  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68818",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution.",
      "recall_number": "Z-2549-2014",
      "product_description": "Trauma Large Screws:  CANN BONE SCREW 7.0MMX100MMLG  CANN BONE SCREW 7.0X105LG  CANN BONE SCREW 7.0X105LG  CANN BONE SCREW 7.0MMX105MMLG  CANN BONE SCREW 7.0X110LG  CANN BONE SCREW 7.0X110LG  CANN BONE SCREW 7.0MMX110MMLG  CANN BONE SCREW 7.0X115LG  CANN BONE SCREW 7.0X115LG  CANN BONE SCREW 7.0MMX115MMLG  CANN BONE SCREW 7.0X120LG  CANN BONE SCREW 7.0X120LG  CANN BONE SCREW 7.0MMX120MMLG  CANN BONE SCREW 7.0X125LG  CANN BONE SCREW 7.0MMX125MMLG  CANN BONE SCREW 7.0X130LG  CANN BONE SCREW 7.0MMX130MMLG  RECON SCR LGE 5.5MMDX105MM  RECON SCR LGE 5.5MMDX110MM  RECON SCR LGE 5.5MMDX115MM  RECON SCR LGE 5.5MMDX120MM  RECON SCR LGE 5.5MMDX125MM  RECON SCR LGE 5.5MMDX130MM  ITST ANTI-ROT SCR 6.5MMDX105MM  ITST ANTI-ROT SCR 6.5MMDX110MM  PERI.  SCR 3.5MM X105MML  PERI SCR. 6.5MM X 100MM L  PERI SCR. 6.5MM X 140MM L  PERI SCR. 6.5MM X 100MM L  PERI SCR. 6.5MM X 100MM L  3.5MM CORT SCR X 105MM SELFTAP  3.5MM CORT SCR X 110MM SELFTAP  5.5MM X 100MM CANN LKNG SCREW  Z NAIL 6.0 X 105 CANC SCREW PT  Z NAIL 6.0 X 110 CANC SCREW PT  Z NAIL 6.0 X 115 CANC SCREW PT  Z NAIL 6.0 X 120 CANC SCREW PT  Z NAIL 6.0 X 125 CANC SCREW PT  Z NAIL 6.0 X 130 CANC SCREW PT  Z NAIL 6.0 X 135 CANC SCREW PT  Z NAIL 6.0 X 140 CANC SCREW PT  Z NAIL 6.0 X 105 CANC SCREW FA  Z NAIL 6.0 X 110 CANC SCREW FA  Z NAIL 6.0 X 115 CANC SCREW FA  Z NAIL 6.0 X 120 CANC SCREW FA  Z NAIL 6.0 X 125 CANC SCREW FA  Z NAIL 6.0 X 130 CANC SCREW FA  Z NAIL 6.0 X 135 CANC SCREW FA  Z NAIL 6.0 X 140 CANC SCREW FA  Z NAIL 6.0 X 105 CANC SCREW PT  Z NAIL 6.0 X 110 CANC SCREW PT  Z NAIL 6.0 X 115 CANC SCREW PT  Z NAIL 6.0 X 120 CANC SCREW PT  Z NAIL 6.0 X 125 CANC SCREW PT  Z NAIL 6.0 X 130 CANC SCREW PT  Z NAIL 6.0 X 135 CANC SCREW PT  Z NAIL 6.0 X 140 CANC SCREW PT  Z NAIL 6.0 X 105 CANC SCREW FA  Z NAIL 6.0 X 110 CANC SCREW FA  Z NAIL 6.0 X 115CANC SCREW FA  Z NAIL 6.0 X 120CANC SCREW FA  Z NAIL 6.0 X 125 CANC SCREW FA  Z NAIL 6.0 X 130 CANC SCREW FA  Z NAIL 6.0 X 135 CANC SCREW FA  Z NAIL 6.0 X 140 CANC SCREW FA  3.5X105 CORT SCR  3.5 X 105 CORT SCR SELF TAP  3.5X110 CORT SCR  3.5 X 110 CORT SCR SELF TAP  3.5X115 CORT SCR  3.5 X 115 CORT SCR SELF TAP  3.5X120 CORT SCR  3.5 X 120 CORT SCR SELF TAP  3.5X125 CORT SCR  3.5X130 CORT SCR  3.5X135 CORT SCR  3.5X140 CORT SCR  3.5X145 CORT SCR  3.5X150 CORT SCR  4.5 X 105 CORT SCR SELF TAP  4.5 X 110 CORT SCR SELF TAP  4.5 X 120 CORT SCR SELF TAP  6.5X105 CANC SCR, FULLY THDED  6.5X105 CANC SCR, 16MM THD  6.5X105 CANC SCR, 32MM THD  6.5X110 CANC SCR, FULLY THDED  6.5X110 CANC SCR, 16MM THD  6.5X110 CANC SCR, 32MM THD  6.5X115 CANC SCR, 32MM THD  6.5X120 CANC SCR, 32MM THD",
      "product_quantity": "972,125 units",
      "reason_for_recall": "Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from  each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.",
      "recall_initiation_date": "20140722",
      "center_classification_date": "20140826",
      "termination_date": "20150915",
      "report_date": "20140903",
      "code_info": "Item Numbers 00114610099, 00114610500, 00114610532, 00114610599, 00114611000, 00114611032, 00114611099, 00114611500, 00114611532, 00114611599, 00114612000, 00114612032, 00114612099, 00114612500, 00114612599, 00114613000, 00114613099, 00225210555, 00225211055, 00225211555, 00225212055, 00225212555, 00225213055, 00225710565, 00225711065, 47234712005, 47234712400, 47234712440, 47234712500, 47234712600, 47234810535, 47234811035, 47235910055, 47248310560, 47248311060, 47248311560, 47248312060, 47248312560, 47248313060, 47248313560, 47248314060, 47248410560, 47248411060, 47248411560, 47248412060, 47248412560, 47248413060, 47248413560, 47248414060, 47348310560, 47348311060, 47348311560, 47348312060, 47348312560, 47348313060, 47348313560, 47348314060, 47348410560, 47348411060, 47348411560, 47348412060, 47348412560, 47348413060, 47348413560, 47348414060, 47483510500, 47483510501, 47483511000, 47483511001, 47483511500, 47483511501, 47483512000, 47483512001, 47483512500, 47483513000, 47483513500, 47483514000, 47483514500, 47483515000, 47484510501, 47484511001, 47484512001, 47486510500, 47486510501, 47486510502, 47486511000, 47486511001, 47486511002, 47486511502, 47486512002  EXPIRATION DATE PRIOR TO 2024-07"
    }
  ]
}