      "status": "Terminated",
      "city": "Mishawaka",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71611",
      "recalling_firm": "Bayer Healthcare, LLC",
      "address_1": "430 S Beiger St",
      "address_2": "N/A",
      "postal_code": "46544-3207",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution -- USA, Australia, Austria, Belgium, Canada, Croatia, Denmark, Finland, France, Germany, Greece, Ireland, Italy, Luxembourg, Netherlands, Norway, Portugal, Slovenia, Spain, Sweden, Switzerland, UK, Albania, Azerbaijan, Bulgaria, Czech Republic, Estonia, Hungary, Korea, Latvia, Lithuania, Philippines, Romania, Serbia, Slovakia, and Thailand.",
      "recall_number": "Z-2482-2015",
      "product_description": "Contour Next Control Solution for Contour Next Blood Glucose Test System.  Liquid quality control solution.  Plastic bottle, 2.5mL.    Indicated for use with fresh fingertip capillary whole blood samples.",
      "product_quantity": "5 million bottles of control solution",
      "reason_for_recall": "If the control solution bottle is not thoroughly shaken prior to use, the control test result may be higher than the control range and may not be recognized and marked as a control test by the blood glucose meter.  Also, the control solution, used as a quality control test in blood glucose meter systems, may read outside the specified range too high or too low.",
      "recall_initiation_date": "20150622",
      "center_classification_date": "20150828",
      "termination_date": "20160523",
      "report_date": "20150909",
      "code_info": "SKU# 7314 (Level 2) and SKU# 7315 (Level 1)  All Lots"
    }
  ]
}