  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68818",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution.",
      "recall_number": "Z-2547-2014",
      "product_description": "Endo & Unipolar Femoral Heads:  ENDO FEMORAL HEAD 41MM DIA  ENDO FEMORAL HEAD 42MM DIA  ENDO FEMORAL HEAD 43MM DIA  ENDO FEMORAL HEAD 44MM DIA  ENDO FEMORAL HEAD 45MM DIA  ENDO FEMORAL HEAD 46MM DIA  ENDO FEMORAL HEAD 47MM DIA  ENDO FEMORAL HEAD 48MM DIA  ENDO FEMORAL HEAD 49MM DIA  ENDO FEMORAL HEAD 50MM DIA  ENDO FEMORAL HEAD 51MM DIA  ENDO FEMORAL HEAD 52MM DIA  ENDO FEMORAL HEAD 53MM DIA  ENDO FEMORAL HEAD 54MM DIA  ENDO FEMORAL HEAD 55MM DIA  ENDO FEMORAL HEAD 57MM DIA  ENDO FEMORAL HEAD 60MM DIA  ENDO FEMORAL HEAD 63MM DIA  ENDO FEM HEAD 41MM  ENDO FEM HEAD 42MM  ENDO FEM HEAD 43MM  ENDO PROS HEAD 44MM  ENDO FEM HEAD 45MM  ENDO FEM HEAD 46MM  ENDO FEM HEAD 47MM  ENDO FEM HEAD 48MM  ENDO FEM HEAD 49MM  ENDO FEM HEAD 50MM  ENDO FEM HEAD 51MM  ENDO FEM HEAD 52MM  ENDO FEM HEAD 53MM  ENDO FEM HEAD 54MM  ENDO FEM HEAD 55MM  ENDO FEM HEAD 57MM  ENDO FEM HEAD 60MM  ENDO FEM HEAD 63MM  UNIPOLAR COMPONENT, 41MM  UNIPOLAR COMPONENT, 42MM  UNIPOLAR COMPONENT, 43MM  UNIPOLAR COMPONENT, 44MM  UNIPOLAR COMPONENT, 45MM  UNIPOLAR COMPONENT, 46MM  UNIPOLAR COMPONENT, 47MM  UNIPOLAR COMPONENT, 48MM  UNIPOLAR COMPONENT, 49MM  UNIPOLAR COMPONENT, 50MM  UNIPOLAR COMPONENT, 51MM  UNIPOLAR COMPONENT, 52MM  UNIPOLAR COMPONENT, 53MM  UNIPOLAR COMPONENT, 54MM  UNIPOLAR COMPONENT, 55MM  UNIPOLAR COMPONENT, 56MM  UNIPOLAR COMPONENT, 57MM  UNIPOLAR COMPONENT, 58MM  UNIPOLAR COMPONENT, 59MM  UNIPOLAR COMPONENT, 60MM",
      "product_quantity": "972,125 units",
      "reason_for_recall": "Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from  each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.",
      "recall_initiation_date": "20140722",
      "center_classification_date": "20140826",
      "termination_date": "20150915",
      "report_date": "20140903",
      "code_info": "Item Numbers   00781804100, 00781804200, 00781804300, 00781804400, 00781804500, 00781804600, 00781804700, 00781804800, 00781804900, 00781805000, 00781805100, 00781805200, 00781805300, 00781805400, 00781805500, 00781805700, 00781806000, 00781806300, 00902604101, 00902604201, 00902604301, 00902604401, 00902604501, 00902604601, 00902604701, 00902604801, 00902604901, 00902605001, 00902605101, 00902605201, 00902605301, 00902605401, 00902605501, 00902605701, 00902606001, 00902606301, 451000041, 451000042, 451000043, 451000044, 451000045, 451000046, 451000047, 451000048, 451000049, 451000050, 451000051, 451000052, 451000053, 451000054, 451000055, 451000056, 451000057, 451000058, 451000059, 451000060  EXPIRATION DATE PRIOR TO 2024-01"
    }
  ]
}