  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68818",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution.",
      "recall_number": "Z-2552-2014",
      "product_description": "Knee Stem Extensions & Impactor Sleeve:  STEM IMPLANT 20MMDX145MM  STEM IMPLANT 22MMDX145MM  STEM IMPLANT 24MMDX145MM  SHARP FLUTED ST 20MMDX75MM  SHARP FLUTED ST 22MMDX75MM  SHARP FLTD ST 24MMDX75MM  SHARP FLUTED ST 20MMDX130MM  SHARP FLUTED ST 22MMDX130MM  SHARP FLUTED ST 24MMDX130MM  CEMENTED STEM EXT 13MM X 145MM  IMPACTOR SLEEVE*  NGK STEM EXT 12X245MM (200MM)  NGK STEM EXT 13X245MM (200MM)  NGK STEM EXT 14X245MM (200MM)  NGK STEM EXT 15X245MM (200MM)  NGK STEM EXT 16X245MM (200MM)  NGK STEM EXT 17X245MM (200MM)  NGK STEM EXT 18X245MM (200MM)  NGK STEM EXT 20X245MM (200MM)  NGK STEM EXT 22X245MM (200MM)  SMOOTH REV STM 9.5MM X 19  SMOOTH REV STM 11.5MM X 1  SMOOTH REV STM 13.5MM X 1  SMOOTH REV STM 15.5MM X 1  SMOOTH REV STM 17.5MM X 1  REV FEM STM 10.5MM X 200M  REV FEM STM 12.5MM X 200M  REV FEM STM 14.5MM X 200M  REV FEM STM 16.5MM X 200M  18.5X200MM REV STEM  REV FEM STM 10.5MM X 250M  REV FEM STM 12.5MM X 250M  REV FEM STM 14.5MM X 250M  REV FEM STM 16.5MM X 250M  12.5X155MM FLUTED OFFST R  14.5X155MM FLUTED OFFST R  16.5X155MM FLUTED OFFST R  18.5X155MM FLUTED OFFST R  20.5X155MM FLUTED OFFST R  12.5X175MM FLUTED REV STM  FLUTE REV STEM - 14.5MM X  FLUTE REV STEM - 16.5MM X  FLUTE REV STEM - 18.5MM X  FLUTE REV STEM - 20.5MM X  FLUTE REV STEM - 22.5MM X  FLUTED REV STM 10.5MM X 1  FLUTED REV STM 11.5MM X 1  FLUTED REV STM 12.5MM X 1  FLUTED REV STM 13.5MM X 1  FLUTED REV STM 14.5MM X 1  FLUTED REV STM 16.5MM X 1  FLUTED REV STM 18.5MM X 1",
      "product_quantity": "972,125 units",
      "reason_for_recall": "Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from  each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.",
      "recall_initiation_date": "20140722",
      "center_classification_date": "20140826",
      "termination_date": "20150915",
      "report_date": "20140903",
      "code_info": "Item Numbers:00598801020, 00598801022, 00598801024, 00598801520,  00598801522, 00598801524, 00598801620, 00598801622, 00598801624,  00598801913, 31810604200, 32598801212, 32598801213, 32598801214,  32598801215, 32598801216, 32598801217, 32598801218, 32598801220,  32598801222, 621519095, 621519115, 621519135, 621519155, 621519175,  621520105, 621520125, 621520145, 621520165, 621520185, 621525105, 621525125, 621525145, 621525165, 681515125, 681515145, 681515165,  681515185, 681515205, 681517125, 681517145, 681517165, 681517185,  681517205, 681517225, 681519105, 681519115, 681519125, 681519135,  681519145, 681519165, 681519185      EXPIRATION DATE PRIOR TO 2024-07"
    }
  ]
}