  "results": [
    {
      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68818",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution.",
      "recall_number": "Z-2546-2014",
      "product_description": "Knee Instrument Fixation Screws:  48MM HEADED SCREW   48MM HEADLESS SCREW  MIS HEADED SCREW, 48MM  MIS HEADLESS SCREW, 48MM",
      "product_quantity": "972,125 units",
      "reason_for_recall": "Zimmer is recalling sterile-packaged implants and instruments packaged in six different package configurations due to packaging design verification test failures. Specifically, multiple test samples from  each of the six configurations failed simulated distribution and shipping testing when the devices either breached inner or outer package trays or seal integrity was lost.",
      "recall_initiation_date": "20140722",
      "center_classification_date": "20140826",
      "termination_date": "20150915",
      "report_date": "20140903",
      "code_info": "Item Numbers 00579104100, 00579104200, 00598304048, 00598304148  Exp Date Prior to 2023-11"
    }
  ]
}