  "results": [
    {
      "country": "United States",
      "city": "Mounds View",
      "address_1": "8200 Coral Sea St Ne",
      "reason_for_recall": "There is a rare communication sequence during the first device interrogation with a SmartSync Device Manager that may result in the temporary suspension of some device features (i.e., battery measurements, Capture Management\", Atrial Lead Position Check\", EffectivCRT\" algorithms, and AdaptivCRT\").  This rare interaction results in temporary suspension of automatic threshold testing and output adjustments, and suspension of auto-optimization of CRT therapy. The issue is unlikely to result in clinical impact to the patient, and features are restored upon next programmer device interrogation or presence of a magnet.",
      "address_2": "",
      "product_quantity": "2773 devices",
      "code_info": "serial numbers:  SPM001078A  SPM001117A  SPM001120A  SPM001139A  SPM001140A  SPM001141A  SPM001174A  SPM001231A  SPM001250A  SPM001251A  SPM001252A  SPM001253A  SPM001262A  SPM001265A  SPM001266A  SPM001270A  SPM001281A  SPM001284A  SPM001285A  SPM001363A  SPM001377A  SPM001394A  SPM001396A  SPM001398A  SPM001403A  SPM001404A  SPM001407A  SPM001419A  SPM001424A  SPM001425A  SPM001441A  SPM001445A  SPM001449A  SPM001456A  SPM001463A  SPM001465A  SPM001474A  SPM001477A  SPM001478A  SPM001482A  SPM001484A  SPM001486A  SPM001491A  SPM001519A  SPM001544A  SPM001545A  SPM001552A  SPM001554A  SPM001555A  SPM001556A  SPM001557A  SPM001559A  SPM001560A  SPM001561A  SPM001562A  SPM001571A  SPM001581A  SPM001584A  SPM001587A  SPM001588A  SPM001633A  SPM001659A  SPM001660A  SPM001661A  SPM001662A  SPM001663A  SPM001664A  SPM001665A  SPM001666A  SPM001667A  SPM001668A  SPM001669A  SPM001670A  SPM001671A  SPM001672A  SPM001673A  SPM001674A  SPM001675A  SPM001676A  SPM001677A  SPM001683A  SPM001684A  SPM001685A  SPM001686A  SPM001704A  SPM001705A  SPM001706A  SPM001707A  SPM001709A  SPM001710A  SPM001711A  SPM001712A  SPM001713A  SPM001714A  SPM001715A  SPM001716A  SPM001719A  SPM001720A  SPM001721A  SPM001723A  SPM001732A  SPM001733A  SPM001734A  SPM001735A  SPM001738A  SPM001739A  SPM001745A  SPM001746A  SPM001747A  SPM001748A  SPM001749A  SPM001751A  SPM001753A  SPM001754A  SPM001756A  SPM001759A  SPM001760A  SPM001766A  SPM001771A  SPM001774A  SPM001775A  SPM001776A  SPM001779A  SPM001781A  SPM001782A  SPM001784A  SPM001786A  SPM001787A  SPM001793A  SPM001795A  SPM001796A  SPM001798A  SPM001799A  SPM001800A  SPM001802A  SPM001803A  SPM001805A  SPM001806A  SPM001807A  SPM001808A  SPM001809A  SPM001812A  SPM001813A  SPM001816A  SPM001817A  SPM001818A  SPM001842A  SPM001846A  SPM001875A  SPM001878A  SPM001880A  SPM001883A  SPM001884A  SPM001885A  SPM001889A  SPM001896A  SPM001901A  SPM001929A  SPM001935A  SPM001936A  SPM001961A  SPM001967A  SPM001971A  SPM001973A  SPM001974A  SPM001975A  SPM001976A  SPM001977A  SPM001982A  SPM001983A  SPM001984A  SPM001985A  SPM001986A  SPM001988A  SPM001989A  SPM002001A  SPM002005A  SPM002024A  SPM002035A  SPM002041A  SPM002068A  SPM002072A  SPM002085A  SPM002107A  SPM002108A  SPM002109A  SPM002110A  SPM002113A  SPM002114A  SPM002115A  SPM002117A  SPM002121A  SPM002122A  SPM002125A  SPM002129A  SPM002130A  SPM002131A  SPM002132A  SPM002134A  SPM002145A  SPM002150A  SPM002179A  SPM002198A  SPM002205A  SPM002213A  SPM002217A  SPM002219A  SPM002232A  SPM002268A  SPM002269A  SPM002292A  SPM002294A  SPM002295A  SPM002310A  SPM002312A  SPM002315A  SPM002316A  SPM002335A  SPM002342A  SPM002343A  SPM002344A  SPM002347A  SPM002365A  SPM002367A  SPM002373A  SPM002374A  SPM002375A  SPM002386A  SPM002416A  SPM002424A  SPM002451A  SPM002456A  SPM002458A  SPM002459A  SPM002460A  SPM002461A  SPM002462A  SPM002463A  SPM002464A  SPM002465A  SPM002538A  SPM002540A  SPM002542A  SPM002546A  SPM002547A  SPM002553A  SPM002554A  SPM002555A  SPM002557A  SPM002558A  SPM002559A  SPM002560A  SPM002561A  SPM002562A  SPM002563A  SPM002565A  SPM002568A  SPM002569A  SPM002572A  SPM002595A  SPM002597A  SPM002599A  SPM002600A  SPM002601A  SPM002602A  SPM002603A  SPM002604A  SPM002606A  SPM002608A  SPM002609A  SPM002610A  SPM002611A  SPM002612A  SPM002613A  SPM002614A  SPM002615A  SPM002616A  SPM002617A  SPM002618A  SPM002656A  SPM002657A  SPM002658A  SPM002659A  SPM002660A  SPM002661A  SPM002662A  SPM002663A  SPM002664A  