      "status": "Terminated",
      "city": "Golden",
      "state": "CO",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85650",
      "recalling_firm": "CME America, LLC",
      "address_1": "14998 W 6th Ave Ste 830",
      "address_2": "N/A",
      "postal_code": "80401-5025",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S: CA, IN, LA, MO, FL, IL, ME, MN, CO, VA, OK, NE, GA, AL, KY, IA, MS, ID, TX, SC, OH, SD, NV, MI, WY, NC, PA, TN, KS, NJ, OR, AR, NM, MA, NY, DC, RI, MD, WA, WI, AZ, NH, UT, HI, AK, MT, ND, VT, CT.    OUS (Foreign): Canada, UK",
      "recall_number": "Z-2559-2020",
      "product_description": "BodyGuard Microset, REF: A120-112XSFK",
      "product_quantity": "1273892",
      "reason_for_recall": "Infusion administration sets may have a delivery inaccuracy over plus/minus 13%, ranging from minus 42.8% to plus 17.5%. Use of the infusion pump system could cause over-infusion or under-infusion of therapy",
      "recall_initiation_date": "20200616",
      "center_classification_date": "20200715",
      "termination_date": "20220509",
      "report_date": "20200722",
      "code_info": "All lots within expiry"
    }
  ]
}