  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Hartland",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74846",
      "recalling_firm": "Merge Healthcare, Inc.",
      "address_1": "900 Walnut Ridge Dr",
      "address_2": "N/A",
      "postal_code": "53029-8347",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US Distribution to states of: CA, PA, TX, and  MT.",
      "recall_number": "Z-2551-2016",
      "product_description": "Merge Unity Z3D software.    Merge Unity PACS (formerly DR Systems PACS) is a medical image and information management system that allows viewing, selection, processing, printing, telecommunications, and media interchange of medical images from a variety of diagnostic imaging systems.",
      "product_quantity": "9 sites potentially have the affected version",
      "reason_for_recall": "The software is unable to accurately determine the calcium score of scans with a slice thickness not equal to 3 mm.",
      "recall_initiation_date": "20150723",
      "center_classification_date": "20160816",
      "termination_date": "20161007",
      "report_date": "20160824",
      "code_info": "Version 11.0"
    }
  ]
}