  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68818",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution.",
      "recall_number": "Z-2550-2014",
      "product_description": "Trauma Tubes and Plates:  TUBE & SCP PLT 90DX150LG  TUBE & SCP PLT 90DX205LG  TUBE & SCP PLT 90DX255LG  TUBE & SCP PLT 95DX150LG  TUBE & SCP PLT 95DX205LG  TUBE & SCP PLT 95DX255LG  TUBE&SCP PLT 130DX8H  TUBE&SCP PLT 130DX10H  TUBE&SCP PLT 130DX12H  TUBE&SCP PLT 130DX14H  TUBE&SCP PLT 135DX8H  TUBE&SCP PLT 135DX10H  TUBE&SCP PLT 135DX12H  TUBE&SCP PLT 135DX14H  TUBE&SCP PLT 140DX8H  TUBE&SCP PLT 140DX10H  TUBE&SCP PLT 140DX12H  TUBE&SCP PLT 140DX14H  TUBE&SCP PLT 145DX8H  TUBE&SCP PLT 145DX10H  TUBE&SCP PLT 145DX12H  TUBE&SCP PLT 145DX14H  TUBE&SCP PLT 150DX6H  TUBE&SCP PLT 150DX8H  TUBE&SCP PLT 150DX10H  TUBE&SCP PLT 150DX12H  TUBE&SCP PLT 150DX14H  VERSA-FX II SC TUBE PLT 90DX6H  VERSA-FX II SC TUBE PLT 90DX8H  VERSA-FXII SC TUBE PLT 90DX10H  VERSA-FXII SC TUBE PLT 90DX12H  VERSA-FXII SC TUBE PLT 90DX14H  VERSA-FXII SC TUBE PLT 95DX6H  VERSA-FXII SC TUBE PLT 95DX8H  VERSA-FXII SC TUBE PLT 95DX10H  VERSA-FXII SC TUBE PLT 95DX12H  VERSA-FXII SC TUBE PLT 95DX14H  VERSA-FX KEYLESS 90D X 8HOLE  VERSA-FX KEYLESS 90D X 11HOLE  VERSA-FX KEYLESS 95D X 8HOLE  VERSA-FX KEYLESS 95D X 11HOLE  VERSA-FX KEYLESS 95D X 14HOLE  VERSA-FX II STD TUBE 130DX8H  VERSA-FX II STD TUBE 130DX10H  VERSA-FX II STD TUBE 130DX12H  VERSA-FX II STD TUBE 130DX14H  VERSA-FX II STD TUBE 135DX8H  VERSA-FX II STD TUBE 135DX10H  VERSA-FX II STD TUBE 135DX12H  VERSA-FX II STD TUBE 135DX14H  VERSA-FX II STD TUBE 140DX8H  VERSA-FX II STD TUBE 140DX10H  VERSA-FX II STD TUBE 140DX12H  VERSA-FX II STD TUBE 140DX14H  VERSA-FX II STD TUBE 145DX8H  VERSA-FX II STD TUBE 145DX10H  VERSA-FX II STD TUBE 145DX12H  VERSA-FX II STD TUBE 145DX14H  VERSA-FX II STD TUBE 150DX8H  VERSA-FX II STD TUBE 150DX10H  VERSA-FX II STD TUBE 150DX12H  VERSA-FX II STD TUBE 150DX14H  VERSA-FX II KEYLESS TB 130DX8H  VERSA-FX II KEYLESS TB 135DX8H  VERSA-FX II KEYLESS TB 140DX8H  VERSA-FX II KEYLESS TB 145DX8H  VERSA-FX II KEYLESS TB 150DX8H  VERSA-FX KEYLESS 90D X 8HOLE  VERSA-FX KEYLESS 90D X 11HOLE  VERSA-FX KEYLESS 95D X 8HOLE  VERSA-FX KEYLESS 95D X 11HOLE  VERSA-FX KEYLESS 95D X 14HOLE",
      "product_quantity": "972,125 units",
      "reason_for_recall": "Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from  each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.",
      "recall_initiation_date": "20140722",
      "center_classification_date": "20140826",
      "termination_date": "20150915",
      "report_date": "20140903",
      "code_info": "Item Numbers:  00118109006, 00118109008, 00118109010, 00118109506, 00118109508, 00118109510, 00118113008, 00118113010, 00118113012, 00118113014, 00118113508, 00118113510, 00118113512, 00118113514, 00118114008, 00118114010, 00118114012, 00118114014, 00118114508, 00118114510, 00118114512, 00118114514, 00118115006, 00118115008, 00118115010, 00118115012, 00118115014, 00119309006, 00119309008, 00119309010, 00119309012, 00119309014, 00119309506, 00119309508, 00119309510, 00119309512, 00119309514, 00119809008, 00119809011, 00119809508, 00119809511, 00119809514, 00119313008, 00119313010, 00119313012, 00119313014, 00119313508, 00119313510, 00119313512, 00119313514, 00119314008, 00119314010, 00119314012, 00119314014, 00119314508, 00119314510, 00119314512, 00119314514, 00119315008, 00119315010, 00119315012, 00119315014, 00119413008, 00119413508, 00119414008, 00119414508, 00119415008, 00119809008, 00119809011, 00119809508, 00119809511, 00119809514            EXPIRATION DATE PRIOR TO 2024-07"
    }
  ]
}