  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Buffalo Grove",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80229",
      "recalling_firm": "Leica Microsystems, Inc.",
      "address_1": "1700 Leider Ln",
      "address_2": "N/A",
      "postal_code": "60089-6622",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution US Nationwide in the states of :  CA, CO, CT, FL, HI, ID, IL, KS, KY, LA, MD, MI, MN, MO, NC, NJ, NY, OH, PA, SD, TN, TX.      International Distribution: Canada, AUSTRALIA, BELGIUM, Bolivia, BRAZIL, BULGARIA, CHILE, CHINA, DENMARK, EGYPT, EL SALVADOR, FINLAND,   FRANCE, GERMANY, HONGKONG, HUNGARY, INDIA, ISRAEL, ITALY, JAPAN, KUWAIT, MALAYSIA, PHILIPPINES, POLAND, PORTUGAL, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, TAIWAN,THAILAND, TURKEY, UK",
      "recall_number": "Z-2555-2018",
      "product_description": "Leica Microsystems products labeled as the following: Leica M320 Microscope Systems with Article Numbers 10448419, 10448420, 10448485, 10448486    Product Usage:  Intended use of this device is an optical instrument for magnifying and illuminating specimens. It can be applied for observation and documentation and for human and veterinary medical treatment. Not for use in Eye Operations.",
      "product_quantity": "8,564 Total Units",
      "reason_for_recall": "Possible failure of screws which hold the gas spring fixation bracket in place.  Due to this failure, the gas spring, which is fixed to the upper part of the parallelogram, cannot hold up the weight of the optics carrier.",
      "recall_initiation_date": "20180604",
      "center_classification_date": "20180726",
      "termination_date": "20210325",
      "report_date": "20180801",
      "code_info": "A. Article Number 10448419 (Serial Numbers 160714001, 10814001, 200814001, 200814002, 90914001, 021014001, 21214001, 31214001, 81214001, 190115001, 190115002);      B Article Number 10448420 (Serial Numbers 180514001, 180514002, 190514001, 190514002, 190514003, 190514004, 200514001, 200514002, 200514003, 200514004, 210514001, 210514002, 210514003, 210514004, 220514001, 220514002, 220514003, 260514001, 260514002, 270514001, 270514002, 270514003, 270514004, 270514005, 280514001, 280514002, 280514003, 290514001, 300514001, 300514002, 300514003, 20614001, 40614001, 50614001, 50614002, 50614003, 60614001, 90614001, 90614002, 100614001, 100614002, 100614003, 120614001, 120614002, 120614003, 130614001, 130614002, 130614003, 130614004, 90514002, 160614001, 160614002, 160614003, 160614004, 170614001, 190614001, 200614001, 200614002, 200614003, 230614001, 230614002, 230614003, 240614001, 250614001, 250614002, 250614003, 250614004, 250614005, 270614001, 300614001, 300614002, 300614003, 10714001, 10714002, 10714003, 10714004, 20714001, 20714002, 20714003, 30714001, 30714002, 40714001, 40714002, 40714003, 70714001, 70714002, 70714003, 80714001, 80714002, 80714003, 90714001, 100714001, 110714001, 140714001, 160714001, 160714002, 170714001, 170714002, 180714001, 180714002, 180714003, 210714001, 210714002, 210714003, 220714001, 220714002, 220714003, 220714004, 230714001, 250714001, 250714002, 250714003, 250714004, 250714005, 250714006, 290714001, 290714002, 290714003, 290714004, 300714001, 300714002, 300714003, 300714004, 310714001, 310714002, 310714003, 310714004, 10814001, 10814002, 040814001, 040814002, 040814003, 040814004, 050814001, 050814002, 050814003, 050814004, 060814001, 060814002, 070814001, 070814002, 070814003, 070814004, 080814001, 080814002, 120814001, 120814002, 130814001, 130814002, 130814003, 130814004, 140814001, 140814002, 140814003, 140814004, 150814001, 150814002, 150814003, 160814001, 160814002, 160814003, 180814001, 180814002, 180814003, 190814001, 190814002, 190814003, 190814004, 190814005, 190814006, 200814001, 200814002, 210814001, 210814002, 210814003, 210814004, 220814001, 220814002, 220814003, 250814001, 250814002, 270814001, 270814002, 270814003, 270814004, 270814005, 270814006, 280814001, 280814002, 280814003, 280814004, 290814001, 290814002, 10914001, 10914002, 10914003, 10914004, 20914001, 20914002, 30914001, 30914002, 30914003, 80714002, 70814003, 40914001, 40914002, 50914001, 50914002, 80914001, 80914002, 80914003, 80914004, 80914005, 90914001, 90914002, 90914003, 100914001, 100914002, 100914003, 100914004, 110914001, 110914002, 110914003, 120914001, 120914002, 120914003, 120914004, 120914005, 130914001, 130914002, 130914003, 150914001, 160914001, 160914002, 160914003, 160914004, 160914005, 170914001, 170914002, 170914003, 170914004, 170914005, 180914001, 180914002, 180914003, 180914004, 180914005, 190914001, 190914002, 190914003, 200914001, 200914002, 200914003, 200914004, 200914005, 220914001, 220914002, 220914003, 220914004, 230914001, 230914002, 90614001, 90614002, 240914001, 240914002, 240914003, 250914001, 250914002, 100614002, 100614003, 260914001, 260914002, 260914003, 260914004, 290914001, 