{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98994",
      "recalling_firm": "Medline Industries, LP",
      "address_1": "3 Lakes Dr",
      "address_2": "",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.",
      "recall_number": "Z-2506-2026",
      "product_description": "MEDLINE Medical Procedure Kits labeled as:    KIT URO ROBOT CYSTECTOMY, D AND C PACK, MAD D & C, VPH LITHOTOMY MINOR, TRMC- PERI GYN PACK, RICH MINOR LITHOTOMY, etc. (see recall documents for a full list of products)",
      "product_quantity": "32438 units",
      "reason_for_recall": "During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.",
      "recall_initiation_date": "20260504",
      "center_classification_date": "20260617",
      "report_date": "20260624",
      "code_info": "Medline Kit Number/SKU DYNJ83242B:  UDI/DI 10198459157028 (each), 40198459157029 (case), Lot Number 25HBR321;  Medline Kit Number/SKU DYNJ83242B:  UDI/DI 10198459157028 (each), 40198459157029 (case), Lot Number 25GBD202;  Medline Kit Number/SKU DYNJ83242B:  UDI/DI 10198459157028 (each), 40198459157029 (case), Lot Number 25FBK813;  Medline Kit Number/SKU DYNJ83242B:  UDI/DI 10198459157028 (each), 40198459157029 (case), Lot Number 25EBG207;  Medline Kit Number/SKU DYNJ61059B:  UDI/DI 10195327193454 (each), 40195327193455 (case), Lot Number 23FBK499;  Medline Kit Number/SKU DYNJ61059B:  UDI/DI 10195327193454 (each), 40195327193455 (case), Lot Number 23EBV071;  Medline Kit Number/SKU DYNJ61059B:  UDI/DI 10195327193454 (each), 40195327193455 (case), Lot Number 22KBK089;  Medline Kit Number/SKU DYNJ61059B:  UDI/DI 10195327193454 (each), 40195327193455 (case), Lot Number 22IBV707;  Medline Kit Number/SKU DYNJ61059B:  UDI/DI 10195327193454 (each), 40195327193455 (case), Lot Number 22HBC310;  Medline Kit Number/SKU DYNJ61059C:  UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 26DMG168;  Medline Kit Number/SKU DYNJ61059C:  UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 26CMD637;  Medline Kit Number/SKU DYNJ61059C:  UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 25KMC849;  Medline Kit Number/SKU DYNJ61059C:  UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 25JMA672;  Medline Kit Number/SKU DYNJ61059C:  UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 25DMC189;  Medline Kit Number/SKU DYNJ61059C:  UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 25CMH655;  Medline Kit Number/SKU DYNJ61059C:  UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 25CMC735;  Medline Kit Number/SKU DYNJ61059C:  UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 24KMA168;  Medline Kit Number/SKU DYNJ61059C:  UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 24IMD969;  Medline Kit Number/SKU DYNJ61059C:  UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 24GMF018;  Medline Kit Number/SKU DYNJ61059C:  UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 24EMA387;  Medline Kit Number/SKU DYNJ61059C:  UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 23LBR964;  Medline Kit Number/SKU DYNJ61059C:  UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 23KBR832;  Medline Kit Number/SKU DYNJ61059C:  UDI/DI 10195327451653 (each), 40195327451654 (case), Lot Number 23JBE549;  Medline Kit Number/SKU DYNJT2135D:  UDI/DI 10889942399239 (each), 40889942399230 (case), Lot Number 21LME546;  Medline Kit Number/SKU DYNJT2135D:  UDI/DI 10889942399239 (each), 40889942399230 (case), Lot Number 21KMG048;  Medline Kit Number/SKU DYNJT2135D:  UDI/DI 10889942399239 (each), 40889942399230 (case), Lot Number 21JMH240;  Medline Kit Number/SKU DYNJT2135D:  UDI/DI 10889942399239 (each), 40889942399230 (case), Lot Number 21IMC541;  Medline Kit Number/SKU DYNJT2135D:  UDI/DI 10889942399239 (each), 40889942399230 (case), Lot Number 21GMD231;  Medline Kit Number/SKU DYNJT2135D:  UDI/DI 10889942399239 (each), 40889942399230 (case), Lot Number 21DME148;  Medline Kit Number/SKU DYNJT2135D:  UDI/DI 10889942399239 (each), 40889942399230 (case), Lot Number 21CMC690;  Medline Kit Number/SKU DYNJ51750C:  UDI/DI 10195327583781 (each), 40195327583782 (case), Lot Number 25HBB990;  Medline Kit Number/SKU DYNJ51750C:  UDI/DI 10195327583781 (each), 40195327583782 (case), Lot Number 25FBO533;  Medline Kit Number/SKU DYNJ51750C:  UDI/DI 10195327583781 (each), 40195327583782 (case), Lot Number 25EBA604;  Medline Kit Number/SKU DYNJ51750C:  UDI/DI 10195327583781 (each), 40195327583782 (case), Lot Number 25CBB580;  Medline Kit Number/SKU DYNJ51750C:  UDI/DI 10195327583781 (each), 40195327583782 (case), Lot Number 25ABE166;  Medline Kit Number/SKU DYNJ51750C:  UDI/DI 10195327583781 (each), 40195327583782 (case), Lot Number 24IBR571;  Medline Kit Number/SKU DYNJ51750C:  UDI/DI 10195327583781 (each), 40195327583782 (case), Lot Number 24HBU707;  Medline Kit Number/SKU DYNJ51750C:  UDI/DI 10195327583781 (each), 40195327583782 (case), Lot Number 24HBA096;  Medline Kit Number/SKU DYNJ51750C:  UDI/DI 10195327583781 (each), 40195327583782 (case), Lot Number 24DBC284;  Medline Kit Number/SKU DYNJ51750C:  UDI/DI 10195327583781 (each), 40195327583782 (case), Lot Number 24CBF095;  Medline Kit Number/SKU DYNJ51750C:  UDI/DI 10195327583781 (each), 40195327583782 (case), Lot Number 24BBA310;  Medline Kit Number/SKU DYNJ51750C:  UDI/DI 10195327583781 (each), 40195327583782 (case), Lot Number 24ABM210;  Medline Kit Number/SKU DYNJ58848A:  UDI/DI 10193489750089 (each), 40193489750080 (case), Lot Number 21DBB421;  Medline Kit Number/SKU DYNJ58848A:  UDI/DI 10193489750089 (each), 40193489750080 (case), Lot Number 21BBO116;  Medline Kit Number/SKU DYNJ41314:  UDI/DI 10884389909201 (each), 40884389909202 (case), Lot Number 21GBR980;  Medline Kit Number/SKU DYNJ41314:  UDI/DI 10884389909201 (each), 40884389909202 (case), Lot Number 21GBR981;  Medline Kit Number/SKU DYNJ41314:  UDI/DI 10884389909201 (each), 40884389909202 (case), Lot Number 21DBR243;  Medline Kit Number/SKU DYNJ41314:  UDI/DI 10884389909201 (each), 40884389909202 (case), Lot Number 21BBB699;  Medline Kit Number/SKU DYNJ41314B:  UDI/DI 10198459572173 (each), 40198459572174 (case), Lot Number 26ABW138;  Medline Kit Number/SKU DYNJ41314B:  UDI/DI 10198459572173 (each), 40198459572174 (case), Lot Number 25LBT658;  Medline Kit Number/SKU DYNJ67159:  UDI/DI 10193489421637 (each), 40193489421638 (case), Lot Number 23JBH935;  Medline Kit Number/SKU DYNJ67159:  UDI/DI 10193489421637 (each), 40193489421638 (case), Lot Number 23HBM023;  Medline Kit Number/SKU DYNJ67159:  UDI/DI 10193489421637 (each), 40193489421638 (case), Lot Number 23HBA226;  Medline Kit Number/SKU DYNJ67159:  UDI/DI 10193489421637 (each), 40193489421638 (case), Lot Number 23GBG856;  Medline Kit Number/SKU DYNJ67159:  UDI/DI 10193489421637 (each), 40193489421638 (case), Lot Number 23FBD345;  Medline Kit Number/SKU DYNJ67159:  UDI/DI 10193489421637 (each), 40193489421638 (case), Lot Number 23EBD455;  Medline Kit Number/SKU DYNJ67159:  UDI/DI 10193489421637 (each), 40193489421638 (case), Lot Number 23BBN023;  Medline Kit Number/SKU DYNJ67159:  UDI/DI 10193489421637 (each), 40193489421638 (case), Lot Number 22KBC161;  Medline Kit Number/SKU DYNJ67159:  UDI/DI 10193489421637 (each), 40193489421638 (case), Lot Number 22JBL911;  Medline Kit Number/SKU DYNJ67159:  UDI/DI 10193489421637 (each), 40193489421638 (case), Lot Number 22IBR338;  Medline Kit Number/SKU DYNJ67159:  UDI/DI 10193489421637 (each), 40193489421638 (case), Lot Number 22HBE563;  Medline Kit Number/SKU DYNJ67159:  UDI/DI 10193489421637 (each), 40193489421638 (case), Lot Number 22GBB332;  Medline Kit Number/SKU DYNJ67159:  UDI/DI 10193489421637 (each), 40193489421638 (case), Lot Number 22DBN951;  Medline Kit Number/SKU DYNJ67159:  UDI/DI 10193489421637 (each), 40193489421638 (case), Lot Number 22BBT782;  Medline Kit Number/SKU DYNJ67159:  UDI/DI 10193489421637 (each), 40193489421638 (case), Lot Number 21LBM798;  Medline Kit Number/SKU DYNJ67159:  UDI/DI 10193489421637 (each), 40193489421638 (case), Lot Number 21VBD256;  Medline Kit Number/SKU DYNJ67159:  UDI/DI 10193489421637 (each), 40193489421638 (case), Lot Number 21IBH631;  Medline Kit Number/SKU DYNJ67159:  UDI/DI 10193489421637 (each), 40193489421638 (case), Lot Number 21FBD506;  Medline Kit Number/SKU DYNJ67159B:  UDI/DI 10195327585716 (each), 40195327585717 (case), Lot Number 24ABN500;  Medline Kit Number/SKU DYNJ67159B:  UDI/DI 10195327585716 (each), 40195327585717 (case), Lot Number 24ABG172;  Medline Kit Number/SKU DYNJ67159C:  UDI/DI 10195327669423 (each), 40195327669424 (case), Lot Number 24JBW695;  Medline Kit Number/SKU DYNJ67159C:  UDI/DI 10195327669423 (each), 40195327669424 (case), Lot Number 24JBA856;  Medline Kit Number/SKU DYNJ67159C:  UDI/DI 10195327669423 (each), 40195327669424 (case), Lot Number 24GBY089;  Medline Kit Number/SKU DYNJ67159C:  UDI/DI 10195327669423 (each), 40195327669424 (case), Lot Number 24EBD720;  Medline Kit Number/SKU DYNJ67159C:  UDI/DI 10195327669423 (each), 40195327669424 (case), Lot Number 24DBG026;  Medline Kit Number/SKU DYNJ67159D:  UDI/DI 10198459294143 (each), 40198459294144 (case), Lot Number 26BBH393;  Medline Kit Number/SKU DYNJ67159D:  UDI/DI 10198459294143 (each), 40198459294144 (case), Lot Number 25LBL862;  Medline Kit Number/SKU DYNJ67159D:  UDI/DI 10198459294143 (each), 40198459294144 (case), Lot Number 25JBK760;  Medline Kit Number/SKU DYNJ67159D:  UDI/DI 10198459294143 (each), 40198459294144 (case), Lot Number 25HBK215;  Medline Kit Number/SKU DYNJ67159D:  UDI/DI 10198459294143 (each), 40198459294144 (case), Lot Number 25FBJ199;  Medline Kit Number/SKU DYNJ67159D:  UDI/DI 10198459294143 (each), 40198459294144 (case), Lot Number 25DBB533;  Medline Kit Number/SKU DYNJT6852:  UDI/DI 10198459583513 (each), 40198459583514 (case), Lot Number 26BBT868;  Medline Kit Number/SKU DYNJT6852:  UDI/DI 10198459583513 (each), 40198459583514 (case), Lot Number 25KBL900;  Medline Kit Number/SKU DYNJ82901B:  UDI/DI 10198459437335 (each), 40198459437336 (case), Lot Number 25KMI697;  Medline Kit Number/SKU DYNJ82901B:  UDI/DI 10198459437335 (each), 40198459437336 (case), Lot Number 25KMH448;  Medline Kit Number/SKU DYNJ34432J:  UDI/DI 10193489404456 (each), 40193489404457 (case), Lot Number 23CBH486;  Medline Kit Number/SKU DYNJ34432J:  UDI/DI 10193489404456 (each), 40193489404457 (case), Lot Number 23ABL372;  Medline Kit Number/SKU DYNJ34432J:  UDI/DI 10193489404456 (each), 40193489404457 (case), Lot Number 22JBD064;  Medline Kit Number/SKU DYNJ34432J:  UDI/DI 10193489404456 (each), 40193489404457 (case), Lot Number 22GBZ380;  Medline Kit Number/SKU DYNJ34432J:  UDI/DI 10193489404456 (each), 40193489404457 (case), Lot Number 22FBS118;  Medline Kit Number/SKU DYNJ34432J:  UDI/DI 10193489404456 (each), 40193489404457 (case), Lot Number 22CBX023;  Medline Kit Number/SKU DYNJ34432J:  UDI/DI 10193489404456 (each), 40193489404457 (case), Lot Number 22ABU279;  Medline Kit Number/SKU DYNJ34432J:  UDI/DI 10193489404456 (each), 40193489404457 (case), Lot Number 22ABG172;  Medline Kit Number/SKU DYNJ34432J:  UDI/DI 10193489404456 (each), 40193489404457 (case), Lot Number 21LBS917;  Medline Kit Number/SKU DYNJ34432J:  UDI/DI 10193489404456 (each), 40193489404457 (case), Lot Number 21JBW610;  Medline Kit Number/SKU DYNJ34432J:  UDI/DI 10193489404456 (each), 40193489404457 (case), Lot Number 21JBF070;  Medline Kit Number/SKU DYNJ34432J:  UDI/DI 10193489404456 (each), 40193489404457 (case), Lot Number 21FBJ399;  Medline Kit Number/SKU DYNJ34432J:  UDI/DI 10193489404456 (each), 40193489404457 (case), Lot Number 21EBP668;  Medline Kit Number/SKU DYNJ34432J:  UDI/DI 10193489404456 (each), 40193489404457 (case), Lot Number 21BBJ995;  Medline Kit Number/SKU DYNJ34432J:  UDI/DI 10193489404456 (each), 40193489404457 (case), Lot Number 21ABA353;  Medline Kit Number/SKU DYNJ34432K:  UDI/DI 10195327384401 (each), 40195327384402 (case), Lot Number 23HBB977;  Medline Kit Number/SKU DYNJ34432K:  UDI/DI 10195327384401 (each), 40195327384402 (case), Lot Number 23GBC923;  Medline Kit Number/SKU DYNJ34432K:  UDI/DI 10195327384401 (each), 40195327384402 (case), Lot Number 23EBJ509;  Medline Kit Number/SKU DYNJ34432L:  UDI/DI 10195327511654 (each), 40195327511655 (case), Lot Number 24JBM566;  Medline Kit Number/SKU DYNJ34432L:  UDI/DI 10195327511654 (each), 40195327511655 (case), Lot Number 24IBD224;  Medline Kit Number/SKU DYNJ34432L:  UDI/DI 10195327511654 (each), 40195327511655 (case), Lot Number 24HBA058;  Medline Kit Number/SKU DYNJ34432L:  UDI/DI 10195327511654 (each), 40195327511655 (case), Lot Number 24GBB789;  Medline Kit Number/SKU DYNJ34432L:  UDI/DI 10195327511654 (each), 40195327511655 (case), Lot Number 24EBJ806;  Medline Kit Number/SKU DYNJ34432L:  UDI/DI 10195327511654 (each), 40195327511655 (case), Lot Number 24DBT403;  Medline Kit Number/SKU DYNJ34432L:  UDI/DI 10195327511654 (each), 40195327511655 (case), Lot Number 24DBF904;  Medline Kit Number/SKU DYNJ34432L:  UDI/DI 10195327511654 (each), 40195327511655 (case), Lot Number 24BBQ443;  Medline Kit Number/SKU DYNJ34432L:  UDI/DI 10195327511654 (each), 40195327511655 (case), Lot Number 23LBQ471;  Medline Kit Number/SKU DYNJ34432L:  UDI/DI 10195327511654 (each), 40195327511655 (case), Lot Number 23KBU644;  Medline Kit Number/SKU DYNJ906947:  UDI/DI 10193489299731 (each), 40193489299732 (case), Lot Number 22OBC517;  Medline Kit Number/SKU DYNJ906947:  UDI/DI 10193489299731 (each), 40193489299732 (case), Lot Number 22CBW739;  Medline Kit Number/SKU DYNJ906947:  UDI/DI 10193489299731 (each), 40193489299732 (case), Lot Number 22ABC151;  Medline Kit Number/SKU DYNJ906947A:  UDI/DI 10195327128296 (each), 40195327128297 (case), Lot Number 23DBE955;  Medline Kit Number/SKU DYNJ906947A:  UDI/DI 10195327128296 (each), 40195327128297 (case), Lot Number 23ABP498;  Medline Kit Number/SKU DYNJ906947A:  UDI/DI 10195327128296 (each), 40195327128297 (case), Lot Number 22HBH329;  Medline Kit Number/SKU DYNJ906947A:  UDI/DI 10195327128296 (each), 40195327128297 (case), Lot Number 22FBM732;  Medline Kit Number/SKU DYNJ906947B:  UDI/DI 10195327380984 (each), 40195327380985 (case), Lot Number 23LBK760;  Medline Kit Number/SKU DYNJ906947B:  UDI/DI 10195327380984 (each), 40195327380985 (case), Lot Number 23LBF829;  Medline Kit Number/SKU DYNJ906947B:  UDI/DI 10195327380984 (each), 40195327380985 (case), Lot Number 23KBW900;  Medline Kit Number/SKU DYNJ906947B:  UDI/DI 10195327380984 (each), 40195327380985 (case), Lot Number 23JBI914;  Medline Kit Number/SKU DYNJ906947B:  UDI/DI 10195327380984 (each), 40195327380985 (case), Lot Number 23HBG246;  Medline Kit Number/SKU DYNJ906947B:  UDI/DI 10195327380984 (each), 40195327380985 (case), Lot Number 23EBU035;  Medline Kit Number/SKU DYNJ906947C:  UDI/DI 10195327569006 (each), 40195327569007 (case), Lot Number 24JBK865;  Medline Kit Number/SKU DYNJ906947C:  UDI/DI 10195327569006 (each), 40195327569007 (case), Lot Number 24JBA344;  Medline Kit Number/SKU DYNJ906947C:  UDI/DI 10195327569006 (each), 40195327569007 (case), Lot Number 24IBN106;  Medline Kit Number/SKU DYNJ906947C:  