{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Framingham",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74520",
      "recalling_firm": "Ab Sciex",
      "address_1": "500 Old Connecticut Path",
      "address_2": "N/A",
      "postal_code": "01701-4574",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (Nationwide) Internationally to   Canada, Austria, China, Netherlands, Norway, India, Korea, Germany, France, Switzerland, Ireland, Italy, Singapore, Russia, UK, Malaysia,",
      "recall_number": "Z-2529-2016",
      "product_description": "AB SCIEX QTRAP¿ 4500MD LC/MS/MS System  Mass Spectrometer  for In-Vitro Diagnostic Use.  Instrument Part Number (REF): 5031231  3200MD series and 4500MD series are mass spectrometers intended to identify inorganic or organic compounds (e.g., lead, mercury, and drugs) in human specimens by ionizing the compound under investigation and separating the resulting ions by means of an electrical and magnetic field according to their mass. It is intended for in vitro diagnostic purposes. For in vitro diagnostic use.",
      "product_quantity": "28 units",
      "reason_for_recall": "MultiQuant MD software where under certain conditions a user  can be presented with incorrect quantitative results when using the Sum Multiple Ions feature.",
      "recall_initiation_date": "20160616",
      "center_classification_date": "20160812",
      "termination_date": "20170302",
      "report_date": "20160824",
      "code_info": "Software Version:  MultiQuant MD 3.0, MultiQuant MD 3.0.1 and MultiQuant MD 3.0.2 software"
    }
  ]
}