290914002, 290914003, 300914001, 300914002, 300914003, 021014001, 021014002, 031014001, 031014002, 031014003, 031014004, 41014001, 41014002, 41014003, 41014004, 41014005, 41014006, 41014007, 61014001, 61014002, 71014001, 71014002, 71014003, 71014004, 81014001, 81014002, 81014003, 81014004, 81014005, 91014001, 91014002, 101014001, 101014002, 101014003, 101014004, 121014001, 121014002, 121014003, 121014004, 121014005, 121014006, 131014001, 141014001, 141014002, 141014003, 141014004, 151014001, 151014002, 151014003, 151014004, 161014001, 161014002, 161014003, 171014001, 171014002, 171014003, 171014004, 181014001, 181014002, 181014003, 181014004, 191014001, 191014002, 191014003, 191014004, 201014001, 201014002, 201014003, 211014001, 211014002, 211014003, 211014004, 231014001, 231014002, 231014003, 241014001, 241014002, 241014003, 241014004, 241014005, 281014001, 281014002, 281014003, 281014004, 281014005, 291014001, 291014002, 291014003, 291014004, 291014005, 291014006, 301014001, 301014002, 311014001, 311014002, 311014003, 311014004, 11114001, 11114002, 11114003, 11114004, 11114005, 31114001, 31114002, 31114003, 31114004, 31114005, 41114001, 41114002, 41114003, 51114001, 51114002, 51114003, 51114004, 81114001, 81114002, 81114003, 81114004, 101114001, 101114002, 101114003, 111114001, 111114002, 121114001, 121114002, 121114003, 121114004, 131114001, 131114002, 141114001, 141114002, 141114003,  161114001, 161114002, 161114003, 161114004, 161114005, 161114006, 171114001, 171114002, 171114003, 171114004, 181114001, 181114002, 181114003, 191114001, 191114002, 191114003, 201114001, 201114002,  201114003, 201114004, 201114005, 211114001, 211114002, 211114003, 211114004, 241114001, 241114002, 241114003, 241114004, 251114001, 251114002, 251114003, 251114004, 261114001, 261114002, 261114003,  261114004, 271114001, 271114002, 281114001, 281114002, 11214001, 11214002, 21214001, 31214001, 31214002, 31214003, 41214001, 41214002, 41214003, 51214001, 51214002, 71214001, 71214002, 81214001, 81214002,  81214003, 91214001, 91214002, 91214003, 91214004, 101214001, 101214002, 101214003, 111214001, 111214002, 111214003, 121214001, 121214002, 121214003, 121214004, 151214001, 151214002, 151214003, 151214004, 151214005, 151214006, 151214007, 151214008, 161214001, 161214002, 171214001, 171214002, 171214003, 181214001, 181214002, 181214003, 181214004, 191214001, 191214002, 191214003, 191214004, 191214005, 191214006, 211214001, 221214001, 221214002, 241214001, 241214002, 241214003, 261214001, 261214002, 261214003, 261214004, 271214001, 271214002, 271214003, 281214001, 281214002, 281214003, 20115001, 20115002, 20115003, 20115004, 50115001, 50115002, 60115001, 60115002, 60115003, 60115004, 60115005, 60115006, 70115001, 70115002, 80115001, 80115002, 90115001, 90115002, 90115003, 90115004, 110115001,  110115002, 120115002, 120115003, 120115004, 130115001, 130115002, 130115003, 130115004, 140115001, 140115002, 140115003, 140115004, 150115001, 150115002, 160115001, 160115002, 190115001, 190115002,  200115001, 200115002, 200115003, 200115004, 210115002, 210115004, 130215004, 110515005, 120115004, 130115003, 50915001);      C. Article Number 10448485 (Serial Numbers 180514001, 180514002, 230514001, 230514002, 230514003, 230514004, 260514001, 260514002, 280514001, 290514001, 20614001, 20614002, 20614003, 30614001, 30614002, 40614001, 40614002, 60614001, 60614002, 90614001, 90614002, 110614001, 110614002, 170614001, 170614002, 170614003, 170614004, 170614005, 180614001, 190614001, 190614002, 190614003, 190614004, 190614005, 200614001, 230614001, 300614001, 20714001, 30714001, 30714002, 40714001, 80714001, 90714001, 90714002, 100714001, 100714002, 110714001, 110714002, 140714001, 140714001, 170714001, 180714001, 210714001, 080814001, 080814002, 120814001, 120814002, 150814001, 220814001, 260814001, 260814002, 260814003, 260814004, 290814001, 290814002, 20914001, 20914002, 50914001, 50914002, 110914001, 130914001, 150914001, 150914002, 230914001, 240914001,  250914001, 250914002, 290914001, 300914001, 011014001, 011014002, 011014003, 011014004, 021014001, 41014001, 61014001, 91014001, 91014002, 101014001, 121014001, 151014001, 211014001, 231014001, 241014001, 241014002, 301014001, 301014002, 301014003, 311014001, 11114001, 11114002, 11114003, 41114001, 81114001, 101114001, 111114001, 111114002, 191114001, 211114001, 271114001, 271114002, 281114001, 11214001, 21214001, 21214002, 51214001, 71214001, 71214002, 71214003, 101214001, 111214001, 161214001, 161214002, 171214001, 191214001, 191214002, 221214001, 221214002, 231214001, 231214002, 231214003, 241214001, 251214001, 251214002, 251214003, 70115001, 80115001, 80115002, 150115001, 150115002, 160115001, 160115002);      D. Article Number 10448486 (Serial Numbers 110614001, 110614002, 120614001, 310114001, 060814001, 060814002, 160814001, 61014001, 121014001, 281014001, 131114001, 131114002, 181114001, 231214001)"
    }
  ]
}