UDI/DI 10195327569006 (each), 40195327569007 (case), Lot Number 24GBR916;  Medline Kit Number/SKU DYNJ906947C:  UDI/DI 10195327569006 (each), 40195327569007 (case), Lot Number 24GBK603;  Medline Kit Number/SKU DYNJ906947C:  UDI/DI 10195327569006 (each), 40195327569007 (case), Lot Number 24FBI305;  Medline Kit Number/SKU DYNJ906947C:  UDI/DI 10195327569006 (each), 40195327569007 (case), Lot Number 24EBD043;  Medline Kit Number/SKU DYNJ906947C:  UDI/DI 10195327569006 (each), 40195327569007 (case), Lot Number 24DBD823;  Medline Kit Number/SKU DYNJ906947C:  UDI/DI 10195327569006 (each), 40195327569007 (case), Lot Number 24CBB548;  Medline Kit Number/SKU DYNJ906947C:  UDI/DI 10195327569006 (each), 40195327569007 (case), Lot Number 24BBL769;  Medline Kit Number/SKU DYNJ906947D:  UDI/DI 10198459156892 (each), 40198459156893 (case), Lot Number 25JBU960;  Medline Kit Number/SKU DYNJ906947D:  UDI/DI 10198459156892 (each), 40198459156893 (case), Lot Number 25JBD859;  Medline Kit Number/SKU DYNJ906947D:  UDI/DI 10198459156892 (each), 40198459156893 (case), Lot Number 25GBW571;  Medline Kit Number/SKU DYNJ906947D:  UDI/DI 10198459156892 (each), 40198459156893 (case), Lot Number 25GBM727;  Medline Kit Number/SKU DYNJ906947D:  UDI/DI 10198459156892 (each), 40198459156893 (case), Lot Number 25EBR822;  Medline Kit Number/SKU DYNJ906947D:  UDI/DI 10198459156892 (each), 40198459156893 (case), Lot Number 25CBL210;  Medline Kit Number/SKU DYNJ906947D:  UDI/DI 10198459156892 (each), 40198459156893 (case), Lot Number 25ABN213;  Medline Kit Number/SKU DYNJ906947D:  UDI/DI 10198459156892 (each), 40198459156893 (case), Lot Number 25ABD653;  Medline Kit Number/SKU DYNJ906947D:  UDI/DI 10198459156892 (each), 40198459156893 (case), Lot Number 24LBD925;  Medline Kit Number/SKU DYNJ906947D:  UDI/DI 10198459156892 (each), 40198459156893 (case), Lot Number 24KBU887;  Medline Kit Number/SKU DYNJ912132:  UDI/DI 10198459639593 (each), 40198459639594 (case), Lot Number 26CBB112;  Medline Kit Number/SKU DYNJ912132:  UDI/DI 10198459639593 (each), 40198459639594 (case), Lot Number 26ABO288;  Medline Kit Number/SKU CDS983579J:  UDI/DI 10195327295387 (each), 40195327295388 (case), Lot Number 25LMC038;  Medline Kit Number/SKU CDS983579J:  UDI/DI 10195327295387 (each), 40195327295388 (case), Lot Number 25LMA314;  Medline Kit Number/SKU CDS983579J:  UDI/DI 10195327295387 (each), 40195327295388 (case), Lot Number 25IMI321;  Medline Kit Number/SKU CDS983579J:  UDI/DI 10195327295387 (each), 40195327295388 (case), Lot Number 25HMA488;  Medline Kit Number/SKU CDS983579J:  UDI/DI 10195327295387 (each), 40195327295388 (case), Lot Number 25FMC285;  Medline Kit Number/SKU CDS983579J:  UDI/DI 10195327295387 (each), 40195327295388 (case), Lot Number 25DMG929;  Medline Kit Number/SKU CDS983579J:  UDI/DI 10195327295387 (each), 40195327295388 (case), Lot Number 25CMG346;  Medline Kit Number/SKU CDS983579J:  UDI/DI 10195327295387 (each), 40195327295388 (case), Lot Number 25BMB083;  Medline Kit Number/SKU CDS983579J:  UDI/DI 10195327295387 (each), 40195327295388 (case), Lot Number 24LME935;  Medline Kit Number/SKU CDS983579J:  UDI/DI 10195327295387 (each), 40195327295388 (case), Lot Number 24JMB664;  Medline Kit Number/SKU CDS983579J:  UDI/DI 10195327295387 (each), 40195327295388 (case), Lot Number 24HMD395;  Medline Kit Number/SKU CDS983579J:  UDI/DI 10195327295387 (each), 40195327295388 (case), Lot Number 24EME998;  Medline Kit Number/SKU CDS983579J:  UDI/DI 10195327295387 (each), 40195327295388 (case), Lot Number 24BMI128;  Medline Kit Number/SKU CDS983579J:  UDI/DI 10195327295387 (each), 40195327295388 (case), Lot Number 23LMG888;  Medline Kit Number/SKU CDS983579J:  UDI/DI 10195327295387 (each), 40195327295388 (case), Lot Number 23GMI303;  Medline Kit Number/SKU CDS983579J:  UDI/DI 10195327295387 (each), 40195327295388 (case), Lot Number 23FMD541;  Medline Kit Number/SKU CDS983579J:  UDI/DI 10195327295387 (each), 40195327295388 (case), Lot Number 23EMD600;  Medline Kit Number/SKU CDS983579J:  UDI/DI 10195327295387 (each), 40195327295388 (case), Lot Number 23DMH786;  Medline Kit Number/SKU CDS983911J:  UDI/DI 10193489402667 (each), 40193489402668 (case), Lot Number 23JBC180;  Medline Kit Number/SKU CDS983911J:  UDI/DI 10193489402667 (each), 40193489402668 (case), Lot Number 23IBM022;  Medline Kit Number/SKU CDS983911J:  UDI/DI 10193489402667 (each), 40193489402668 (case), Lot Number 23CBT642;  Medline Kit Number/SKU CDS983911J:  UDI/DI 10193489402667 (each), 40193489402668 (case), Lot Number 23ABU424;  Medline Kit Number/SKU CDS983911J:  UDI/DI 10193489402667 (each), 40193489402668 (case), Lot Number 23ABE617;  Medline Kit Number/SKU CDS983911J:  UDI/DI 10193489402667 (each), 40193489402668 (case), Lot Number 22LBM285;  Medline Kit Number/SKU CDS983911J:  UDI/DI 10193489402667 (each), 40193489402668 (case), Lot Number 22LBL299;  Medline Kit Number/SKU CDS983911J:  UDI/DI 10193489402667 (each), 40193489402668 (case), Lot Number 22LBC398;  Medline Kit Number/SKU CDS983911J:  UDI/DI 10193489402667 (each), 40193489402668 (case), Lot Number 22KBI115;  Medline Kit Number/SKU CDS983911J:  UDI/DI 10193489402667 (each), 40193489402668 (case), Lot Number 22JBM360;  Medline Kit Number/SKU CDS983911J:  UDI/DI 10193489402667 (each), 40193489402668 (case), Lot Number 22HBW123;  Medline Kit Number/SKU CDS983911J:  UDI/DI 10193489402667 (each), 40193489402668 (case), Lot Number 22EBU168;  Medline Kit Number/SKU CDS983911J:  UDI/DI 10193489402667 (each), 40193489402668 (case), Lot Number 22NBF154;  Medline Kit Number/SKU CDS983911J:  UDI/DI 10193489402667 (each), 40193489402668 (case), Lot Number 22NBD200;  Medline Kit Number/SKU CDS983911J:  UDI/DI 10193489402667 (each), 40193489402668 (case), Lot Number 21KBA484;  Medline Kit Number/SKU CDS983911J:  UDI/DI 10193489402667 (each), 40193489402668 (case), Lot Number 21IBJ325;  Medline Kit Number/SKU CDS983911J:  UDI/DI 10193489402667 (each), 40193489402668 (case), Lot Number 21IBC193;  Medline Kit Number/SKU CDS983911J:  UDI/DI 10193489402667 (each), 40193489402668 (case), Lot Number 21FBF895;  Medline Kit Number/SKU CDS983911J:  UDI/DI 10193489402667 (each), 40193489402668 (case), Lot Number 21FBF898;  Medline Kit Number/SKU CDS983911L:  UDI/DI 10198459238208 (each), 40198459238209 (case), Lot Number 25JBE929;  Medline Kit Number/SKU CDS983911L:  UDI/DI 10198459238208 (each), 40198459238209 (case), Lot Number 25JBF476;  Medline Kit Number/SKU CDS983911L:  UDI/DI 10198459238208 (each), 40198459238209 (case), Lot Number 25GBW710;  Medline Kit Number/SKU CDS983911L:  UDI/DI 10198459238208 (each), 40198459238209 (case), Lot Number 25EBN090;  Medline Kit Number/SKU CDS983911L:  UDI/DI 10198459238208 (each), 40198459238209 (case), Lot Number 25CBQ682;  Medline Kit Number/SKU CDS983911L:  UDI/DI 10198459238208 (each), 40198459238209 (case), Lot Number 25BBK668;  Medline Kit Number/SKU DYNJ65924D:  UDI/DI 10198459177200 (each), 40198459177201 (case), Lot Number 25DBH769;  Medline Kit Number/SKU DYNJ65924D:  UDI/DI 10198459177200 (each), 40198459177201 (case), Lot Number 24KBL400;  Medline Kit Number/SKU DYNJ65924F:  UDI/DI 10198459475993 (each), 40198459475994 (case), Lot Number 26DBE515;  Medline Kit Number/SKU DYNJ65924F:  UDI/DI 10198459475993 (each), 40198459475994 (case), Lot Number 26CBJ573;  Medline Kit Number/SKU DYNJ900273D:  UDI/DI 10888277759534 (each), 40888277759535 (case), Lot Number 22LBC989;  Medline Kit Number/SKU DYNJ900273D:  UDI/DI 10888277759534 (each), 40888277759535 (case), Lot Number 22GBJ158;  Medline Kit Number/SKU DYNJ900273D:  UDI/DI 10888277759534 (each), 40888277759535 (case), Lot Number 22FBG763;  Medline Kit Number/SKU DYNJ900273D:  UDI/DI 10888277759534 (each), 40888277759535 (case), Lot Number 22EBC244;  Medline Kit Number/SKU DYNJ900273D:  UDI/DI 10888277759534 (each), 40888277759535 (case), Lot Number 22EBA136;  Medline Kit Number/SKU DYNJ900273D:  UDI/DI 10888277759534 (each), 40888277759535 (case), Lot Number 22ABS158;  Medline Kit Number/SKU DYNJ900273D:  UDI/DI 10888277759534 (each), 40888277759535 (case), Lot Number 22ABC145;  Medline Kit Number/SKU DYNJ900273D:  UDI/DI 10888277759534 (each), 40888277759535 (case), Lot Number 21IBV683;  Medline Kit Number/SKU DYNJ900273D:  UDI/DI 10888277759534 (each), 40888277759535 (case), Lot Number 21GBM229;  Medline Kit Number/SKU DYNJ900273D:  UDI/DI 10888277759534 (each), 40888277759535 (case), Lot Number 21EBO675;  Medline Kit Number/SKU DYNJ900273D:  UDI/DI 10888277759534 (each), 40888277759535 (case), Lot Number 21ABD425;  Medline Kit Number/SKU DYNJ900273F:  UDI/DI 10195327348205 (each), 40195327348206 (case), Lot Number 25JBK123;  Medline Kit Number/SKU DYNJ900273F:  UDI/DI 10195327348205 (each), 40195327348206 (case), Lot Number 25GBV261;  Medline Kit Number/SKU DYNJ900273F:  UDI/DI 10195327348205 (each), 40195327348206 (case), Lot Number 25GBC113;  Medline Kit Number/SKU DYNJ900273F:  UDI/DI 10195327348205 (each), 40195327348206 (case), Lot Number 25BBN516;  Medline Kit Number/SKU DYNJ900273F:  UDI/DI 10195327348205 (each), 40195327348206 (case), Lot Number 25ABC632;  Medline Kit Number/SKU DYNJ900273F:  UDI/DI 10195327348205 (each), 40195327348206 (case), Lot Number 24KBD011;  Medline Kit Number/SKU DYNJ900273F:  UDI/DI 10195327348205 (each), 40195327348206 (case), Lot Number 24IBT024;  Medline Kit Number/SKU DYNJ900273F:  UDI/DI 10195327348205 (each), 40195327348206 (case), Lot Number 24IBE080;  Medline Kit Number/SKU DYNJ900273F:  UDI/DI 10195327348205 (each), 40195327348206 (case), Lot Number 24HBP714;  Medline Kit Number/SKU DYNJ900273F:  UDI/DI 10195327348205 (each), 40195327348206 (case), Lot Number 24FBL611;  Medline Kit Number/SKU DYNJ900273F:  UDI/DI 10195327348205 (each), 40195327348206 (case), Lot Number 24BBV258;  Medline Kit Number/SKU DYNJ900273F:  UDI/DI 10195327348205 (each), 40195327348206 (case), Lot Number 24BBO683;  Medline Kit Number/SKU DYNJ900273F:  UDI/DI 10195327348205 (each), 40195327348206 (case), Lot Number 23LBT922;  Medline Kit Number/SKU DYNJ900273F:  UDI/DI 10195327348205 (each), 40195327348206 (case), Lot Number 23LBM669;  Medline Kit Number/SKU DYNJ900273F:  UDI/DI 10195327348205 (each), 40195327348206 (case), Lot Number 23IBC933;  Medline Kit Number/SKU DYNJ900273F:  UDI/DI 10195327348205 (each), 40195327348206 (case), Lot Number 23IBD125;  Medline Kit Number/SKU DYNJ900273F:  UDI/DI 10195327348205 (each), 40195327348206 (case), Lot Number 23HBC371;  Medline Kit Number/SKU DYNJ900273F:  UDI/DI 10195327348205 (each), 40195327348206 (case), Lot Number 23GBB272;  Medline Kit Number/SKU DYNJ900273F:  UDI/DI 10195327348205 (each), 40195327348206 (case), Lot Number 23EBN120;  Medline Kit Number/SKU DYNJ900273F:  UDI/DI 10195327348205 (each), 40195327348206 (case), Lot Number 23CBL737;  Medline Kit Number/SKU CDS984596F:  UDI/DI 10193489484649 (each), 40193489484640 (case), Lot Number 23KMI895;  Medline Kit Number/SKU CDS984596F:  UDI/DI 10193489484649 (each), 40193489484640 (case), Lot Number 23KMD181;  Medline Kit Number/SKU CDS984596F:  UDI/DI 10193489484649 (each), 40193489484640 (case), Lot Number 23GMF223;  Medline Kit Number/SKU CDS984596F:  UDI/DI 10193489484649 (each), 40193489484640 (case), Lot Number 23EMF755;  Medline Kit Number/SKU CDS984596F:  UDI/DI 10193489484649 (each), 40193489484640 (case), Lot Number 23CMF122;  Medline Kit Number/SKU CDS984596F:  UDI/DI 10193489484649 (each), 40193489484640 (case), Lot Number 23CMB203;  Medline Kit Number/SKU CDS984596F:  UDI/DI 10193489484649 (each), 40193489484640 (case), Lot Number 23BMH561;  Medline Kit Number/SKU CDS984596F:  UDI/DI 10193489484649 (each), 40193489484640 (case), Lot Number 23BMF867;  Medline Kit Number/SKU CDS984596F:  UDI/DI 10193489484649 (each), 40193489484640 (case), Lot Number 23BMB770;  Medline Kit Number/SKU CDS984596F:  UDI/DI 10193489484649 (each), 40193489484640 (case), Lot Number 22LMC186;  Medline Kit Number/SKU CDS984596F:  UDI/DI 10193489484649 (each), 40193489484640 (case), Lot Number 22JMH127;  Medline Kit Number/SKU CDS984596F:  UDI/DI 10193489484649 (each), 40193489484640 (case), Lot Number 22HMF508;  Medline Kit Number/SKU CDS984596F:  UDI/DI 10193489484649 (each), 40193489484640 (case), Lot Number 22GME420;  Medline Kit Number/SKU CDS984596F:  UDI/DI 10193489484649 (each), 40193489484640 (case), Lot Number 22EMF689;  Medline Kit Number/SKU CDS984596F:  UDI/DI 10193489484649 (each), 40193489484640 (case), Lot Number 22CMG452;  Medline Kit Number/SKU CDS984596F:  UDI/DI 10193489484649 (each), 40193489484640 (case), Lot Number 22AMF650;  Medline Kit Number/SKU CDS984596F:  UDI/DI 10193489484649 (each), 40193489484640 (case), Lot Number 21KMH512;  Medline Kit Number/SKU CDS984596F:  UDI/DI 10193489484649 (each), 40193489484640 (case), Lot Number 21JMB022;  Medline Kit Number/SKU CDS984596F:  UDI/DI 10193489484649 (each), 40193489484640 (case), Lot Number 21GMD813;  Medline Kit Number/SKU CDS984596F:  UDI/DI 10193489484649 (each), 40193489484640 (case), Lot Number 21FMF915;  Medline Kit Number/SKU CDS984596F:  UDI/DI 10193489484649 (each), 40193489484640 (case), Lot Number 21FMF059;  Medline Kit Number/SKU CDS984596F:  UDI/DI 10193489484649 (each), 40193489484640 (case), Lot Number 21DKA276;  Medline Kit Number/SKU CDS984596F:  UDI/DI 10193489484649 (each), 40193489484640 (case), Lot Number 21CKA260;  Medline Kit Number/SKU CDS984596F:  UDI/DI 10193489484649 (each), 40193489484640 (case), Lot Number 21BKA125;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 26DBC220;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 26BBH093;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 26ABQ164;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 25KBQ260;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 25HBS223;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 25CBU691;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 25BBO684;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 25BBD033;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 25ABI075;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 24LBD957;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 24JBN479;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 24IBI110;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 24EBN280;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 24CBG271;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 24BBB062;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 24BBB068;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 23LBI203;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 23HBD159;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 23ABP326;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 23ABP470;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 22KBM274;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 22IBA124;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 22HBF411;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 22FBO642;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 