{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Allentown",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88531",
      "recalling_firm": "B. Braun Medical, Inc.",
      "address_1": "901 Marcon Blvd",
      "address_2": "N/A",
      "postal_code": "18109-9512",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution and the country of Canada.",
      "recall_number": "Z-2545-2021",
      "product_description": "Outlook Pump Sets w/3 CARESITE Needle-Free Injection Sites: single-use for use with the Outlook Safety Infusion System and/or Horizon NXT Pump for infusion of parenteral fluids  Material ID: 352413",
      "product_quantity": "585,576 units",
      "reason_for_recall": "Leakage of the pump set within the cassette portion of the pump sets, potential for delays in administration of medications, under-delivery or incomplete dosing and bloodstream infections.",
      "recall_initiation_date": "20210810",
      "center_classification_date": "20210924",
      "termination_date": "20240916",
      "report_date": "20211006",
      "code_info": "Lot Code: 0061755638 0061757839 0061759426 0061763839 0061763893 0061763894 0061766190 0061767014 0061767027 0061768466 0061769094 0061772192 0061772193 0061775455 0061776025 0061776298 0061776299 0061776300 0061776301 0061777154 00VL752555 00VL754131  DI: 04046964182211"
    }
  ]
}