  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "76629",
      "recalling_firm": "Zimmer Biomet, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide)    Foreign: Argentina, Australia, Belgium, Brazil, Bridgend Wales, Canada, Chile, Colombia, Dominican Republic, England, Eschbach, France, India, Italy, Japan, Jordan, Korea, Malaysia, Mexico, Netherlands, Nicaragua, Panama, Peru, Russia, Saudi Arabia, Singapore, Spain, Switzerland, Taiwan, Thailand",
      "recall_number": "Z-2553-2017",
      "product_description": "Fem IM Nail 15mmdx44cm, Sterile,   Rod, fixation, intramedullary and accessories- Specific Knee, Hip and Nail Implants",
      "product_quantity": "9867",
      "reason_for_recall": "Zimmer Biomet Inc. initiated a field action on 02/28/2017 for sterile-packaged implants packaged in two different package configurations due to packaging design verification test failures.",
      "recall_initiation_date": "20170228",
      "center_classification_date": "20170614",
      "termination_date": "20180227",
      "report_date": "20170621",
      "code_info": "Item Number: 00225244015, Lot Number Expiry Date Before: March 2027"
    }
  ]
}