  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Skaneateles Falls",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "88481",
      "recalling_firm": "Welch Allyn Inc",
      "address_1": "4341 State Street Rd",
      "address_2": "N/A",
      "postal_code": "13153-5300",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution and the countries of Afghanistan,  Andorra, Australia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belgium, Bermuda, Bolivia, Botswana, Brazil, Brunei Dar-es-S, Canada, Chile, China,  Colombia, Congo, Costa Rica, C¿te D'ivoire (Ivory Coast), Croatia, Cyprus, Czech Republic,  Denmark, Dominican Rep., Ecuador, Estonia, Finland, France, Frenc.Polynesia, French, Guiana, Germany, Greece, Guadeloupe, Guam, Guatemala, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jordan, Kenya,  Kuwait, Lebanon, Lesotho, Libya, Luxembourg, Malaysia, Maldives, Malta, Martinique,  Mexico, Monaco, Myanmar, N.Mariana Islnd, Netherlands (the), New Zealand, Norway,  Oman, Pakistan, Panama, Philippines, Poland, Portugal, Puerto Rico, Qatar, Reunion Romania, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden,  Switzerland, Taiwan, Thailand, Trinidad, Tobago, Tunisia, Turkey, Ukraine, United Kingdom, Utd.Arab Emir., Vietnam, Zimbabwe.",
      "recall_number": "Z-2482-2021",
      "product_description": "Braun ThermoScan PRO 6000 Ear Thermometer-indicated for the intermittent measurement of human body temperature for patients having ages ranging from normal weight (full term) newborn to geriatric adults in a professional use environment.  Model:  6000",
      "product_quantity": "1,036,740 units",
      "reason_for_recall": "If the device is exposed to fluid ingress and is used before the cleaning fluid has completely dried, there is a risk of the device overheating potentially causing a burn to the user or the ear canal of the patient",
      "recall_initiation_date": "20210720",
      "center_classification_date": "20210916",
      "report_date": "20210922",
      "code_info": "All Devices in distribution.  UDI: 00732094309027 DDDYYKXXXXX  DDD = 3 DIGIT JULIAN DAY (EXAMPLE - 234). YY = LAST 2 DIGITS OF JULIAN YEAR (EXAMPLE - 13) K = KEYTRONIC XXXXX = 5 DIGIT BASE 10 SERIALIZED NUMBER.",
      "more_code_info": ""
    }
  ]
}