  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Northridge",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68277",
      "recalling_firm": "Medtronic MiniMed Inc.",
      "address_1": "18000 Devonshire St",
      "address_2": "N/A",
      "postal_code": "91325-1219",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution-Nationwide in US",
      "recall_number": "Z-2544-2014",
      "product_description": "Medtronic MiniMed Paradigm Insulin Infusion Pumps,     Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K, MMT-723, MMT-723K, MMT-554, and MMT-754.    MiniMed 530G: MMT-551, and MMT-751     Paradigm insulin infusion pumps are intended for use to deliver basal and bolus insulin to persons with diabetes.",
      "product_quantity": "559,374 units total (444,374 units in US)",
      "reason_for_recall": "Medtronic is recalling the Medtronic MiniMed Paradigm Insulin Infusion Pumps because they have received reports that users have accidentally programmed the pump to deliver the maximum bolus amount.",
      "recall_initiation_date": "20140313",
      "center_classification_date": "20140822",
      "termination_date": "20150407",
      "report_date": "20140903",
      "code_info": "Paradigm:  MMT-523NAB, MMT-723KNB,  MMT-523NAH, MMT-723KNH,  MMT -523NAL, MMT-723KNL,  MMT-523NAP, MMT-723KNP,  MMT-523NAS, MMT-723KNS,  MMT-723NAB, MMT-723NAH, MMT-723NAL, MMT-723NAP, MMT-723NAS, MMT-523KNB, MMT-523KNH, MMT-523KNL, MMT-523KNP, MMT-523KNS.    MiniMed 530G: MMT-551NAM, MMT-551NAH, MMT-551NAL, MMT-551NAP, MMT-551NAS, MMT-751NAB, MMT-751NAH, MMT-751NAL, MMT-751NAP, MMT-751NAS."
    }
  ]
}