SPM002665A  SPM002698A  SPM002699A  SPM002703A  SPM002704A  SPM002705A  SPM002706A  SPM002708A  SPM002709A  SPM002710A  SPM002711A  SPM002712A  SPM002714A  SPM002736A  SPM002756A  SPM002885A  SPM002997A  SPM003000A  SPM003001A  SPM003035A  SPM003036A  SPM003037A  SPM003038A  SPM003042A  SPM003043A  SPM003046A  SPM003047A  SPM003048A  SPM003051A  SPM003053A  SPM003054A  SPM003058A  SPM003059A  SPM003060A  SPM003061A  SPM003063A  SPM003064A  SPM003065A  SPM003066A  SPM003067A  SPM003068A  SPM003069A  SPM003070A  SPM003071A  SPM003073A  SPM003074A  SPM003078A  SPM003079A  SPM003080A  SPM003081A  SPM003082A  SPM003084A  SPM003085A  SPM003087A  SPM003088A  SPM003092A  SPM003105A  SPM003108A  SPM003134A  SPM003136A  SPM003137A  SPM003140A  SPM003142A  SPM003410A  SPM003411A  SPM003444A  SPM003446A  SPM003447A  SPM003448A  SPM003449A  SPM003450A  SPM003451A  SPM003465A  SPM003472A  SPM003473A  SPM003474A  SPM003475A  SPM003476A  SPM003477A  SPM003478A  SPM003479A  SPM003480A  SPM003481A  SPM003482A  SPM003516A  SPM003517A  SPM003518A  SPM003519A  SPM003520A  SPM003521A  SPM003522A  SPM003523A  SPM003524A  SPM003525A  SPM003529A  SPM003530A  SPM003531A  SPM003532A  SPM003533A  SPM003534A  SPM003535A  SPM003536A  SPM003537A  SPM003538A  SPM003557A  SPM003558A  SPM003559A  SPM003565A  SPM003587A  SPM003589A  SPM003590A  SPM003591A  SPM003592A  SPM003593A  SPM003594A  SPM003595A  SPM003596A  SPM003597A  SPM003598A  SPM003599A  SPM003600A  SPM003601A  SPM003602A  SPM003605A  SPM003607A  SPM003608A  SPM003610A  SPM003611A  SPM003612A  SPM003668A  SPM003669A  SPM003679A  SPM003680A  SPM003682A  SPM003683A  SPM003684A  SPM003685A  SPM003686A  SPM003687A  SPM003689A  SPM003690A  SPM003691A  SPM003692A  SPM003697A  SPM003698A  SPM003699A  SPM003700A  SPM003701A  SPM003702A  SPM003703A  SPM003704A  SPM003705A  SPM003706A  SPM003708A  SPM003712A  SPM003716A  SPM003717A  SPM003718A  SPM003719A  SPM003721A  SPM003722A  SPM003723A  SPM003725A  SPM003741A  SPM003742A  SPM003744A  SPM003747A  SPM003748A  SPM003749A  SPM003751A  SPM003762A  SPM003843A  SPM003850A  SPM003852A  SPM003877A  SPM003916A  SPM003993A  SPM004088A  SPM004309A  SPM004351A  SPM004460A  SPM004463A  SPM004465A  SPM004466A  SPM004474A  SPM004475A  SPM004477A  SPM004481A  SPM004482A  SPM004484A  SPM004487A  SPM004489A  SPM004501A  SPM004503A  SPM004505A  SPM004506A  SPM004507A  SPM004511A  SPM004512A  SPM004513A  SPM004517A  SPM004518A  SPM004519A  SPM004520A  SPM004521A  SPM004524A  SPM004525A  SPM004526A  SPM004527A  SPM004528A  SPM004529A  SPM004530A  SPM004536A  SPM004548A  SPM004561A  SPM004571A  SPM004573A  SPM004574A  SPM004575A  SPM004576A  SPM004577A  SPM004578A  SPM004579A  SPM004580A  SPM004581A  SPM004582A  SPM004583A  SPM004584A  SPM004585A  SPM004589A  SPM004590A  SPM004591A  SPM004592A  SPM004594A  SPM004595A  SPM004606A  SPM004607A  SPM004618A  SPM004619A  SPM004620A  SPM004621A  SPM004622A  SPM004623A  SPM004624A  SPM004625A  SPM004626A  SPM004627A  SPM004653A  SPM004654A  SPM004670A  SPM004671A  SPM004672A  SPM004673A  SPM004674A  SPM004675A  SPM004676A  SPM004677A  SPM004678A  SPM004679A  SPM004680A  SPM004681A  SPM004682A  SPM004683A  SPM004684A  SPM004685A  SPM004686A  SPM004687A  SPM004688A  SPM004689A  SPM004690A  SPM004691A  SPM004692A  SPM004693A  SPM004694A  SPM004695A  SPM004696A  SPM004697A  SPM004698A  SPM004700A  SPM004701A  SPM004702A  SPM004704A  SPM004707A  SPM004708A  SPM004709A  SPM004712A  SPM004713A  SPM004714A  SPM004715A  SPM004717A  SPM004718A  SPM004719A  SPM004720A  SPM004721A  SPM004723A  SPM004724A  SPM004725A  SPM004726A  SPM004727A  SPM004728A  SPM004729A  SPM004730A  SPM004731A  SPM004732A  SPM004735A  SPM004737A  SPM004738A  SPM004743A  SPM004746A  SPM004747A  SPM004749A  SPM004750A  SPM004756A  SPM004764A  SPM004766A  SPM004769A  SPM004771A  SPM004772A  SPM004773A  SPM004774A  SPM004775A  SPM004776A  SPM004777A  SPM004778A  SPM004779A  SPM004781A  SPM004782A  SPM004783A  SPM004784A  SPM004786A  SPM004787A  SPM004788A  SPM004791A  SPM004793A  SPM004795A  SPM004796A  SPM004797A  SPM004798A  SPM004799A  SPM004800A  SPM004801A  SPM004802A  SPM004805A  SPM004931A  SPM004932A  SPM004933A  SPM004964A  SPM004965A  SPM004982A  SPM004984A  SPM004986A  SPM004987A  SPM004988A  SPM004989A  SPM004990A  SPM004991A  SPM004992A  SPM004993A  SPM004994A  SPM004995A  SPM004996A  SPM004997A  