{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Keene",
      "state": "NH",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68739",
      "recalling_firm": "Smiths Medical ASD, Inc.",
      "address_1": "10 Bowman Dr",
      "address_2": "N/A",
      "postal_code": "03431-5043",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA including AK, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV and Internationally to CA, CY, GB, IE, IT, JE, KW, MT, PL, RU, SA, ZA.",
      "recall_number": "Z-2541-2014",
      "product_description": "H4052 Endo Tube Holder, Pedi 3.0mm    Pediatric Endotracheal Intubation Kit",
      "product_quantity": "135,518 (110,246 nationwide)",
      "reason_for_recall": "The tube fastener on the ET Tube Holder may compress the tracheal tube when used which may cause difficulty when locking the ET Tube Holder in place or difficulty when passing a  suction catheter.",
      "recall_initiation_date": "20140707",
      "center_classification_date": "20140828",
      "termination_date": "20210526",
      "report_date": "20140903",
      "code_info": "Lot Numbers: 2602399, 2588625, 2580799, 2576366, 2523679, 2513119, 2492246, 2481855, 2474647, 2465829, 2457754, 2443379, 2438616, 2433451, 2428215, 2412146, 2394061, 2380352, 2376566, 2365454, 2361579, 2357822, 2347642,   2343312, 2343311, 2329584, 2322827, 2309473, 2298755, 2298754, 2286596, 2271359, 2250054, 2244752, 2244751, 2228364, 2211930, 2208196, 2191931, 2188460, 2179231, 2175057, 2154535, 2152054, 2146301, 2142179, 2135984, 2128636, 2120121, 2116100, 2109019, 2090224, 2077180, 2068123, 2063991, 2063990, 11889, 11870, 11859, 11697, 11695, 11694, 11657, 11654, 11406"
    }
  ]
}