  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Englewood",
      "state": "CO",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88553",
      "recalling_firm": "Paragon 28, Inc.",
      "address_1": "14445 Grasslands Dr",
      "address_2": "",
      "postal_code": "80112-7062",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "U.S. Nationwide distribution in the states of CA, FL, GA, ID, KS, KY, LA, MI, MO, NC, NY, TX, VA, WA and WV.  O.U.S.: N/A",
      "recall_number": "Z-2488-2021",
      "product_description": "Phantom Nail System, Drill, 4.6 x 300mm, Cannulated, 3/16\" Sq. Connection    p28 P99-110-4630",
      "product_quantity": "14 drill bits",
      "reason_for_recall": "A thin wall condition between the inner cannulation and the flutes of the drill which potentially could cause a fracture through the wall of the drill to the cannulation beneath.",
      "recall_initiation_date": "20210730",
      "center_classification_date": "20210917",
      "termination_date": "20211215",
      "report_date": "20210929",
      "code_info": "Model Number: P99-110-4630 UDI Code: (01)00889795056849(10)HN120550 Lot Number: HN120550",
      "more_code_info": ""
    }
  ]
}