SPM004999A  SPM005001A  SPM005002A  SPM005005A  SPM005008A  SPM005009A  SPM005013A  SPM005021A  SPM005023A  SPM005024A  SPM005025A  SPM005027A  SPM005028A  SPM005029A  SPM005031A  SPM005032A  SPM005033A  SPM005034A  SPM005035A  SPM005036A  SPM005037A  SPM005038A  SPM005039A  SPM005040A  SPM005041A  SPM005042A  SPM005044A  SPM005045A  SPM005047A  SPM005048A  SPM005049A  SPM005050A  SPM005051A  SPM005052A  SPM005053A  SPM005054A  SPM005055A  SPM005056A  SPM005057A  SPM005058A  SPM005060A  SPM005061A  SPM005065A  SPM005066A  SPM005067A  SPM005068A  SPM005069A  SPM005074A  SPM005075A  SPM005076A  SPM005077A  SPM005078A  SPM005080A  SPM005081A  SPM005083A  SPM005084A  SPM005085A  SPM005087A  SPM005088A  SPM005089A  SPM005091A  SPM005092A  SPM005093A  SPM005095A  SPM005096A  SPM005098A  SPM005124A  SPM005127A  SPM005140A  SPM005142A  SPM005143A  SPM005154A  SPM005156A  SPM005157A  SPM005158A  SPM005173A  SPM005179A  SPM005181A  SPM005183A  SPM005185A  SPM005192A  SPM005195A  SPM005197A  SPM005200A  SPM005203A  SPM005213A  SPM005219A  SPM005231A  SPM005232A  SPM005246A  SPM005251A  SPM005253A  SPM005262A  SPM005278A  SPM005281A  SPM005282A  SPM005301A  SPM005302A  SPM005305A  SPM005308A  SPM005310A  SPM005311A  SPM005312A  SPM005313A  SPM005318A  SPM005322A  SPM005323A  SPM005324A  SPM005325A  SPM005326A  SPM005327A  SPM005328A  SPM005329A  SPM005330A  SPM005333A  SPM005334A  SPM005335A  SPM005337A  SPM005341A  SPM005344A  SPM005347A  SPM005352A  SPM005353A  SPM005357A  SPM005358A  SPM005360A  SPM005362A  SPM005363A  SPM005364A  SPM005366A  SPM005367A  SPM005368A  SPM005369A  SPM005372A  SPM005374A  SPM005376A  SPM005379A  SPM005380A  SPM005381A  SPM005386A  SPM005389A  SPM005391A  SPM005392A  SPM005393A  SPM005394A  SPM005402A  SPM005404A  SPM005407A  SPM005408A  SPM005410A  SPM005411A  SPM005413A  SPM005416A  SPM005417A  SPM005421A  SPM005424A  SPM005425A  SPM005429A  SPM005430A  SPM005431A  SPM005432A  SPM005433A  SPM005437A  SPM005439A  SPM005440A  SPM005441A  SPM005443A  SPM005444A  SPM005445A  SPM005446A  SPM005447A  SPM005448A  SPM005450A  SPM005452A  SPM005453A  SPM005456A  SPM005457A  SPM005458A  SPM005460A  SPM005463A  SPM005464A  SPM005465A  SPM005468A  SPM005472A  SPM005505A  SPM005506A  SPM005553A  SPM005556A  SPM005561A  SPM005564A  SPM005566A  SPM005572A  SPM005577A  SPM005578A  SPM005580A  SPM005589A  SPM005596A  SPM005597A  SPM005598A  SPM005600A  SPM005601A  SPM005602A  SPM005606A  SPM005607A  SPM005608A  SPM005609A  SPM005623A  SPM005631A  SPM005632A  SPM005637A  SPM005640A  SPM005643A  SPM005645A  SPM005649A  SPM005650A  SPM005651A  SPM005652A  SPM005653A  SPM005654A  SPM005656A  SPM005661A  SPM005663A  SPM005668A  SPM005670A  SPM005677A  SPM005679A  SPM005680A  SPM005681A  SPM005682A  SPM005684A  SPM005685A  SPM005687A  SPM005696A  SPM005697A  SPM005700A  SPM005703A  SPM005704A  SPM005708A  SPM005712A  SPM005778A  SPM005837A  SPM005867A  SPM005879A  SPM005884A  SPM005888A  SPM005891A  SPM005943A  SPM005968A  SPM006001A  SPM006005A  SPM006007A  SPM006011A  SPM006021A  SPM006022A  SPM006023A  SPM006024A  SPM006028A  SPM006029A  SPM006030A  SPM006034A  SPM006040A  SPM006041A  SPM006045A  SPM006046A  SPM006051A  SPM006054A  SPM006055A  SPM006057A  SPM006058A  SPM006059A  SPM006062A  SPM006063A  SPM006064A  SPM006065A  SPM006068A  SPM006070A  SPM006072A  SPM006073A  SPM006076A  SPM006078A  SPM006080A  SPM006082A  SPM006109A  SPM006110A  SPM006111A  SPM006122A  SPM006123A  SPM006126A  SPM006127A  SPM006129A  SPM006131A  SPM006132A  SPM006134A  SPM006135A  SPM006136A  SPM006143A  SPM006145A  SPM006146A  SPM006147A  SPM006148A  SPM006151A  SPM006152A  SPM006154A  SPM006155A  SPM006158A  SPM006161A  SPM006162A  SPM006163A  SPM006167A  SPM006169A  SPM006171A  SPM006174A  SPM006175A  SPM006176A  SPM006178A  SPM006186A  SPM006188A  SPM006189A  SPM006192A  SPM006194A  SPM006195A  SPM006196A  SPM006198A  SPM006200A  SPM006201A  SPM006203A  SPM006204A  SPM006205A  SPM006206A  SPM006208A  SPM006213A  SPM006217A  SPM006218A  SPM006221A  SPM006224A  SPM006225A  SPM006226A  SPM006227A  SPM006229A  SPM006230A  SPM006234A  SPM006235A  SPM006236A  SPM006238A  SPM006241A  SPM006243A  SPM006244A  SPM006246A  SPM006247A  SPM006249A  SPM006250A  SPM006252A  SPM006253A  SPM006256A  SPM006260A  SPM006264A  SPM006267A  SPM006270A  SPM006271A  SPM006274A  SPM006275A  SPM006277A  