  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Williamston",
      "address_1": "100 Centurion Way",
      "reason_for_recall": "False elevation of carboxyhemoglobin (COHb) results in blood samples collected using BD Vacutainer¿ Plastic Lithium Heparin tubes  with the IL GEM 4000 instrument.",
      "address_2": "",
      "product_quantity": "140 units",
      "code_info": "Lot Number: 2019050790  exp. Date: 10/31/2019",
      "center_classification_date": "20200709",
      "distribution_pattern": "US Nationwide distribution including in the state of IL.",
      "state": "MI",
      "product_description": "Centurion DIFFICULT IV KIT  containing BD's Vacutainer Plastic Lithium Heparin Tubes (BD Vacutainer-BD Codes 367884 & 367960; Centurion code 45177 (VACUTAINER GREEN TOP) and 84815 (VACUTAINER 3ML LT. GREEN).  Centurion Kit Code:  IV8860",
      "report_date": "20200715",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Centurion Medical Products Corporation",
      "recall_number": "Z-2545-2020",
      "initial_firm_notification": "E-Mail",
      "product_type": "Devices",
      "event_id": "85494",
      "termination_date": "20210621",
      "more_code_info": "",
      "recall_initiation_date": "20190601",
      "postal_code": "48895-9086",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}