{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "West Chester",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71862",
      "recalling_firm": "Synthes (USA) Products LLC",
      "address_1": "1301 Goshen Pkwy",
      "address_2": "N/A",
      "postal_code": "19380-5986",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution including US nationwide, Canada and Germany.",
      "recall_number": "Z-2488-2015",
      "product_description": "VEPTR Vertical Expandable Prosthetic Titanium Rib, inferior cradles also known as caudal rib supports.  The VEPTR/VEPTR II devices mechanically stabilize and/or correct thoracic deformities to allow the chest and lungs to grow.",
      "product_quantity": "3262",
      "reason_for_recall": "The last hole on certain VEPTR Inferior Cradles may be partially formed.  Fully lengthening the construct to the partially formed last hole may result in mechanical failure of the construct due to rib sleeve breakage, and may require  surgury to exchange components.  Effects may also include pain, loss of deformity correction, and/or soft tissue/visceral injury.",
      "recall_initiation_date": "20150729",
      "center_classification_date": "20150831",
      "termination_date": "20160818",
      "report_date": "20150909",
      "code_info": "part numbers: 497.087  497.088  497.089  497.093  497.094  497.095  497.098  497.099  497.100  497.225  497.226  497.227  497.228  497.229  497.230  497.231  497.232  497.233  497.234  497.235  497.236  497.237  497.238  497.239  497.241  497.242  497.243  497.244  497.245  497.246  497.247  497.248  497.249  lot numbers: 4550443  4938481  4947780  4947788  4955528  4955529  4955532  5830333  5830334  6005310  6011867  7371262  7371275  7424595  7459468  7613026  4814417  4884492  4938495  4947969  4955498  4955501  5770304  5830335  6011868  7371263  7459469  7613027  4686201  4814414  4938497  4938501  4938766  4947971  6011869  4362289  4393923  4816422  4935344  4938512  4938515  4938759  4947977  5767359  5770641  5812094  6011844  7371280  7423480  7613044  4362288  4816420  4938517  4938519  4938758  4947978  6024576  7371281  4362286  4816414  4816415  4938521  4938523  4938756  5811904  6011884  4362283  4562149  4617874  4816115  4816117  4884494  4938540  4947924  4947925  4955412  4955413  6005302  6011850  7371288  7844724  4362282  4685618  4698047  4816113  4938543  4938549  4947926  6011872  7371266  7423469  7544000  7613030  7775437  4362281  4393924  4685619  4816110  4938503  4938560  4947927  4947928  6011873  4362280  4393905  4550457  4568693  4617037  4686343  4686346  4694014  4694234  4755789  4755791  4755793  4821421  4882741  4882742  4882743  4882744  4947893  4947894  4947895  4947896  4947897  4947898  4947899  4947900  4969661  4969663  5816072  5857598  5883097  5883098  6031197  6031198  6079441  6087165  4362279  4393906  4550458  4568694  4617038  4686347  4686349  4686351  4694235  4709517  4755785  4755786  4755787  4821415  4882745  4882747  4882748  4938735  4947901  4947902  4947903  4947905  4947910  4969659  4969660  5642635  5702649  5744112  5744238  5883100  5883101  6011907  6024580  6087170  6087171  6087172  6087173  4362278  4393910  4550459  4617039  4686353  4694236  4709518  4755783  4755784  4821410  4882750  4882751  4882752  4882753  4938737  4947911  4947912  4947913  4969658  5668754  5697400  5770445  5857607  5872385  5883103  5883105  6021058  6024577  6031199  6031200  6063205  4362277  4550460  4568695  4617040  4686356  4694237  4755478  4755479  4755480  4821407  4882754  4882755  4882756  4882757  4947914  4947915  4947916  4947917  4947918  5426428  5770298  5839271  5857616  5883106  5883107  6016089  6031201  6031202  4568700  4617041  4686359  4694021  4694238  4755472  4755474  4755476  4821404  4882758  4882759  4882760  4882761  4947919  4947920  4947921  4969657  5839270  5872094  5872378  6024578  6031203  6031204  4362276  4686289  4697976  4755470  4755471  4885455  4947922  4947980  4947985  6011852  6031205  4362275  4550461  4686290  4938738  4947987  4947988  4947990  6011885  6031206  4362274  4686291  4697978  4755469  4885458  4938739  4947992  4947994  4947996  5407126  6030989  6030990  6031207  6052890  4362273  4697984  4755468  4885459  4938740  4947997  4947998  4948000  4948001  6011847  6031208  4686293  4820546  4885461  4938741  4948003  4948005  6011853  4362271  4686363  4755466  4755467  4885471  4948007  4948008  4948010  4948012  4969634  5365436  5668735  5668742  5680726  5872325  6009825  6016101  6031209  4362270  4393912  4686365  4695024  4820936  4885476  4938742  4938744  4948014  5407131  5407133  6011874  6031031  6031210  4686366  4695025  4755464  4820933  4820934  4885480  4938746  4948015  5407129  5407130  6031211  4362269  4686367  4695026  4755462  4820931  4820932  4938747  4948017  6011854  4686368  4820929  4948019  4948020  6011876  6024579  4362268  4617047  4686202  4814412  4885508  4948022  4948024  4969628  5642583  4686203  4885509  4948026  4948027  4969627  5759696  5835992  6011896  4208597  4686204  4885510  4938751  4947982  4948033  4948034  4969626  5505857  5505862  5816076  6011855  6367928  6882133  7544008  9807745  4362267  4617048  4885511  4938753  4947983  4947984  4947989  4969625  5767354  5770536  6005308  6011856  4362266  4885512  4938754  4938755  4947991  4947993  4969624  5682975  5770675  6011857  4208598  4938757  4938760  4938761  4938765  4947999  4948002  4948004  5770692  5812078  5812079  6011877  6087168  6367930  7599075  7613037  9807752  4362265  4816096  4938669  4938768  4948006  4948009  4969623  5835993  5891761  6005313  4362264  4816094  4885514  4938672  4948013  4969622  5649087  5744115  5816079  5835994  5861990  6011913  4208599  4617050  4695408  4816091  4816092  4885515  4938674  4948016  4948018  4948021  4969621  5767416  5770683  6005309  6011858  6052891  6916285  7404349  7544315"
    }
  ]
}