22DBG492;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 22BBV242;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 22BBM416;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 21JBU176;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 21IBD412;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 21HBO756;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 21GBH376;  Medline Kit Number/SKU CDS985201A:  UDI/DI 10889942942312 (each), 40889942942313 (case), Lot Number 21CBT396;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 25HBI452;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 25FBT604;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 25DBR483;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 25CBK782;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 25ABP633;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 24KBU153;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 24KBF638;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 24IBD754;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 24HBA458;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 24GBH767;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 24FBN482;  Medline Kit Number/SKU CDS982891F:  UDI/",
      "more_code_info": "DI 10889942216956 (each), 40889942216957 (case), Lot Number 24CBT645;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 24BBD960;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 23LBI900;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 23LBJ483;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 23JBN017;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 23IBO475;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 23HBJ305;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 23FBS905;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 23FBC159;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 23DBD766;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 23CBD784;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 23ABS326;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 23ABS332;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 22KBO723;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 22IBJ729;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 22IBD815;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 22HBS217;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 22GBU420;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 22FBM811;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 22EBH358;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 22OBK783;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 22BBL319;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 22BBL320;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 21VBB534;  Medline Kit Number/SKU CDS982891F:  UDI/DI 10889942216956 (each), 40889942216957 (case), Lot Number 21IBL561;  Medline Kit Number/SKU DYNJ85870:  UDI/DI 10195327561314 (each), 40195327561315 (case), Lot Number 25EBA326;  Medline Kit Number/SKU DYNJ85870:  UDI/DI 10195327561314 (each), 40195327561315 (case), Lot Number 25ABI116;  Medline Kit Number/SKU DYNJ85870:  UDI/DI 10195327561314 (each), 40195327561315 (case), Lot Number 24LBC287;  Medline Kit Number/SKU DYNJ85870:  UDI/DI 10195327561314 (each), 40195327561315 (case), Lot Number 24KBU492;  Medline Kit Number/SKU DYNJ85870:  UDI/DI 10195327561314 (each), 40195327561315 (case), Lot Number 24JBE491;  Medline Kit Number/SKU DYNJ85870:  UDI/DI 10195327561314 (each), 40195327561315 (case), Lot Number 24HBL345;  Medline Kit Number/SKU DYNJ85870:  UDI/DI 10195327561314 (each), 40195327561315 (case), Lot Number 24FBT623;  Medline Kit Number/SKU DYNJ85870:  UDI/DI 10195327561314 (each), 40195327561315 (case), Lot Number 24FBB337;  Medline Kit Number/SKU DYNJ85870:  UDI/DI 10195327561314 (each), 40195327561315 (case), Lot Number 24CBK547;  Medline Kit Number/SKU DYNJ85870:  UDI/DI 10195327561314 (each), 40195327561315 (case), Lot Number 24ABK669;  Medline Kit Number/SKU DYNJ85870:  UDI/DI 10195327561314 (each), 40195327561315 (case), Lot Number 23LBH611;  Medline Kit Number/SKU DYNJ33603D:  UDI/DI 10888277925298 (each), 40888277925299 (case), Lot Number 21FMC493;  Medline Kit Number/SKU DYNJ40506B:  UDI/DI 10889942626540 (each), 40889942626541 (case), Lot Number 21KMF678;  Medline Kit Number/SKU DYNJ40506B:  UDI/DI 10889942626540 (each), 40889942626541 (case), Lot Number 21JMG968;  Medline Kit Number/SKU DYNJ40506B:  UDI/DI 10889942626540 (each), 40889942626541 (case), Lot Number 21IMB622;  Medline Kit Number/SKU DYNJ40506B:  UDI/DI 10889942626540 (each), 40889942626541 (case), Lot Number 21HMB675;  Medline Kit Number/SKU DYNJ40506B:  UDI/DI 10889942626540 (each), 40889942626541 (case), Lot Number 21EMD611;  Medline Kit Number/SKU DYNJ40506B:  UDI/DI 10889942626540 (each), 40889942626541 (case), Lot Number 21DMA524;  Medline Kit Number/SKU DYNJ40506B:  UDI/DI 10889942626540 (each), 40889942626541 (case), Lot Number 21BMA422;  Medline Kit Number/SKU DYNJ40506B:  UDI/DI 10889942626540 (each), 40889942626541 (case), Lot Number 21AME493;  Medline Kit Number/SKU DYNJ43571C:  UDI/DI 10198459091490 (each), 40198459091491 (case), Lot Number 25HBC803;  Medline Kit Number/SKU DYNJ43571C:  UDI/DI 10198459091490 (each), 40198459091491 (case), Lot Number 25CBC685;  Medline Kit Number/SKU DYNJ43571D:  UDI/DI 10198459522222 (each), 40198459522223 (case), Lot Number 26BBL904;  Medline Kit Number/SKU DYNJ43571D:  UDI/DI 10198459522222 (each), 40198459522223 (case), Lot Number 26ABA655;  Medline Kit Number/SKU DYNJ43571D:  UDI/DI 10198459522222 (each), 40198459522223 (case), Lot Number 25LBC485;  Medline Kit Number/SKU DYNJT2182I:  UDI/DI 10193489729221 (each), 40193489729222 (case), Lot Number 22GME833;  Medline Kit Number/SKU DYNJT2182I:  UDI/DI 10193489729221 (each), 40193489729222 (case), Lot Number 22EMF340;  Medline Kit Number/SKU DYNJT2182I:  UDI/DI 10193489729221 (each), 40193489729222 (case), Lot Number 22DMA340;  Medline Kit Number/SKU DYNJT2182I:  UDI/DI 10193489729221 (each), 40193489729222 (case), Lot Number 21LMD408;  Medline Kit Number/SKU DYNJT2182I:  UDI/DI 10193489729221 (each), 40193489729222 (case), Lot Number 21JME295;  Medline Kit Number/SKU DYNJT2182I:  UDI/DI 10193489729221 (each), 40193489729222 (case), Lot Number 21JMB250;  Medline Kit Number/SKU DYNJT2182I:  UDI/DI 10193489729221 (each), 40193489729222 (case), Lot Number 21JMB254;  Medline Kit Number/SKU DYNJT2182I:  UDI/DI 10193489729221 (each), 40193489729222 (case), Lot Number 21GMG146;  Medline Kit Number/SKU DYNJT2182I:  UDI/DI 10193489729221 (each), 40193489729222 (case), Lot Number 21EME681;  Medline Kit Number/SKU DYNJT2182I:  UDI/DI 10193489729221 (each), 40193489729222 (case), Lot Number 21EMB537;  Medline Kit Number/SKU DYNJT2182I:  UDI/DI 10193489729221 (each), 40193489729222 (case), Lot Number 20LMF612;  Medline Kit Number/SKU DYNJT2182J:  UDI/DI 10195327263553 (each), 40195327263554 (case), Lot Number 23EMH160;  Medline Kit Number/SKU DYNJT2182J:  UDI/DI 10195327263553 (each), 40195327263554 (case), Lot Number 23DMI189;  Medline Kit Number/SKU DYNJT2182J:  UDI/DI 10195327263553 (each), 40195327263554 (case), Lot Number 23DMC554;  Medline Kit Number/SKU DYNJT2182J:  UDI/DI 10195327263553 (each), 40195327263554 (case), Lot Number 22KMF620;  Medline Kit Number/SKU DYNJT2182J:  UDI/DI 10195327263553 (each), 40195327263554 (case), Lot Number 22JMC529;  Medline Kit Number/SKU DYNJT2182K:  UDI/DI 10195327434502 (each), 