      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "63125",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "N/A",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution - US Nationwide including the states of: AZ, CA, CO,FL, MN, IL, NJ, NY, OH, OR and the countries of Austrailia, Utd. Arab Emir., France, Germany, India, and Italy.",
      "recall_number": "Z-2493-2012",
      "product_description": "Cemented Tibial Drill    Product Usage:  The cemented tibial drill is used to prepare the tibial bone prior to the impaction of the applicable stemmed tibial broach. The cemented tibial drill has engraved lines that are size specific to guide the user as to the depth to insert the drill into the tibial bone.",
      "product_quantity": "60",
      "reason_for_recall": "Zimmer is initiating a recall of the Cemented Tibial Drill for the Zimmer Persona Personalized Knee System due to  three reports being received as of August 29. 2012 for tibial cortex perforation during preparation of the tibial bone.  The investigation into the reports that have been received found that the perforations are occurring while using the  Cemented Tibial Drill. The perforations have o",
      "recall_initiation_date": "20120907",
      "center_classification_date": "20120928",
      "termination_date": "20140417",
      "report_date": "20121010",
      "code_info": "Part 42-5399-018-00, Lots 62034571, 62089766"
    }
  ]
}