece, Guam, Guyana, Hong Kong, Hungary, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kenya, Kuwait, Lebanon, Libya, Malaysia, Mauritius, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turks and Caicos Islands, United Arab Emirates, United Kingdom,  US Virgin Islands, Venezuela and Vietnam.",
      "state": "IN",
      "product_description": "Affinity¿ Four Birthing Bed,     Product Usage:  Product is intended to be used as birthing beds for women of child bearing age in an LDR (Labor, Delivery, Recovery) or LDRP (Labor, Delivery, Recovery, Postpartum) setting within the acute care Labor and Delivery market.",
      "report_date": "20140910",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Hill-Rom, Inc.",
      "recall_number": "Z-2562-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "68650",
      "termination_date": "20170714",
      "more_code_info": "",
      "recall_initiation_date": "20140626",
      "postal_code": "47006-7520",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}