  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80417",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "The devices were distributed to the following US states:  AL, CA, CO, FL, IL, KY, MA, MI, MN, MS, ME, NJ, NY, OH, TN, TX, UT, and VA.",
      "recall_number": "Z-2524-2018",
      "product_description": "ARTIS Pheno, Fluoroscopic X-Ray system, Model Number 10849000",
      "product_quantity": "40 in the United States",
      "reason_for_recall": "During manual LAO/RAO rotation at maximum speed or a combination of LAO/RAO and Cran/Caud rotation at maximum speed, the possibility exists for the safety monitoring within the system control to become activated. As a result, the C-arm stops the rotation and can drop up to 40 mm (1.6 ).",
      "recall_initiation_date": "20180530",
      "center_classification_date": "20180720",
      "termination_date": "20200214",
      "report_date": "20180801",
      "code_info": "164033,   164071,   164034,   164095,   164057,   164054,   164055,   164080,   164072,   164088,   164048,   164094,   164068,   164019,   164013,   164073,   164025,   164026,   164027,   164060,   164083,   164021,   164066,   164075,   164074,   164084,   164086,   164098,   164028,   164053,   164063,   164064,   164024,   164049,   164018,   164022,   164061,   164100,   164097,   164011"
    }
  ]
}