SPM006278A  SPM006281A  SPM006283A  SPM006284A  SPM006285A  SPM006286A  SPM006287A  SPM006288A  SPM006289A  SPM006290A  SPM006300A  SPM006301A  SPM006303A  SPM006304A  SPM006305A  SPM006307A  SPM006309A  SPM006311A  SPM006312A  SPM006314A  SPM006315A  SPM006316A  SPM006317A  SPM006319A  SPM006320A  SPM006321A  SPM006322A  SPM006323A  SPM006324A  SPM006326A  SPM006327A  SPM006329A  SPM006330A  SPM006332A  SPM006334A  SPM006335A  SPM006339A  SPM006340A  SPM006342A  SPM006343A  SPM006345A  SPM006346A  SPM006349A  SPM006350A  SPM006353A  SPM006357A  SPM006359A  SPM006363A  SPM006377A  SPM006389A  SPM006390A  SPM006394A  SPM006398A  SPM006399A  SPM006402A  SPM006411A  SPM006412A  SPM006414A  SPM006416A  SPM006419A  SPM006422A  SPM006424A  SPM006425A  SPM006435A  SPM006436A  SPM006444A  SPM006451A  SPM006452A  SPM006460A  SPM006463A  SPM006466A  SPM006469A  SPM006472A  SPM006473A  SPM006477A  SPM006478A  SPM006479A  SPM006481A  SPM006482A  SPM006483A  SPM006484A  SPM006485A  SPM006486A  SPM006487A  SPM006488A  SPM006489A  SPM006496A  SPM006501A  SPM006502A  SPM006504A  SPM006505A  SPM006506A  SPM006507A  SPM006508A  SPM006509A  SPM006510A  SPM006512A  SPM006514A  SPM006516A  SPM006518A  SPM006521A  SPM006522A  SPM006525A  SPM006527A  SPM006534A  SPM006535A  SPM006536A  SPM006538A  SPM006540A  SPM006547A  SPM006550A  SPM006552A  SPM006555A  SPM006556A  SPM006560A  SPM006568A  SPM006569A  SPM006570A  SPM006574A  SPM006582A  SPM006583A  SPM006585A  SPM006587A  SPM006589A  SPM006592A  SPM006596A  SPM006598A  SPM006600A  SPM006603A  SPM006606A  SPM006609A  SPM006610A  SPM006613A  SPM006615A  SPM006617A  SPM006619A  SPM006620A  SPM006622A  SPM006623A  SPM006624A  SPM006625A  SPM006628A  SPM006651A  SPM006654A  SPM006660A  SPM006662A  SPM006666A  SPM006668A  SPM006675A  SPM006679A  SPM006683A  SPM006694A  SPM006702A  SPM006704A  SPM006706A  SPM006711A  SPM006712A  SPM006714A  SPM006722A  SPM006724A  SPM006725A  SPM006732A  SPM006741A  SPM006744A  SPM006757A  SPM006766A  SPM006775A  SPM006781A  SPM006784A  SPM006785A  SPM006788A  SPM006789A  SPM006799A  SPM006806A  SPM006807A  SPM006809A  SPM006810A  SPM006811A  SPM006812A  SPM006817A  SPM006828A  SPM006835A  SPM006838A  SPM006840A  SPM006841A  SPM006844A  SPM006847A  SPM006855A  SPM006857A  SPM006867A  SPM006874A  SPM006878A  SPM006885A  SPM006886A  SPM006887A  SPM006892A  SPM006894A  SPM006928A  SPM006941A  SPM006943A  SPM006946A  SPM006954A  SPM006956A  SPM006957A  SPM006959A  SPM006960A  SPM006962A  SPM006967A  SPM006970A  SPM006971A  SPM006973A  SPM006978A  SPM006990A  SPM006996A  SPM006997A  SPM007000A  SPM007001A  SPM007004A  SPM007008A  SPM007027A  SPM007028A  SPM007029A  SPM007030A  SPM007036A  SPM007038A  SPM007040A  SPM007045A  SPM007050A  SPM007052A  SPM007053A  SPM007064A  SPM007065A  SPM007075A  SPM007081A  SPM007082A  SPM007083A  SPM007091A  SPM007109A  SPM007110A  SPM007113A  SPM007118A  SPM007120A  SPM007123A  SPM007124A  SPM007125A  SPM007127A  SPM007140A  SPM007150A  SPM007175A  SPM007185A  SPM007192A  SPM007194A  SPM007197A  SPM007206A  SPM007207A  SPM007208A  SPM007209A  SPM007210A  SPM007211A  SPM007223A  SPM007224A  SPM007236A  SPM007237A  SPM007251A  SPM007259A  SPM007271A  SPM007284A  SPM007288A  SPM007290A  SPM007291A  SPM007301A  SPM007303A  SPM007305A  SPM007319A  SPM007322A  SPM007339A  SPM007350A  SPM007355A  SPM007359A  SPM007364A  SPM007370A  SPM007371A  SPM007372A  SPM007375A  SPM007379A  SPM007381A  SPM007382A  SPM007388A  SPM007389A  SPM007391A  SPM007397A  SPM007398A  SPM007399A  SPM007403A  SPM007405A  SPM007419A  SPM007424A  SPM007425A  SPM007431A  SPM007434A  SPM007436A  SPM007437A  SPM007444A  SPM007447A  SPM007450A  SPM007451A  SPM007455A  SPM007462A  SPM007464A  SPM007465A  SPM007479A  SPM007480A  SPM007482A  SPM007500A  SPM007513A  SPM007523A  SPM007524A  SPM007527A  SPM007529A  SPM007552A  SPM007559A  SPM007560A  SPM007565A  SPM007573A  SPM007575A  SPM007576A  SPM007577A  SPM007581A  SPM007585A  SPM007589A  SPM007592A  SPM007593A  SPM007594A  SPM007597A  SPM007611A  SPM007617A  SPM007619A  SPM007623A  SPM007626A  SPM007627A  SPM007628A  SPM007632A  SPM007634A  SPM007635A  SPM007636A  SPM007644A  SPM007645A  SPM007646A  SPM007651A  SPM007653A  SPM007655A  SPM007660A  SPM007666A  SPM007667A  SPM007668A  SPM007669A  SPM007671A  SPM007676A  SPM007679A  SPM007680A  SPM007681A  