  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Completed",
      "city": "Glenwood",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92647",
      "recalling_firm": "Landauer",
      "address_1": "2 Science Rd",
      "address_2": "",
      "postal_code": "60425-1531",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide",
      "recall_number": "Z-2538-2023",
      "product_description": "nanoDot D2DNS, for custom calibrate, OTO, Model Number 03033-OTO; radiation monitoring dosimeter used with the microSTAR readers",
      "product_quantity": "147 units",
      "reason_for_recall": "LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.",
      "recall_initiation_date": "20230712",
      "center_classification_date": "20230912",
      "report_date": "20230920",
      "code_info": "UDI/DI 0860003399903, all batch numbers",
      "more_code_info": ""
    }
  ]
}