  "results": [
    {
      "status": "Terminated",
      "city": "Hazelwood",
      "state": "MO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "77026",
      "recalling_firm": "Biomerieux Inc",
      "address_1": "595 Anglum Rd",
      "address_2": "N/A",
      "postal_code": "63042-2320",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Nationwide distribution in the U.S, including military and government sites.   The product was also distributed to Algeria, Angola, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia - Herzegovina, Brazil, Brunei, Bulgaria, Burkina Faso, Cambodia, Cameroun, Canada, Chad, Chile, China, Colombia, Congo, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Dutch Antilles, Ecuador, Egypt, El Salvador, Estonia, Finland, France, Gabon, Gambia, Georgia, Germany, Greece, Grenada, Guatemala, Guinea, Haiti, Honduras, Hungary, India, Indonesia, Iraq, Israel, Italy, Ivory Coast, Japan, Jordan, Kazakhstan, Kenya,  Uganda, Kuwait, Laos, Latvia, Lebanon, Lithuania, Malaysia, Mali, Malta, Mexico, Mongolia, Morocco, Mozambique, Myanmar, Netherlands, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Syria, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Venezuela, Vietnam,  and Zimbabwe.",
      "recall_number": "Z-2518-2017",
      "product_description": "VITEK¿ 2 Gram Negative  Susceptibility card (AST-N355), IVD, REF 421351, 20 cards per carton.",
      "product_quantity": "3,419 cartons",
      "reason_for_recall": "The integrity of the product container may be compromised thereby exposing product to moisture which could lead to antibiotic degradation (loss of potency).",
      "recall_initiation_date": "20170420",
      "center_classification_date": "20170613",
      "termination_date": "20200722",
      "report_date": "20170621",
      "code_info": "7950149203 24/Apr/18  7950197203 11/Jun/18  7950243403 27/Jul/18"
    }
  ]
}