{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "NORTH HAVEN",
      "address_1": "60 NIDDLETOWN AVE 2ND F",
      "reason_for_recall": "Device may be missing a sled component. The sled component is responsible for staple deployment. This  may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal  contents.",
      "address_2": "",
      "product_quantity": "",
      "code_info": "N7L0711KX, N7L1141KX, N7M0853KX, N8B0419KX, N7L0912KX,  N7M0241KX, N8A0271KX, N8C0012KX, N7L0973KX, N7M0353KX,  N8A0684KX, N8C0143KX, N7L1038KX, N7M0366KX, N8A0924KX,  N7M0996KX, N7L0807KX, N7M0548KX, N8A1143KX, N8A0173KX,   N7L1050KX, N7M0617KX, N8B0161KX, N8A0488KX, N7L1125KX,  N7M0745KX, N8B0283KX & N8C0136KX",
      "center_classification_date": "20180719",
      "distribution_pattern": "Nationwide U.S.A.   Internationally to: Albania, Algeria, Austria, Belgium, Bulgaria, Croatia,  Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece,  Guadeloupe, Hungary, Iceland, Ireland, Israel, Italy, Ivory Coast, Jordan,  Kazakhstan, Kuwait, Latvia, Lebanon, Luxembourg, Macedonia, Malta,  Martinique, Moldova, Morocco, Namibia, Netherlands, New Caledonia,  Norway, Pakistan, Poland, Portugal, Reunion, Romania, Russian federation,  Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Spain, Sweden,  Switzerland, Tunisia, Turkey, United Arab Emirates & United Kingdom",
      "state": "CT",
      "product_description": "Endo GIA 60 mm Articulating Medium/Thick Reload with Tri-StapleTechnology, Item Code EGIA60AMT",
      "report_date": "20180725",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "COVIDIEN MEDTRONIC",
      "recall_number": "Z-2504-2018",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "80423",
      "more_code_info": "",
      "recall_initiation_date": "20180522",
      "postal_code": "06473",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Ongoing"
    }
  ]
}