  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Shanghai",
      "state": "",
      "country": "China",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85804",
      "recalling_firm": "Shanghai United Imaging Healthcare Co., Ltd.",
      "address_1": "No. 2258 Chengbei Rd, Jiading Ind.",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Distributed to two customers. Both customers located in Houston, Texas.",
      "recall_number": "Z-2522-2020",
      "product_description": "uCT 760 System, Computed Tomography X-ray System",
      "product_quantity": "5 units",
      "reason_for_recall": "Two  issues were identified with the computed tomography x-ray system including an intermittent issue which may affect image data leading to artifacts; and intermittent scout scanning interruption due to occasional angle signal drift.  This  may result in the need for rescanning of the patient with the patient receiving an additional dose of radiation and possible use of additional contrast media.",
      "recall_initiation_date": "20200508",
      "center_classification_date": "20200706",
      "termination_date": "20220120",
      "report_date": "20200715",
      "code_info": "Model uCT 760; Serial Numbers: 600229, 600230, 600232, 600271 with software version R001.11.0.729741-Re-1100, and serial number 600282 with  software version R001.11.0.785930-Re-1150.",
      "more_code_info": ""
    }
  ]
}