40195327434503 (case), Lot Number 26BMF250;  Medline Kit Number/SKU DYNJT2182K:  UDI/DI 10195327434502 (each), 40195327434503 (case), Lot Number 25JMD070;  Medline Kit Number/SKU DYNJT2182K:  UDI/DI 10195327434502 (each), 40195327434503 (case), Lot Number 25GMJ864;  Medline Kit Number/SKU DYNJT2182K:  UDI/DI 10195327434502 (each), 40195327434503 (case), Lot Number 25FMA215;  Medline Kit Number/SKU DYNJT2182K:  UDI/DI 10195327434502 (each), 40195327434503 (case), Lot Number 25DMD128;  Medline Kit Number/SKU DYNJT2182K:  UDI/DI 10195327434502 (each), 40195327434503 (case), Lot Number 24LMG936;  Medline Kit Number/SKU DYNJT2182K:  UDI/DI 10195327434502 (each), 40195327434503 (case), Lot Number 24KMG823;  Medline Kit Number/SKU DYNJT2182K:  UDI/DI 10195327434502 (each), 40195327434503 (case), Lot Number 24IMI096;  Medline Kit Number/SKU DYNJT2182K:  UDI/DI 10195327434502 (each), 40195327434503 (case), Lot Number 24FME595;  Medline Kit Number/SKU DYNJT2182K:  UDI/DI 10195327434502 (each), 40195327434503 (case), Lot Number 24EMJ896;  Medline Kit Number/SKU DYNJT2182K:  UDI/DI 10195327434502 (each), 40195327434503 (case), Lot Number 23LMG401;  Medline Kit Number/SKU DYNJT2182K:  UDI/DI 10195327434502 (each), 40195327434503 (case), Lot Number 23KMB937;  Medline Kit Number/SKU DYNJT2182K:  UDI/DI 10195327434502 (each), 40195327434503 (case), Lot Number 23IMG801;  Medline Kit Number/SKU DYNJT2182K:  UDI/DI 10195327434502 (each), 40195327434503 (case), Lot Number 23GMA355;  Medline Kit Number/SKU DYNJT2182K:  UDI/DI 10195327434502 (each), 40195327434503 (case), Lot Number 23FMF575;  Medline Kit Number/SKU DYNJT4318:  UDI/DI 10198459447358 (each), 40198459447359 (case), Lot Number 25KBR934;  Medline Kit Number/SKU DYNJT4318:  UDI/DI 10198459447358 (each), 40198459447359 (case), Lot Number 25JBV072;  Medline Kit Number/SKU DYNJT4318:  UDI/DI 10198459447358 (each), 40198459447359 (case), Lot Number 25HBM756;  Medline Kit Number/SKU DYNJT6353:  UDI/DI 10198459547713 (each), 40198459547714 (case), Lot Number 25LBK247;  Medline Kit Number/SKU DYNJT6353A:  UDI/DI 10198459700996 (each), 40198459700997 (case), Lot Number 26CBD905;  Medline Kit Number/SKU DYNJ43571B:  UDI/DI 10889942030057 (each), 40889942030058 (case), Lot Number 24GBS679;  Medline Kit Number/SKU DYNJ43571B:  UDI/DI 10889942030057 (each), 40889942030058 (case), Lot Number 24FBT446;  Medline Kit Number/SKU DYNJ43571B:  UDI/DI 10889942030057 (each), 40889942030058 (case), Lot Number 24DBP456;  Medline Kit Number/SKU DYNJ43571B:  UDI/DI 10889942030057 (each), 40889942030058 (case), Lot Number 23EBP885;  Medline Kit Number/SKU DYNJ43571B:  UDI/DI 10889942030057 (each), 40889942030058 (case), Lot Number 23DBC289;  Medline Kit Number/SKU DYNJ43571B:  UDI/DI 10889942030057 (each), 40889942030058 (case), Lot Number 23CBN047;  Medline Kit Number/SKU DYNJ43571B:  UDI/DI 10889942030057 (each), 40889942030058 (case), Lot Number 22GBP400;  Medline Kit Number/SKU DYNJ43571B:  UDI/DI 10889942030057 (each), 40889942030058 (case), Lot Number 22OBC463;  Medline Kit Number/SKU DYNJ43571B:  UDI/DI 10889942030057 (each), 40889942030058 (case), Lot Number 22OBB289;  Medline Kit Number/SKU DYNJ43571B:  UDI/DI 10889942030057 (each), 40889942030058 (case), Lot Number 22BBT741;  Medline Kit Number/SKU DYNJ43571B:  UDI/DI 10889942030057 (each), 40889942030058 (case), Lot Number 21HBP856;  Medline Kit Number/SKU DYNJ43571B:  UDI/DI 10889942030057 (each), 40889942030058 (case), Lot Number 21DBR386;  Medline Kit Number/SKU DYNJ43571B:  UDI/DI 10889942030057 (each), 40889942030058 (case), Lot Number 21BBF624;  Medline Kit Number/SKU CDS983057G:  UDI/DI 10193489286557 (each), 40193489286558 (case), Lot Number 21GBD835;  Medline Kit Number/SKU CDS983057G:  UDI/DI 10193489286557 (each), 40193489286558 (case), Lot Number 21CBG778;  Medline Kit Number/SKU CDS983057G:  UDI/DI 10193489286557 (each), 40193489286558 (case), Lot Number 21BBD920;  Medline Kit Number/SKU CDS983057I:  UDI/DI 10193489954272 (each), 40193489954273 (case), Lot Number 23FBP396;  Medline Kit Number/SKU CDS983057I:  UDI/DI 10193489954272 (each), 40193489954273 (case), Lot Number 23EBK937;  Medline Kit Number/SKU CDS983057I:  UDI/DI 10193489954272 (each), 40193489954273 (case), Lot Number 23DBG092;  Medline Kit Number/SKU CDS983057I:  UDI/DI 10193489954272 (each), 40193489954273 (case), Lot Number 23BBG730;  Medline Kit Number/SKU CDS983057I:  UDI/DI 10193489954272 (each), 40193489954273 (case), Lot Number 23ABK587;  Medline Kit Number/SKU CDS983057I:  UDI/DI 10193489954272 (each), 40193489954273 (case), Lot Number 22KBW058;  Medline Kit Number/SKU CDS983057I:  UDI/DI 10193489954272 (each), 40193489954273 (case), Lot Number 22HBP142;  Medline Kit Number/SKU CDS983057I:  UDI/DI 10193489954272 (each), 40193489954273 (case), Lot Number 22GBK972;  Medline Kit Number/SKU CDS983057I:  UDI/DI 10193489954272 (each), 40193489954273 (case), Lot Number 22FBQ273;  Medline Kit Number/SKU CDS983057I:  UDI/DI 10193489954272 (each), 40193489954273 (case), Lot Number 22CBW361;  Medline Kit Number/SKU CDS983057I:  UDI/DI 10193489954272 (each), 40193489954273 (case), Lot Number 21LBH702;  Medline Kit Number/SKU CDS983057I:  UDI/DI 10193489954272 (each), 40193489954273 (case), Lot Number 21JBK075;  Medline Kit Number/SKU CDS983057I:  UDI/DI 10193489954272 (each), 40193489954273 (case), Lot Number 21HBL053;  Medline Kit Number/SKU DYNJ909989A:  UDI/DI 10198459221781 (each), 40198459221782 (case), Lot Number 25CBM171;  Medline Kit Number/SKU DYNJ909989C:  UDI/DI 10198459437946 (each), 40198459437947 (case), Lot Number 26CBU149;  Medline Kit Number/SKU DYNJ909989C:  UDI/DI 10198459437946 (each), 40198459437947 (case), Lot Number 26CBD741;  Medline Kit Number/SKU DYNJ909989C:  UDI/DI 10198459437946 (each), 40198459437947 (case), Lot Number 26ABT251;  Medline Kit Number/SKU DYNJ909989C:  UDI/DI 10198459437946 (each), 40198459437947 (case), Lot Number 25KBL337;  Medline Kit Number/SKU DYNJ909989C:  UDI/DI 10198459437946 (each), 40198459437947 (case), Lot Number 25JBN995;  Medline Kit Number/SKU DYNJ0547084N:  UDI/DI 10193489366266 (each), 40193489366267 (case), Lot Number 21BME401;  Medline Kit Number/SKU DYNJ0547084N:  UDI/DI 10193489366266 (each), 40193489366267 (case), Lot Number 21BMB642;  Medline Kit Number/SKU DYNJ0547084N:  UDI/DI 10193489366266 (each), 40193489366267 (case), Lot Number 21BMA102;  Medline Kit Number/SKU DYNJ53820:  UDI/DI 10889942443864 (each), 40889942443865 (case), Lot Number 22CBB527;  Medline Kit Number/SKU DYNJ53820:  UDI/DI 10889942443864 (each), 40889942443865 (case), Lot Number 21LBX592;  Medline Kit Number/SKU DYNJ53820:  UDI/DI 10889942443864 (each), 40889942443865 (case), Lot Number 21KBV148;  Medline Kit Number/SKU DYNJ53820:  UDI/DI 10889942443864 (each), 40889942443865 (case), Lot Number 21KBO735;  Medline Kit Number/SKU DYNJ53820:  UDI/DI 10889942443864 (each), 40889942443865 (case), Lot Number 21KBK868;  Medline Kit Number/SKU DYNJ53820:  UDI/DI 10889942443864 (each), 40889942443865 (case), Lot Number 21HBS913;  Medline Kit Number/SKU DYNJ53820:  UDI/DI 10889942443864 (each), 40889942443865 (case), Lot Number 21GBI825;  Medline Kit Number/SKU CDS984860G:  UDI/DI 10195327274733 (each), 40195327274734 (case), Lot Number 24DDB586;  Medline Kit Number/SKU CDS984860G:  UDI/DI 10195327274733 (each), 40195327274734 (case), Lot Number 24DDC112;  Medline Kit Number/SKU CDS984860G:  UDI/DI 10195327274733 (each), 40195327274734 (case), Lot Number 24DDB288;  Medline Kit Number/SKU