SPM007682A  SPM007683A  SPM007685A  SPM007686A  SPM007687A  SPM007690A  SPM007692A  SPM007693A  SPM007695A  SPM007696A  SPM007698A  SPM007700A  SPM007702A  SPM007704A  SPM007705A  SPM007714A  SPM007719A  SPM007721A  SPM007722A  SPM007723A  SPM007726A  SPM007728A  SPM007729A  SPM007731A  SPM007733A  SPM007740A  SPM007746A  SPM007747A  SPM007748A  SPM007749A  SPM007750A  SPM007752A  SPM007754A  SPM007755A  SPM007758A  SPM007759A  SPM007760A  SPM007762A  SPM007766A  SPM007767A  SPM007772A  SPM007775A  SPM007776A  SPM007777A  SPM007782A  SPM007784A  SPM007785A  SPM007788A  SPM007793A  SPM007795A  SPM007796A  SPM007797A  SPM007799A  SPM007802A  SPM007804A  SPM007805A  SPM007808A  SPM007809A  SPM007811A  SPM007813A  SPM007816A  SPM007818A  SPM007819A  SPM007824A  SPM007825A  SPM007829A  SPM007830A  SPM007831A  SPM007832A  SPM007833A  SPM007835A  SPM007837A  SPM007839A  SPM007840A  SPM007844A  SPM007845A  SPM007849A  SPM007850A  SPM007856A  SPM007857A  SPM007858A  SPM007869A  SPM007872A  SPM007873A  SPM007874A  SPM007875A  SPM007876A  SPM007877A  SPM007879A  SPM007880A  SPM008034A  SPM008039A  SPM008042A  SPM008045A  SPM008046A  SPM008048A  SPM008049A  SPM008052A  SPM008055A  SPM008060A  SPM008061A  SPM008065A  SPM008066A  SPM008067A  SPM008068A  SPM008075A  SPM008083A  SPM008086A  SPM008095A  SPM008101A  SPM008104A  SPM008110A  SPM008117A  SPM008120A  SPM008128A  SPM008129A  SPM008130A  SPM008131A  SPM008133A  SPM008134A  SPM008135A  SPM008136A  SPM008138A  SPM008140A  SPM008141A  SPM008143A  SPM008144A  SPM008145A  SPM008147A  SPM008148A  SPM008149A  SPM008150A  SPM008151A  SPM008152A  SPM008153A  SPM008154A  SPM008156A  SPM008157A  SPM008158A  SPM008160A  SPM008162A  SPM008167A  SPM008169A  SPM008171A  SPM008172A  SPM008174A  SPM008176A  SPM008178A  SPM008179A  SPM008181A  SPM008183A  SPM008186A  SPM008187A  SPM008194A  SPM008195A  SPM008196A  SPM008199A  SPM008201A  SPM008203A  SPM008204A  SPM008205A  SPM008206A  SPM008207A  SPM008208A  SPM008209A  SPM008210A  SPM008212A  SPM008213A  SPM008214A  SPM008216A  SPM008217A  SPM008218A  SPM008219A  SPM008221A  SPM008222A  SPM008223A  SPM008226A  SPM008227A  SPM008229A  SPM008230A  SPM008231A  SPM008238A  SPM008240A  SPM008241A  SPM008242A  SPM008243A  SPM008244A  SPM008245A  SPM008247A  SPM008248A  SPM008251A  SPM008252A  SPM008254A  SPM008255A  SPM008259A  SPM008260A  SPM008261A  SPM008262A  SPM008263A  SPM008264A  SPM008265A  SPM008267A  SPM008271A  SPM008272A  SPM008273A  SPM008275A  SPM008278A  SPM008281A  SPM008282A  SPM008285A  SPM008287A  SPM008289A  SPM008291A  SPM008293A  SPM008294A  SPM008295A  SPM008299A  SPM008302A  SPM008303A  SPM008304A  SPM008305A  SPM008306A  SPM008308A  SPM008309A  SPM008311A  SPM008312A  SPM008314A  SPM008315A  SPM008316A  SPM008320A  SPM008322A  SPM008323A  SPM008328A  SPM008331A  SPM008333A  SPM008335A  SPM008336A  SPM008338A  SPM008339A  SPM008341A  SPM008344A  SPM008347A  SPM008348A  SPM008350A  SPM008352A  SPM008353A  SPM008354A  SPM008355A  SPM008356A  SPM008359A  SPM008360A  SPM008362A  SPM008366A  SPM008367A  SPM008371A  SPM008372A  SPM008373A  SPM008375A  SPM008376A  SPM008378A  SPM008379A  SPM008381A  SPM008382A  SPM008383A  SPM008385A  SPM008390A  SPM008392A  SPM008401A  SPM008404A  SPM008405A  SPM008406A  SPM008409A  SPM008414A  SPM008415A  SPM008418A  SPM008424A  SPM008425A  SPM008430A  SPM008432A  SPM008434A  SPM008437A  SPM008438A  SPM008440A  SPM008441A  SPM008443A  SPM008444A  SPM008448A  SPM008449A  SPM008451A  SPM008452A  SPM008454A  SPM008455A  SPM008456A  SPM008458A  SPM008460A  SPM008461A  SPM008462A  SPM008463A  SPM008466A  SPM008467A  SPM008468A  SPM008471A  SPM008472A  SPM008473A  SPM008474A  SPM008476A  SPM008477A  SPM008479A  SPM008480A  SPM008481A  SPM008483A  SPM008487A  SPM008488A  SPM008489A  SPM008491A  SPM008497A  SPM008498A  SPM008499A  SPM008500A  SPM008501A  SPM008502A  SPM008506A  SPM008509A  SPM008510A  SPM008511A  SPM008512A  SPM008514A  SPM008515A  SPM008516A  SPM008517A  SPM008519A  SPM008520A  SPM008521A  SPM008522A  SPM008523A  SPM008524A  SPM008525A  SPM008527A  SPM008528A  SPM008530A  SPM008532A  SPM008533A  SPM008534A  SPM008535A  SPM008536A  SPM008537A  SPM008539A  SPM008540A  SPM008541A  SPM008543A  SPM008544A  SPM008545A  SPM008547A  SPM008549A  SPM008550A  SPM008556A  SPM008559A  SPM008560A  SPM008561A  SPM008562A  SPM008566A  SPM008568A  