  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Alpharetta",
      "state": "GA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71966",
      "recalling_firm": "Halyard Health, Inc",
      "address_1": "5405 Windward Pkwy",
      "address_2": "N/A",
      "postal_code": "30004-3894",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide and Puerto Rico and Internationally to Canada",
      "recall_number": "Z-2496-2015",
      "product_description": "KimVent* Closed Suction System for Adults, 14 F, D-Tip, T-Piece (Product Code 22059); KimVent* Closed Suction System for Adults, 14 F, T-Piece, Directional Tip (Product Code 8155)  A tracheobronchial suction catheter is a device that is attached to a port of the airway circuit for use as needed to aspirate liquids or semisolids from a patient's upper airway",
      "product_quantity": "N/A",
      "reason_for_recall": "The thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the \"LOCKED\" position.",
      "recall_initiation_date": "20150729",
      "center_classification_date": "20150831",
      "termination_date": "20160112",
      "report_date": "20150909",
      "code_info": "Product Code 22059 - M5076T619, M5082T608, M5085T603; Product Code 8155 - M5082T606"
    }
  ]
}