{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80476",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "HI , IL, MI, NC, NY, VA, WI",
      "recall_number": "Z-2525-2018",
      "product_description": "AXIOM Artis with Laird Cooling Unit for SSFD: Material # 7555118.   Artis is an angiography system developed for single and biplane diagnostic imaging and interventional procedures.",
      "product_quantity": "6 units US",
      "reason_for_recall": "Due to a defective sealing, coolant may inflow into the electrical parts of the system cabinet, which could cause a short circuit and cause the system to shut down suddenly during an ongoing procedure",
      "recall_initiation_date": "20180525",
      "center_classification_date": "20180720",
      "termination_date": "20190516",
      "report_date": "20180801",
      "code_info": "Laird Cooling Unit for SSFD: Material # 7555118."
    }
  ]
}