SPM008570A  SPM008576A  SPM008582A  SPM008584A  SPM008585A  SPM008587A  SPM008589A  SPM008593A  SPM008594A  SPM008595A  SPM008596A  SPM008597A  SPM008601A  SPM008602A  SPM008603A  SPM008605A  SPM008606A  SPM008610A  SPM008612A  SPM008613A  SPM008614A  SPM008617A  SPM008618A  SPM008619A  SPM008621A  SPM008622A  SPM008623A  SPM008624A  SPM008625A  SPM008626A  SPM008627A  SPM008629A  SPM008631A  SPM008632A  SPM008633A  SPM008634A  SPM008635A  SPM008637A  SPM008639A  SPM008640A  SPM008641A  SPM008643A  SPM008645A  SPM008646A  SPM008648A  SPM008649A  SPM008650A  SPM008651A  SPM008652A  SPM008655A  SPM008657A  SPM008660A  SPM008661A  SPM008662A  SPM008664A  SPM008665A  SPM008667A  SPM008668A  SPM008670A  SPM008672A  SPM008673A  SPM008675A  SPM008676A  SPM008678A  SPM008679A  SPM008680A  SPM008681A  SPM008685A  SPM008688A  SPM008689A  SPM008696A  SPM008697A  SPM008698A  SPM008699A  SPM008702A  SPM008703A  SPM008705A  SPM008706A  SPM008707A  SPM008708A  SPM008710A  SPM008711A  SPM008712A  SPM008714A  SPM008715A  SPM008718A  SPM008719A  SPM008721A  SPM008725A  SPM008726A  SPM008727A  SPM008728A  SPM008729A  SPM008730A  SPM008731A  SPM008732A  SPM008733A  SPM008734A  SPM008735A  SPM008742A  SPM008745A  SPM008751A  SPM008752A  SPM008766A  SPM008767A  SPM008769A  SPM008770A  SPM008771A  SPM008772A  SPM008773A  SPM008775A  SPM008776A  SPM008777A  SPM008778A  SPM008780A  SPM008781A  SPM008783A  SPM008788A  SPM008792A  SPM008793A  SPM008795A  SPM008796A  SPM008797A  SPM008800A  SPM008803A  SPM008807A  SPM008814A  SPM008816A  SPM008818A  SPM008823A  SPM008826A  SPM008829A  SPM008830A  SPM008831A  SPM008832A  SPM008836A  SPM008838A  SPM008839A  SPM008840A  SPM008841A  SPM008843A  SPM008844A  SPM008846A  SPM008849A  SPM008853A  SPM008854A  SPM008855A  SPM008856A  SPM008858A  SPM008861A  SPM008894A  SPM008925A  SPM008929A  SPM008932A  SPM008939A  SPM008968A  SPM008979A  SPM008980A  SPM008981A  SPM008982A  SPM008983A  SPM008985A  SPM008986A  SPM008988A  SPM008989A  SPM008990A  SPM008993A  SPM009014A  SPM009019A  SPM009020A  SPM009022A  SPM009023A  SPM009024A  SPM009026A  SPM009028A  SPM009029A  SPM009031A  SPM009033A  SPM009035A  SPM009036A  SPM009037A  SPM009049A  SPM009053A  SPM009058A  SPM009066A  SPM009070A  SPM009078A  SPM009081A  SPM009083A  SPM009084A  SPM009086A  SPM009088A  SPM009100A  SPM009101A  SPM009103A  SPM009105A  SPM009107A  SPM009108A  SPM009109A  SPM009114A  SPM009144A  SPM009150A  SPM009154A  SPM009162A  SPM009164A  SPM009165A  SPM009166A  SPM009168A  SPM009170A  SPM009171A  SPM009173A  SPM009174A  SPM009179A  SPM009180A  SPM009184A  SPM009185A  SPM009186A  SPM009191A  SPM009192A  SPM009193A  SPM009194A  SPM009197A  SPM009198A  SPM009204A  SPM009206A  SPM009212A  SPM009215A  SPM009217A  SPM009223A  SPM009227A  SPM009229A  SPM009230A  SPM009233A  SPM009234A  SPM009236A  SPM009237A  SPM009239A  SPM009240A  SPM009243A  SPM009246A  SPM009247A  SPM009248A  SPM009249A  SPM009257A  SPM009258A  SPM009261A  SPM009262A  SPM009263A  SPM009265A  SPM009266A  SPM009268A  SPM009271A  SPM009275A  SPM009278A  SPM009283A  SPM009284A  SPM009285A  SPM009286A  SPM009290A  SPM009292A  SPM009299A  SPM009300A  SPM009301A  SPM009310A  SPM009314A  SPM009321A  SPM009325A  SPM009326A  SPM009328A  SPM009331A  SPM009332A  SPM009336A  SPM009340A  SPM009342A  SPM009344A  SPM009345A  SPM009347A  SPM009350A  SPM009351A  SPM009354A  SPM009358A  SPM009360A  SPM009361A  SPM009363A  SPM009364A  SPM009368A  SPM009372A  SPM009374A  SPM009375A  SPM009377A  SPM009380A  SPM009388A  SPM009389A  SPM009390A  SPM009394A  SPM009397A  SPM009398A  SPM009400A  SPM009402A  SPM009403A  SPM009404A  SPM009405A  SPM009408A  SPM009409A  SPM009411A  SPM009414A  SPM009418A  SPM009419A  SPM009421A  SPM009422A  SPM009423A  SPM009425A  SPM009426A  SPM009427A  SPM009428A  SPM009429A  SPM009431A  SPM009432A  SPM009435A  SPM009437A  SPM009438A  SPM009439A  SPM009444A  SPM009447A  SPM009448A  SPM009451A  SPM009452A  SPM009453A  SPM009456A  SPM009458A  SPM009460A  SPM009462A  SPM009463A  SPM009465A  SPM009468A  SPM009471A  SPM009472A  SPM009474A  SPM009479A  SPM009480A  SPM009482A  SPM009483A  SPM009485A  SPM009486A  SPM009487A  SPM009490A  SPM009494A  SPM009495A  SPM009496A  SPM009497A  SPM009499A  SPM009501A  SPM009502A  SPM009503A  SPM009504A  SPM009507A  SPM009509A  SPM009511A  SPM009513A  SPM009514A  SPM009515A  SPM009516A  