  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Covington",
      "state": "GA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80375",
      "recalling_firm": "Bard Medical Division",
      "address_1": "8195 Industrial Blvd NE",
      "address_2": "N/A",
      "postal_code": "30014-1497",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Domestic  Distribution: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL,GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV,   NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.    Worldwide Distribution: Canada, Austria, Bahrain, Belgium, Brazil, Brussels  China, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Gibraltar,   Greece, Guam, Hungary, Iran, Italy, Jordan, Korea, Luxemburg, Malaysia, Netherlands, Norway, Oman, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, UAE, UK, Vietnam.",
      "recall_number": "Z-2536-2018",
      "product_description": "Arctic Sun Temperature Management System, Model 5000.     The Arctic Sun¿ 5000 model is a non-invasive, thermal regulating system, indicated for monitoring and controlling patient temperature within a range of 32¿C to 38.5¿C (89.6¿F to 101.3¿F) in adult and pediatric patients of all ages.",
      "product_quantity": "5608 Units Total",
      "reason_for_recall": "An improper wiring connection on the chiller assembly may result in overheating of the connection and potential failure of certain components in the system.",
      "recall_initiation_date": "20180613",
      "center_classification_date": "20180724",
      "termination_date": "20210216",
      "report_date": "20180801",
      "code_info": "Model #5000    Product codes: 5000-00-00, 5000-00-00L, 5000-00-00E,   5000-01-01, 5000-01-01L, 5000-01-02, 5000-01-03, 5000-01-04, 5000-01-05, 5000-00-06, 5000-01-07, 5000-01-08.  All serial numbers are affected."
    }
  ]
}