  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "San Jose",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85802",
      "recalling_firm": "Vitalconnect Inc.",
      "address_1": "224 Airport Pkwy Ste 300",
      "address_2": "",
      "postal_code": "95110-1022",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "U.S. Nationwide distribution including in the states of NJ, NY, VA, OUS: None",
      "recall_number": "Z-2537-2020",
      "product_description": "VistaSolution, physiological signal transmitter. A software graphical user interface  for use by healthcare professionals to display physiological data collected by the wireless remote monitoring system in healthcare settings. This  is a secondary, adjunct patient monitor and is not intended to replace existing standard-of-care patient monitoring.",
      "product_quantity": "7",
      "reason_for_recall": "A healthcare provider (HCP) noticed a patient had a lowered historic SpO2 reading that the software should have triggered a notification to the care team, but no notification was found. A software defect was encountered causing the lack of notification.  The defect could result in exposing a patient to potential hypoxia for a short window of time. The firm is correcting the software to bring the performance back into compliance with specifications.",
      "recall_initiation_date": "20200521",
      "center_classification_date": "20200708",
      "termination_date": "20210513",
      "report_date": "20200715",
      "code_info": "Catalog number: 6200, Vista Solution version 2.1, UDI:  00850490007139",