SPM009518A  SPM009519A  SPM009520A  SPM009522A  SPM009525A  SPM009526A  SPM009527A  SPM009529A  SPM009530A  SPM009532A  SPM009533A  SPM009536A  SPM009538A  SPM009540A  SPM009541A  SPM009545A  SPM009546A  SPM009547A  SPM009548A  SPM009549A  SPM009550A  SPM009551A  SPM009552A  SPM009553A  SPM009554A  SPM009555A  SPM009558A  SPM009566A  SPM009568A  SPM009569A  SPM009570A  SPM009571A  SPM009572A  SPM009573A  SPM009575A  SPM009577A  SPM009578A  SPM009581A  SPM009582A  SPM009584A  SPM009591A  SPM009599A  SPM009600A  SPM009601A  SPM009603A  SPM009604A  SPM009607A  SPM009608A  SPM009617A  SPM009618A  SPM009623A  SPM009624A  SPM009625A  SPM009626A  SPM009630A  SPM009632A  SPM009642A  SPM009651A  SPM009652A  SPM009657A  SPM009658A  SPM009662A  SPM009663A  SPM009665A  SPM009669A  SPM009675A  SPM009676A  SPM009678A  SPM009687A  SPM009827A  SPM009829A  SPM009832A  SPM009834A  SPM009836A  SPM009837A  SPM009839A  SPM009842A  SPM009843A  SPM009844A  SPM009845A  SPM009846A  SPM009847A  SPM009848A  SPM009849A  SPM009850A  SPM009852A  SPM009853A  SPM009854A  SPM009857A  SPM009858A  SPM009864A  SPM009868A  SPM009870A  SPM009871A  SPM009872A  SPM009874A  SPM009878A  SPM009880A  SPM009882A  SPM009883A  SPM009885A  SPM009891A  SPM009894A  SPM009897A  SPM009898A  SPM009900A  SPM009901A  SPM009902A  SPM009904A  SPM009906A  SPM009909A  SPM009913A  SPM009915A  SPM009917A  SPM009918A  SPM009920A  SPM009921A  SPM009924A  SPM009927A  SPM009930A  SPM009931A  SPM009932A  SPM009933A  SPM009934A  SPM009935A  SPM009937A  SPM009938A  SPM009940A  SPM009941A  SPM009942A  SPM009943A  SPM009949A  SPM009951A  SPM009952A  SPM009954A  SPM009956A  SPM009958A  SPM009960A  SPM009962A  SPM009964A  SPM009966A  SPM009971A  SPM009972A  SPM009973A  SPM009975A  SPM009976A  SPM009980A  SPM009981A  SPM009982A  SPM009983A  SPM009985A  SPM009986A  SPM009987A  SPM009988A  SPM009989A  SPM009990A  SPM009991A  SPM009992A  SPM009993A  SPM009995A  SPM009996A  SPM009997A  SPM009998A  SPM009999A  SPM010000A  SPM010001A  SPM010003A  SPM010005A  SPM010007A  SPM010008A  SPM010009A  SPM010010A  SPM010013A  SPM010035A  SPM010036A  SPM010038A  SPM010039A  SPM010040A  SPM010042A  SPM010043A  SPM010045A  SPM010047A  SPM010049A  SPM010052A  SPM010055A  SPM010056A  SPM010058A  SPM010059A  SPM010060A  SPM010061A  SPM010065A  SPM010069A  SPM010071A  SPM010072A  SPM010074A  SPM010075A  SPM010076A  SPM010079A  SPM010080A  SPM010082A  SPM010088A  SPM010089A  SPM010090A  SPM010091A  SPM010092A  SPM010093A  SPM010096A  SPM010099A  SPM010100A  SPM010102A  SPM010103A  SPM010104A  SPM010107A  SPM010108A  SPM010109A  SPM010111A  SPM010114A  SPM010118A  SPM010121A  SPM010126A  SPM010127A  SPM010130A  SPM010132A  SPM010134A  SPM010136A  SPM010139A  SPM010141A  SPM010143A  SPM010144A  SPM010146A  SPM010148A  SPM010149A  SPM010150A  SPM010153A  SPM010154A  SPM010156A  SPM010157A  SPM010158A  SPM010161A  SPM010166A  SPM010168A  SPM010172A  SPM010173A  SPM010174A  SPM010175A  SPM010177A  SPM010178A  SPM010179A  SPM010180A  SPM010182A  SPM010183A  SPM010188A  SPM010189A  SPM010191A  SPM010192A  SPM010195A  SPM010199A  SPM010200A  SPM010201A  SPM010207A  SPM010211A  SPM010213A  SPM010214A  SPM010215A  SPM010217A  SPM010218A  SPM010219A  SPM010220A  SPM010221A  SPM010222A  SPM010224A  SPM010225A  SPM010226A  SPM010228A  SPM010229A  SPM010233A  SPM010234A  SPM010240A  SPM010242A  SPM010244A  SPM010246A  SPM010247A  SPM010248A  SPM010249A  SPM010252A  SPM010253A  SPM010254A  SPM010255A  SPM010256A  SPM010259A  SPM010260A  SPM010261A  SPM010262A  SPM010264A  SPM010265A  SPM010266A  SPM010267A  SPM010269A  SPM010270A  SPM010271A  SPM010273A  SPM010274A  SPM010280A  SPM010281A  SPM010282A  SPM010284A  SPM010287A  SPM010292A  SPM010293A  SPM010294A  SPM010295A  SPM010297A  SPM010298A  SPM010299A  SPM010302A  SPM010306A  SPM010308A  SPM010309A  SPM010310A  SPM010312A  SPM010313A  SPM010316A  SPM010319A  SPM010321A  SPM010323A  SPM010324A  SPM010326A  SPM010331A  SPM010332A  SPM010333A  SPM010334A  SPM010335A  SPM010338A  SPM010340A  SPM010341A  SPM010342A  SPM010343A  SPM010344A  SPM010345A  SPM010346A  SPM010348A  SPM010350A  SPM010352A  SPM010354A  SPM010357A  SPM010360A  SPM010363A  SPM010365A  SPM010367A  SPM010368A  SPM010369A  SPM010371A  SPM010372A  SPM010375A  SPM010376A  SPM010377A  SPM010379A  SPM010380A  SPM010381A  SPM010382A  SPM010384A  SPM010385A  