{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "99045",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "1800 W Center St",
      "address_2": "N/A",
      "postal_code": "46580-2304",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "N/A",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-2529-2026",
      "product_description": "Off-Axis Comprehensive Shoulder System, Large, Augment, Off-Axis, Reamer Guide  Model/Catalog Number: 110040242",
      "product_quantity": "429 units",
      "reason_for_recall": "Five complaints have been received identifying issues where the plastic guides were catching during reaming, leading to breakage.",
      "recall_initiation_date": "20260520",
      "center_classification_date": "20260623",
      "report_date": "20260701",
      "code_info": "Lot Code: Model No. 110040242 ; UDI-DI (01)00889024681989(17)350601(10)099471 ; Lot Number 099471 Model No. 110040242 ; UDI-DI (01)00889024681989(17)350614(10)099473 ; Lot Number 099473 Model No. 110040242 ; UDI-DI (01)00889024681989(17)350804(10)099474 ; Lot Number 099474 Model No. 110040242 ; UDI-DI (01)00889024681989(17)350806(10)099475 ; Lot Number 099475"
    }
  ]
}