  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Oak Creek",
      "state": "WI",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80280",
      "recalling_firm": "Ad-Tech Medical Instrument Corporation",
      "address_1": "400 W Oakview Pkwy",
      "address_2": "N/A",
      "postal_code": "53154-7213",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution in the states of CA, FL, IL, MI, NH, NY, OH, OR, PA, SC, UT, WV.      Worldwide distribution to Canada, Denmark, Germany, India, Israel, Japan, Russia, South Korea, Taiwan, United Kingdom.",
      "recall_number": "Z-2539-2018",
      "product_description": "AD-TECH Medical Instrument Corporation Drill Sleeve Guides DSG-6-3-090-2.4N utilizing raw material (RM0677) from the following lots from vendor 00-0001062., AD0021502, AD0021628.    Product Usage:  The drill sleeve guides are accessories to the Ad-Tech Cranial Drill Bits. The Cranial Drill Bits and accessories are intended to be used for drilling holes in the skull for neurological procedures, such as brain biopsy, brain contacting electrode and electrode accessory device placement. The DSG-6.3-090-2.4N drill sleeve guides are intended to be used with the 2.4mm Cranial Drill Bit diameter.",
      "product_quantity": "65 Total",
      "reason_for_recall": "The diameter of the drill sleeve guide raw material was found to be under tolerance, potentially resulting in the drill bit seizing in the guide during surgery.",
      "recall_initiation_date": "20180502",
      "center_classification_date": "20180725",
      "termination_date": "20190208",
      "report_date": "20180801",
      "code_info": "Catalog Number DSG-6.3-090-2.4N.  Batch Number (Lot Number):  107104 (710220), 114390 (715001), 111423 (716180), 105176 (613121), 113775 (719190), 115847 (712211), 113776 (719190), 115000 (714201), 114724 (711101), 116389 (712121), 117146 (810110)."
    }
  ]
}