SPM010386A  SPM010388A  SPM010389A  SPM010390A  SPM010393A  SPM010395A  SPM010397A  SPM010399A  SPM010401A  SPM010402A  SPM010403A  SPM010404A  SPM010406A  SPM010407A  SPM010408A  SPM010410A  SPM010411A  SPM010412A  SPM010413A  SPM010414A  SPM010415A  SPM010417A  SPM010419A  SPM010420A  SPM010422A  SPM010428A  SPM010434A  SPM010435A  SPM010436A  SPM010438A  SPM010439A  SPM010441A  SPM010443A  SPM010444A  SPM010445A  SPM010449A  SPM010453A  SPM010455A  SPM010457A  SPM010459A  SPM010460A  SPM010461A  SPM010462A  SPM010463A  SPM010464A  SPM010467A  SPM010468A  SPM010469A  SPM010471A  SPM010476A  SPM010481A  SPM010482A  SPM010484A  SPM010490A  SPM010492A  SPM010493A  SPM010494A  SPM010495A  SPM010496A  SPM010503A  SPM010505A  SPM010506A  SPM010512A  SPM010513A  SPM010515A  SPM010517A  SPM010518A  SPM010521A  SPM010526A  SPM010527A  SPM010529A  SPM010534A  SPM010535A  SPM010538A  SPM010540A  SPM010542A  SPM010543A  SPM010545A  SPM010548A  SPM010549A  SPM010550A  SPM010564A  SPM010567A  SPM010568A  SPM010570A  SPM010700A  SPM010860A  SPM010919A  SPM010935A  SPM010972A  SPM010992A  SPM010998A  SPM011008A  SPM011009A  SPM011030A  SPM011054A  SPM011069A  SPM011070A  SPM011079A  SPM011080A  SPM011081A  SPM011096A  SPM011101A  SPM011103A  SPM011106A  SPM011110A  SPM011111A  SPM011113A  SPM011124A  SPM011127A  SPM011130A  SPM011135A  SPM011136A  SPM011141A  SPM011144A  SPM011145A  SPM011146A  SPM011147A  SPM011148A  SPM011149A  SPM011159A  SPM011160A  SPM011163A  SPM011167A  SPM011169A  SPM011173A  SPM011174A  SPM011176A  SPM011179A  SPM011181A  SPM011182A  SPM011184A  SPM011186A  SPM011187A  SPM011190A  SPM011192A  SPM011193A  SPM011194A  SPM011195A  SPM011199A  SPM011201A  SPM011209A  SPM011210A  SPM011211A  SPM011212A  SPM011214A  SPM011215A  SPM011219A  SPM011220A  SPM011221A  SPM011222A  SPM011223A  SPM011224A  SPM011227A  SPM011228A  SPM011229A  SPM011232A  SPM011233A  SPM011234A  SPM011238A  SPM011240A  SPM011241A  SPM011242A  SPM011243A  SPM011246A  SPM011248A  SPM011250A  SPM011253A  SPM011257A  SPM011260A  SPM011263A  SPM011267A  SPM011268A  SPM011270A  SPM011272A  SPM011278A  SPM011281A  SPM011284A  SPM011285A  SPM011286A  SPM011288A  SPM011291A  SPM011293A  SPM011297A  SPM011298A  SPM011301A  SPM011302A  SPM011303A  SPM011305A  SPM011311A  SPM011313A  SPM011314A  SPM011315A  SPM011318A  SPM011325A  SPM011327A  SPM011328A  SPM011329A  SPM011333A  SPM011334A  SPM011335A  SPM011337A  SPM011338A  SPM011340A  SPM011342A  SPM011344A  SPM011346A  SPM011347A  SPM011349A  SPM011351A  SPM011356A  SPM011359A  SPM011369A  SPM011372A  SPM011373A  SPM011383A  SPM011386A  SPM011387A  SPM011390A  SPM011392A  SPM011402A  SPM011405A  SPM011406A  SPM011409A  SPM011414A  SPM011418A  SPM011419A  SPM011424A  SPM011426A  SPM011429A  SPM011433A  SPM011434A  SPM011437A  SPM011438A  SPM011452A  SPM011455A  SPM011457A  SPM011458A  SPM011461A  SPM011479A  SPM011480A  SPM011482A  SPM011485A  SPM011765A  SPM011772A  SPM011782A  SPM011790A  SPM011791A  SPM011795A  SPM011798A  SPM011804A  SPM011806A  SPM011808A  SPM011809A  SPM011812A  SPM011821A  SPM",
      "center_classification_date": "20200705",
      "distribution_pattern": "worldwide",
      "state": "MN",
      "product_description": "CareLink SmartSync Device Manager, Model Number 24970A",
      "report_date": "20200715",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)",
      "recall_number": "Z-2505-2020",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "85816",
      "more_code_info": "011823A  SPM011824A  SPM011827A  SPM011833A  SPM011837A  SPM011840A  SPM011848A  SPM011853A  SPM011854A  SPM011856A  SPM011859A  SPM011863A  SPM011864A  SPM011868A  SPM011870A  SPM011874A  SPM011876A  SPM011878A  SPM011881A  SPM011884A  SPM011893A  SPM011896A  SPM011902A  SPM011904A  SPM011905A  SPM011908A  SPM011912A  SPM011914A  SPM011917A  SPM011920A  SPM011921A  SPM011923A  SPM011941A  SPM011943A  SPM011951A  SPM011953A  SPM011962A  SPM011971A  SPM011998A  SPM012005A  SPM012009A  SPM012014A  SPM012016A  SPM012018A  SPM012024A  SPM012025A  SPM012027A  SPM012035A  SPM012036A  SPM012043A  SPM012046A  SPM012047A  SPM012049A  SPM012062A  SPM012064A  SPM012065A  SPM012071A  SPM012074A",
      "recall_initiation_date": "20200604",
      "postal_code": "55112-4391",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Ongoing"
    }
  ]
}