  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Houston",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71967",
      "recalling_firm": "Trilliant Surgical Ltd.",
      "address_1": "6721 Portwest Dr Ste 160",
      "address_2": "N/A",
      "postal_code": "77024-8019",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "US Distribution to the states of : FL, OH, NJ,  MN, WI, NM and CA.",
      "recall_number": "Z-2490-2015",
      "product_description": "7 Hole VL Gridlock Fibula Plate, Part # 300-60-001 NON-STERILE SINGLE USE ONLY; 9 Hole VL Gridlock Fibula Plate, Part # 300-60-002 NON-STERILE SINGLE USE ONLY; 11 Hole VL Gridlock Fibula Plate, Part # 300-60-003 NON-STERILE SINGLE USE ONLY. For use in trauma and reconstructive procedures in small bones.",
      "product_quantity": "28 units",
      "reason_for_recall": "The affected parts subject to the recall are out of specification, resulting in in the loss of the optimal locking screw functionality of the plates.  The thread depth of the screw holes is insufficient, potentially limiting the ability to engage and lock with associated mating locking screws.",
      "recall_initiation_date": "20150806",
      "center_classification_date": "20150831",
      "termination_date": "20170126",
      "report_date": "20150909",
      "code_info": "Lots: TSL002595, TSL002596, TSL002597"
    }
  ]
}