  "results": [
    {
      "status": "Terminated",
      "city": "Bloomington",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77305",
      "recalling_firm": "Cook Inc.",
      "address_1": "750 N Daniels Way",
      "address_2": "N/A",
      "postal_code": "47404-9120",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic: FL  IN  LA  MS  NC  TX    Foreign:Argentina, Australia, Brazil, Canada, Czech Republic, Egypt, Estonia, France, Germany, Israel, Italy, Netherlands, Norway, Paraguay, Poland, Portugal, Russian Federation, Spain, Switzerland, Tunisia, Turkey, United Kingdom,    VA/DOD:None",
      "recall_number": "Z-2540-2017",
      "product_description": "Vue Optic Visualization Source, Flexor Vue Deflecting Endoscopic System, Endoscope, Fiber Optic, Supplied sterilized by ethylene oxide gas in peel-open packages",
      "product_quantity": "1328 units",
      "reason_for_recall": "Cook Inc. is initiating a voluntary recall of the Vue Optic Visualization Source and the Flexor Vue Deflecting Endoscopic System because it was determined that the current reprocessing instructions do not provide sufficient detailed information for the cleaning, disinfection, and sterilization of these products.",
      "recall_initiation_date": "20170503",
      "center_classification_date": "20170614",
      "termination_date": "20180126",
      "report_date": "20170621",
      "code_info": "Vue Optic Visualization Source, Catalog Number: FVO-150, GPN: G25343  Flexor Vue Deflecting Endoscopic System  Catalog Number: FV-090075-150, GPN: G50972  Catalog Number: FV-090045-150, GPN: G34306"
    }
  ]
}