  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Orleans",
      "state": "",
      "country": "France",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80452",
      "recalling_firm": "Maquet SAS",
      "address_1": "Parc De Limere",
      "address_2": "Avenue De La Pomme De Pin",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of AR, AZ, CA, CO, CT, FL, GA, IA, ID, KS, KY, LA, MD, ME, MO, NE, NY, OH, OR, TN, TX, UT, and WA. Foreign distribution to Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Brazil Cameroon, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Ghana, Hong Kong, Hungary, India, Iran, Iraq, Ireland, Italy, Japan, Jordan, Kenya, Korea, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Namibia, Netherlands, Nigeria, Oman, Pakistan, Panama, Paraguay, Poland, Portugal, South Korea, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, Tanzania, Yemen, and Zimbabwe.",
      "recall_number": "Z-2502-2018",
      "product_description": "VOLISTA StandOP Surgical Light,   ARD568811961  VLT600DF AIM STP;  ARD568821961  VLT600DF AIM STP QL;  ARD568811911  VLT600SF AIM STP;  ARD568821911  VLT600SF AIM STP QL;  ARD568821910  VLT600SF STP QL;  ARD568811901  VLT600SF AIM STP    Product  Usage:  The Maquet VOLISTA surgical lights are designed to illuminate the body of a patient during surgical operations, diagnostics and treatment. StandOP VOLISTA is a system of ceiling-mounted surgical lights suited for installation in surgical suites, examining rooms, doctors surgeries and out-patient consultations.",
      "product_quantity": "151",
      "reason_for_recall": "Certain VOLISTA StandOP Surgical Lights have been identified as having a potential light head detachment issue. Improper assembly of the bracket that connects the light head to the axle fork could result in the light head bracket breaking, causing the light head to detach from the arm and become suspended only from the cabling inside the arm. Use of an affected device may result in serious injury to patient and/or medical staff if the light head detaches and drops, impacting patient and/or medical staff during demonstration, service, or use.",
      "recall_initiation_date": "20180614",
      "center_classification_date": "20180719",
      "termination_date": "20220623",
      "report_date": "20180725",
      "code_info": "All Serial Numbers manufactured from March 10, 2014 until November 17, 2016  6.",
      "more_code_info": ""
    }
  ]
}