  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Alpharetta",
      "state": "GA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71966",
      "recalling_firm": "Halyard Health, Inc",
      "address_1": "5405 Windward Pkwy",
      "address_2": "N/A",
      "postal_code": "30004-3894",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US (nationwide and Puerto Rico and Internationally to Canada",
      "recall_number": "Z-2492-2015",
      "product_description": "KimVent* Closed Suction System for Adults, 12 F, T-Piece (Product Code 2155); KimVent* Closed Suction System for Adults, 14 F, T-Piece (Product Code 2205, 8169, 8248, 22015, 22025, 220125, 220135)     A tracheobronchial suction catheter is a device that is attached to a port of the airway circuit for use as needed to aspirate liquids or semisolids from a patient's upper airway.",
      "product_quantity": "19,250 cases",
      "reason_for_recall": "The thumb valve of some KIMVENT Closed Suction Systems remained partially open after suctioning was completed, even though the valve appeared to be closed or in the \"LOCKED\" position.",
      "recall_initiation_date": "20150729",
      "center_classification_date": "20150831",
      "termination_date": "20160112",
      "report_date": "20150909",
      "code_info": "Product Code 2155 - M5166T402, M5166T622, M5166T631; Product Code 2205 - M5079T606, M5086T603, M5089T612, M5089T632, M5110T611, M5110T626, M5166T642; Product Code 8169 - M5103T608; Product Code 8248 - M5082T632; Product Code 22015 - M5082T603, M5089T609, M5096T607; Product Code 22025 - M5082T627, M5096T603; Product Code 220125  M5082T617, M5096T608; Product Code 220135 - M5054T619, M5079T604, M5082T620, M5089T614, M5096T622"
    }
  ]
}