CDS984860G:  UDI/DI 10195327274733 (each), 40195327274734 (case), Lot Number 24ADA978;  Medline Kit Number/SKU CDS984860G:  UDI/DI 10195327274733 (each), 40195327274734 (case), Lot Number 23KDB957;  Medline Kit Number/SKU CDS984860G:  UDI/DI 10195327274733 (each), 40195327274734 (case), Lot Number 23JDC102;  Medline Kit Number/SKU CDS984860G:  UDI/DI 10195327274733 (each), 40195327274734 (case), Lot Number 23HDC089;  Medline Kit Number/SKU CDS984860G:  UDI/DI 10195327274733 (each), 40195327274734 (case), Lot Number 23GDB274;  Medline Kit Number/SKU CDS984860G:  UDI/DI 10195327274733 (each), 40195327274734 (case), Lot Number 23EDB865;  Medline Kit Number/SKU CDS984860G:  UDI/DI 10195327274733 (each), 40195327274734 (case), Lot Number 23EDB036;  Medline Kit Number/SKU CDS984860G:  UDI/DI 10195327274733 (each), 40195327274734 (case), Lot Number 23BDB418;  Medline Kit Number/SKU CDS984860G:  UDI/DI 10195327274733 (each), 40195327274734 (case), Lot Number 23ADB664;  Medline Kit Number/SKU CDS984860G:  UDI/DI 10195327274733 (each), 40195327274734 (case), Lot Number 23ADA140;  Medline Kit Number/SKU DYNJ908882C:  UDI/DI 10198459705847 (each), 40198459705848 (case), Lot Number 26CBQ081;  Medline Kit Number/SKU DYNJ903399G:  UDI/DI 10195327471712 (each), 40195327471713 (case), Lot Number 23IBI075;  Medline Kit Number/SKU DYNJ903399K:  UDI/DI 10198459522703 (each), 40198459522704 (case), Lot Number 26DBF271;  Medline Kit Number/SKU DYNJ903399K:  UDI/DI 10198459522703 (each), 40198459522704 (case), Lot Number 26ABX052;  Medline Kit Number/SKU DYNJ903399K:  UDI/DI 10198459522703 (each), 40198459522704 (case), Lot Number 25KBC437;  Medline Kit Number/SKU DYNJ0842931J:  UDI/DI 10193489352245 (each), 40193489352246 (case), Lot Number 22LDA417;  Medline Kit Number/SKU DYNJ83817:  UDI/DI 10195327310264 (each), 40195327310265 (case), Lot Number 23EBU906;  Medline Kit Number/SKU DYNJ83817:  UDI/DI 10195327310264 (each), 40195327310265 (case), Lot Number 23EBJ268;  Medline Kit Number/SKU DYNJ83817:  UDI/DI 10195327310264 (each), 40195327310265 (case), Lot Number 23DBN861;  Medline Kit Number/SKU DYNJ83817:  UDI/DI 10195327310264 (each), 40195327310265 (case), Lot Number 23BBJ439;  Medline Kit Number/SKU DYNJ902174D:  UDI/DI 10195327129811 (each), 40195327129812 (case), Lot Number 22HBE047;  Medline Kit Number/SKU DYNJ902174D:  UDI/DI 10195327129811 (each), 40195327129812 (case), Lot Number 22EBO445;  Medline Kit Number/SKU DYNJ902174F:  UDI/DI 10195327187088 (each), 40195327187089 (case), Lot Number 23DBG308;  Medline Kit Number/SKU DYNJ902174F:  UDI/DI 10195327187088 (each), 40195327187089 (case), Lot Number 23BBG705;  Medline Kit Number/SKU DYNJ902174F:  UDI/DI 10195327187088 (each), 40195327187089 (case), Lot Number 22JBR859;  Medline Kit Number/SKU DYNJ902174F:  UDI/DI 10195327187088 (each), 40195327187089 (case), Lot Number 22IBN495;  Medline Kit Number/SKU DYNJ902174F:  UDI/DI 10195327187088 (each), 40195327187089 (case), Lot Number 22GBW268;  Medline Kit Number/SKU DYNJ902174G:  UDI/DI 10195327389390 (each), 40195327389391 (case), Lot Number 25FBS517;  Medline Kit Number/SKU DYNJ902174G:  UDI/DI 10195327389390 (each), 40195327389391 (case), Lot Number 25FBG131;  Medline Kit Number/SKU DYNJ902174G:  UDI/DI 10195327389390 (each), 40195327389391 (case), Lot Number 25CBT671;  Medline Kit Number/SKU DYNJ902174G:  UDI/DI 10195327389390 (each), 40195327389391 (case), Lot Number 25ABW503;  Medline Kit Number/SKU DYNJ902174G:  UDI/DI 10195327389390 (each), 40195327389391 (case), Lot Number 24LBS163;  Medline Kit Number/SKU DYNJ902174G:  UDI/DI 10195327389390 (each), 40195327389391 (case), Lot Number 24IBV342;  Medline Kit Number/SKU DYNJ902174G:  UDI/DI 10195327389390 (each), 40195327389391 (case), Lot Number 24HBB109;  Medline Kit Number/SKU DYNJ902174G:  UDI/DI 10195327389390 (each), 40195327389391 (case), Lot Number 24FBS476;  Medline Kit Number/SKU DYNJ902174G:  UDI/DI 10195327389390 (each), 40195327389391 (case), Lot Number 24EBD091;  Medline Kit Number/SKU DYNJ902174G:  UDI/DI 10195327389390 (each), 40195327389391 (case), Lot Number 24BBT254;  Medline Kit Number/SKU DYNJ902174G:  UDI/DI 10195327389390 (each), 40195327389391 (case), Lot Number 24ABV360;  Medline Kit Number/SKU DYNJ902174G:  UDI/DI 10195327389390 (each), 40195327389391 (case), Lot Number 23HBO344;  Medline Kit Number/SKU DYNJ902174G:  UDI/DI 10195327389390 (each), 40195327389391 (case), Lot Number 23HBM943;  Medline Kit Number/SKU DYNJ902174G:  UDI/DI 10195327389390 (each), 40195327389391 (case), Lot Number 23GBD659;  Medline Kit Number/SKU DYNJ902174G:  UDI/DI 10195327389390 (each), 40195327389391 (case), Lot Number 23EBF906;  Medline Kit Number/SKU DYNJ80825B:  UDI/DI 10195327459987 (each), 40195327459988 (case), Lot Number 26DBC300;  Medline Kit Number/SKU DYNJ80825B:  UDI/DI 10195327459987 (each), 40195327459988 (case), Lot Number 26BBO483;  Medline Kit Number/SKU DYNJ80825B:  UDI/DI 10195327459987 (each), 40195327459988 (case), Lot Number 26ABJ364;  Medline Kit Number/SKU DYNJ80825B:  UDI/DI 10195327459987 (each), 40195327459988 (case), Lot Number 25KBP940;  Medline Kit Number/SKU DYNJ80825B:  UDI/DI 10195327459987 (each), 40195327459988 (case), Lot Number 25IBU145;  Medline Kit Number/SKU DYNJ80825B:  UDI/DI 10195327459987 (each), 40195327459988 (case), Lot Number 25GBD445;  Medline Kit Number/SKU DYNJ80825B:  UDI/DI 10195327459987 (each), 40195327459988 (case), Lot Number 25EBM158;  Medline Kit Number/SKU DYNJ80825B:  UDI/DI 10195327459987 (each), 40195327459988 (case), Lot Number 25CBL164;  Medline Kit Number/SKU DYNJ80825B:  UDI/DI 10195327459987 (each), 40195327459988 (case), Lot Number 25ABW151;  Medline Kit Number/SKU DYNJ80825B:  UDI/DI 10195327459987 (each), 40195327459988 (case), Lot Number 24LBB488;  Medline Kit Number/SKU DYNJ80825B:  UDI/DI 10195327459987 (each), 40195327459988 (case), Lot Number 24JBI981;  Medline Kit Number/SKU DYNJ80825B:  UDI/DI 10195327459987 (each), 40195327459988 (case), Lot Number 24GBF805;  Medline Kit Number/SKU DYNJ80825B:  UDI/DI 10195327459987 (each), 40195327459988 (case), Lot Number 24EBL328;  Medline Kit Number/SKU DYNJ80825B:  UDI/DI 10195327459987 (each), 40195327459988 (case), Lot Number 24BBN982;  Medline Kit Number/SKU DYNJ80825B:  UDI/DI 10195327459987 (each), 40195327459988 (case), Lot Number 24ABB409;  Medline Kit Number/SKU DYNJ80825B:  UDI/DI 10195327459987 (each), 40195327459988 (case), Lot Number 23JBR651;  Medline Kit Number/SKU DYNJ80825B:  UDI/DI 10195327459987 (each), 40195327459988 (case), Lot Number 23GBI463;  Medline Kit Number/SKU DYNJ03387A:  UDI/DI 10884389906033 (each), 40884389906034 (case), Lot Number 22CMG065;  Medline Kit Number/SKU DYNJ03387A:  UDI/DI 10884389906033 (each), 40884389906034 (case), Lot Number 22BME711;  Medline Kit Number/SKU DYNJ03387A:  UDI/DI 10884389906033 (each), 40884389906034 (case), Lot Number 21KMF815;  Medline Kit Number/SKU DYNJ03387A:  UDI/DI 10884389906033 (each), 40884389906034 (case), Lot Number 21IMF945;  Medline Kit Number/SKU DYNJ03387A:  UDI/DI 10884389906033 (each), 40884389906034 (case), Lot Number 21IME838;  Medline Kit Number/SKU DYNJ03387A:  UDI/DI 10884389906033 (each), 40884389906034 (case), Lot Number 21GMD526;  Medline Kit Number/SKU DYNJ03387A:  UDI/DI 10884389906033 (each), 40884389906034 (case), Lot Number 21FMF570;  Medline Kit Number/SKU DYNJ03387A:  UDI/DI 10884389906033 (each), 40884389906034 (case), Lot Number 21DMA953;  Medline Kit Number/SKU DYNJ03387A:  UDI/DI 10884389906033 (each), 