  "results": [
    {
      "status": "Terminated",
      "city": "Alpharetta",
      "state": "GA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71966",
      "recalling_firm": "Halyard Health, Inc",
      "address_1": "5405 Windward Pkwy",
      "address_2": "N/A",
      "postal_code": "30004-3894",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide and Puerto Rico and Internationally to Canada",
      "recall_number": "Z-2513-2015",
      "product_description": "KimVent* Closed Suction Systems for Adults, 14 F, Qwik Clip (Product Code 2273)  A tracheobronchial suction catheter is a device that is attached to a port of the airway circuit for use as needed to aspirate liquids or semisolids from a patient's upper airway.",
      "product_quantity": "N/A",
      "reason_for_recall": "The thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the \"LOCKED\" position.",
      "recall_initiation_date": "20150729",
      "center_classification_date": "20150831",
      "termination_date": "20160112",
      "report_date": "20150909",
      "code_info": "Product Code 2273 - M5076T311, M5082T316, M5103T320"
    }
  ]
}