  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Solon",
      "address_1": "31200 Solon Rd",
      "reason_for_recall": "A vendor changed the manufacturing process of a component used in the canister ring that may affect the performance of the ring.  The deficit display may be higher than the actual deficit when the fluid collection canister is fully loaded.",
      "address_2": "Unit 1",
      "product_quantity": "21",
      "code_info": "Model P4000  Serial numbers affected:   20110004, 20110007, 20110015, 20110025, 20130001-20130032, 20140001-20140005",
      "center_classification_date": "20140827",
      "distribution_pattern": "Nationwide Distribution including HI, MS, NY, OH, PA, VA, WA, WV, and IL.",
      "state": "OH",
      "product_description": "The Thermedx Fluid Management System is designed to provide irrigation fluids to irrigate, distend, and debride while warming the fluid to assist in the prevention of peri-operative hypothermia. The fluid is warmed via an internal infrared mechanism. Additionally the system will hold suction canisters for the collection of \u001cused\u001d irrigant.",
      "report_date": "20140903",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Thermedx LLC",
      "recall_number": "Z-2555-2014",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "68986",
      "termination_date": "20150804",
      "recall_initiation_date": "20140210",
      "postal_code": "44139-3556",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}