40884389906034 (case), Lot Number 21CMB888;  Medline Kit Number/SKU DYNJ03387A:  UDI/DI 10884389906033 (each), 40884389906034 (case), Lot Number 21BMB780;  Medline Kit Number/SKU DYNJ03387A:  UDI/DI 10884389906033 (each), 40884389906034 (case), Lot Number 21AMB961;  Medline Kit Number/SKU DYNJ30086B:  UDI/DI 10888277877269 (each), 40888277877260 (case), Lot Number 26CBL791;  Medline Kit Number/SKU DYNJ30086B:  UDI/DI 10888277877269 (each), 40888277877260 (case), Lot Number 25KBQ796;  Medline Kit Number/SKU DYNJ30086B:  UDI/DI 10888277877269 (each), 40888277877260 (case), Lot Number 25HBG829;  Medline Kit Number/SKU DYNJ30086B:  UDI/DI 10888277877269 (each), 40888277877260 (case), Lot Number 25HBG861;  Medline Kit Number/SKU DYNJ30086B:  UDI/DI 10888277877269 (each), 40888277877260 (case), Lot Number 25ABQ214;  Medline Kit Number/SKU DYNJ30086B:  UDI/DI 10888277877269 (each), 40888277877260 (case), Lot Number 25ABH194;  Medline Kit Number/SKU DYNJ30086B:  UDI/DI 10888277877269 (each), 40888277877260 (case), Lot Number 24JBJ082;  Medline Kit Number/SKU DYNJ30086B:  UDI/DI 10888277877269 (each), 40888277877260 (case), Lot Number 24HBU763;  Medline Kit Number/SKU DYNJ30086B:  UDI/DI 10888277877269 (each), 40888277877260 (case), Lot Number 24EBD563;  Medline Kit Number/SKU DYNJ30086B:  UDI/DI 10888277877269 (each), 40888277877260 (case), Lot Number 24ABT266;  Medline Kit Number/SKU DYNJ30086B:  UDI/DI 10888277877269 (each), 40888277877260 (case), Lot Number 24ABG336;  Medline Kit Number/SKU DYNJ30086B:  UDI/DI 10888277877269 (each), 40888277877260 (case), Lot Number 23IBB896;  Medline Kit Number/SKU DYNJ30086B:  UDI/DI 10888277877269 (each), 40888277877260 (case), Lot Number 23HBM969;  Medline Kit Number/SKU DYNJ30086B:  UDI/DI 10888277877269 (each), 40888277877260 (case), Lot Number 23GBN163;  Medline Kit Number/SKU DYNJ30086B:  UDI/DI 10888277877269 (each), 40888277877260 (case), Lot Number 23CBW178;  Medline Kit Number/SKU DYNJ30086B:  UDI/DI 10888277877269 (each), 40888277877260 (case), Lot Number 23ABL227;  Medline Kit Number/SKU DYNJ30086B:  UDI/DI 10888277877269 (each), 40888277877260 (case), Lot Number 23ABF499;  Medline Kit Number/SKU DYNJ30086B:  UDI/DI 10888277877269 (each), 40888277877260 (case), Lot Number 22IBP245;  Medline Kit Number/SKU DYNJ30086B:  UDI/DI 10888277877269 (each), 40888277877260 (case), Lot Number 22DBR825;  Medline Kit Number/SKU DYNJ30086B:  UDI/DI 10888277877269 (each), 40888277877260 (case), Lot Number 22BBW610;  Medline Kit Number/SKU DYNJ30086B:  UDI/DI 10888277877269 (each), 40888277877260 (case), Lot Number 22BBD864;  Medline Kit Number/SKU DYNJ30086B:  UDI/DI 10888277877269 (each), 40888277877260 (case), Lot Number 21KBM679;  Medline Kit Number/SKU DYNJ30086B:  UDI/DI 10888277877269 (each), 40888277877260 (case), Lot Number 21IBN470;  Medline Kit Number/SKU DYNJ30086B:  UDI/DI 10888277877269 (each), 40888277877260 (case), Lot Number 21HBR595;  Medline Kit Number/SKU DYNJ30086B:  UDI/DI 10888277877269 (each), 40888277877260 (case), Lot Number 21GBJ474;  Medline Kit Number/SKU DYNJ30086B:  UDI/DI 10888277877269 (each), 40888277877260 (case), Lot Number 21ABH496;  Medline Kit Number/SKU DYNJ53623A:  UDI/DI 10198459036361 (each), 40198459036362 (case), Lot Number 26BME479;  Medline Kit Number/SKU DYNJ53623A:  UDI/DI 10198459036361 (each), 40198459036362 (case), Lot Number 25LMG372;  Medline Kit Number/SKU DYNJ53623A:  UDI/DI 10198459036361 (each), 40198459036362 (case), Lot Number 25CMC926;  Medline Kit Number/SKU DYNJ53623A:  UDI/DI 10198459036361 (each), 40198459036362 (case), Lot Number 24FMD996;  Medline Kit Number/SKU DYNJ53623A:  UDI/DI 10198459036361 (each), 40198459036362 (case), Lot Number 25KMD669;  Medline Kit Number/SKU DYNJ51750D:  UDI/DI 10198459496912 (each), 40198459496913 (case), Lot Number 26CBT187;  Medline Kit Number/SKU DYNJ51750D:  UDI/DI 10198459496912 (each), 40198459496913 (case), Lot Number 26ABF642;  Medline Kit Number/SKU DYNJ51750D:  UDI/DI 10198459496912 (each), 40198459496913 (case), Lot Number 25JBT713;  Medline Kit Number/SKU DYNJ51750D:  UDI/DI 10198459496912 (each), 40198459496913 (case), Lot Number 25IBP661;  Medline Kit Number/SKU DYNJ902191G:  UDI/DI 10195327390457 (each), 40195327390458 (case), Lot Number 23FBR342;  Medline Kit Number/SKU DYNJ902191G:  UDI/DI 10195327390457 (each), 40195327390458 (case), Lot Number 23EBW660;  Medline Kit Number/SKU DYNJ902191I:  UDI/DI 10195327474935 (each), 40195327474936 (case), Lot Number 23JBW744;  Medline Kit Number/SKU DYNJ902191I:  UDI/DI 10195327474935 (each), 40195327474936 (case), Lot Number 23JBC157;  Medline Kit Number/SKU DYNJ902191L:  UDI/DI 10198459350177 (each), 40198459350178 (case), Lot Number 26CMF220;  Medline Kit Number/SKU DYNJ902191L:  UDI/DI 10198459350177 (each), 40198459350178 (case), Lot Number 25KBD577;  Medline Kit Number/SKU DYNJ902191L:  UDI/DI 10198459350177 (each), 40198459350178 (case), Lot Number 25HBP496;  Medline Kit Number/SKU DYNJ10127D:  UDI/DI 10195327338084 (each), 40195327338085 (case), Lot Number 23KBD073;  Medline Kit Number/SKU DYNJ10127D:  UDI/DI 10195327338084 (each), 40195327338085 (case), Lot Number 23HBU196;  Medline Kit Number/SKU DYNJ10127D:  UDI/DI 10195327338084 (each), 40195327338085 (case), Lot Number 23EBL329;  Medline Kit Number/SKU DYNJ10127D:  UDI/DI 10195327338084 (each), 40195327338085 (case), Lot Number 23DBQ745;  Medline Kit Number/SKU DYNJ10127F:  UDI/DI 10195327576141 (each), 40195327576142 (case), Lot Number 26ABM131;  Medline Kit Number/SKU DYNJ10127F:  UDI/DI 10195327576141 (each), 40195327576142 (case), Lot Number 25KBL442;  Medline Kit Number/SKU DYNJ10127F:  UDI/DI 10195327576141 (each), 40195327576142 (case), Lot Number 25HBL187;  Medline Kit Number/SKU DYNJ10127F:  UDI/DI 10195327576141 (each), 40195327576142 (case), Lot Number 25EBF377;  Medline Kit Number/SKU DYNJ10127F:  UDI/DI 10195327576141 (each), 40195327576142 (case), Lot Number 25BBK501;  Medline Kit Number/SKU DYNJ10127F:  UDI/DI 10195327576141 (each), 40195327576142 (case), Lot Number 24LBA162;  Medline Kit Number/SKU DYNJ10127F:  UDI/DI 10195327576141 (each), 40195327576142 (case), Lot Number 24JBN581;  Medline Kit Number/SKU DYNJ10127F:  UDI/DI 10195327576141 (each), 40195327576142 (case), Lot Number 24IBK625;  Medline Kit Number/SKU DYNJ10127F:  UDI/DI 10195327576141 (each), 40195327576142 (case), Lot Number 24HBU132;  Medline Kit Number/SKU DYNJ10127F:  UDI/DI 10195327576141 (each), 40195327576142 (case), Lot Number 24CBM709;  Medline Kit Number/SKU DYNJ10127F:  UDI/DI 10195327576141 (each), 40195327576142 (case), Lot Number 24BBS053;  Medline Kit Number/SKU DYNJ10127F:  UDI/DI 10195327576141 (each), 40195327576142 (case), Lot Number 24BBG854;  Medline Kit Number/SKU DYNJ10127F:  UDI/DI 10195327576141 (each), 40195327576142 (case), Lot Number 24ABP138;  Medline Kit Number/SKU DYNJ902663M:  UDI/DI 10198459448911 (each), 40198459448912 (case), Lot Number 26AML079;  Medline Kit Number/SKU DYNJ902663M:  UDI/DI 10198459448911 (each), 40198459448912 (case), Lot Number 25LMJ978;  Medline Kit Number/SKU DYNJ902663M:  UDI/DI 10198459448911 (each), 40198459448912 (case), Lot Number 25IMI287;  Medline Kit Number/SKU DYNJ902663M:  UDI/DI 10198459448911 (each), 40198459448912 (case), Lot Number 25HMH343;  Medline Kit Number/SKU DYNJ902663M:  UDI/DI 10198459448911 (each), 40198459448912 (case), Lot Number 25HMD911;  Medline Kit Number/SKU DYNJ902663N:  UDI/DI 10198459675058 (each), 40198459675059 (case), Lot Number 26DMH785."
    }
  ]
}