  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Orchard Park",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68961",
      "recalling_firm": "Curbell Medical, Inc.",
      "address_1": "20 Centre Dr",
      "address_2": "N/A",
      "postal_code": "14127-4102",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Nationwide, Canada, Ireland, Italy",
      "recall_number": "Z-2554-2014",
      "product_description": "CURBELL MEDICAL CareSense Wireless Advanced LCD Bed and Chair Monitor,  MODEL # CSM-BC400; --- COMMON/USUAL NAME: Bed Patient Monitor;  CLASSIFICATION NAME:  Bed Patient Monitor --- Monitor is connected to a sensor pad that monitors the patient if they attempt to arise from a bed, chair, floor mat, or toilet seat. The monitor will sound an alarm.",
      "product_quantity": "Domestic: 112 units; Foreign: 211 units",
      "reason_for_recall": "The firm became aware of a potential problem that was initiated by a customer complaint.  After consultation with the manufacturer, it was discovered that a resistor was incorrectly placed within the circuit board on the monitor.  This change to the resistor was a planned change to address a product improvement (improve battery drain).  However, this modification rendered the monitor's alarm via the nurse call communication port to not function.",
      "recall_initiation_date": "20140805",
      "center_classification_date": "20140826",
      "termination_date": "20170228",
      "report_date": "20140903",
      "code_info": "Lot 0414"
    }
  ]
}