  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98994",
      "recalling_firm": "Medline Industries, LP",
      "address_1": "3 Lakes Dr",
      "address_2": "",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of AE, BB, BM, BS, CA, CY, GU, NL, PA, QA, SA, VI.",
      "recall_number": "Z-2504-2026",
      "product_description": "MEDLINE Medical Procedure Kits labeled as:    1) D AND C PACK, Medline Kit Number/SKU DYNJ53973C;   2) PK CUST CV A&B CABRINI CO, Medline Kit Number/SKU DYNJ60583C.",
      "product_quantity": "168 units",
      "reason_for_recall": "During an internal review, Medline identified representative samples did not meet applicable performance specifications. Identified issues include irregularities in catheter surface finish, instances where the catheter connector detached prematurely from the catheter funnel, and limited cases where the catheter outer diameter did not meet specification.",
      "recall_initiation_date": "20260504",
      "center_classification_date": "20260617",
      "report_date": "20260624",
      "code_info": "Medline Product Number/SKU DYNJ53973C:  UDI/DI 10195327008864 (each), 40195327008865 (case), Lot Number 21KME435;  Medline Product Number/SKU DYNJ53973C:  UDI/DI 10195327008864 (each), 40195327008865 (case), Lot Number 21JMC247;  Medline Product Number/SKU DYNJ53973C:  UDI/DI 10195327008864 (each), 40195327008865 (case), Lot Number 21JMB678;  Medline Product Number/SKU DYNJ53973C:  UDI/DI 10195327008864 (each), 40195327008865 (case), Lot Number 21IMH919;  Medline Product Number/SKU DYNJ60583C:  UDI/DI 10195327605582 (each), 40195327605583 (case), Lot Number 24CBO389;  Medline Product Number/SKU DYNJ60583C:  UDI/DI 10195327605582 (each), 40195327605583 (case), Lot Number 24ABR872.",
      "more_code_info": ""
    }
  ]
}