  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Houston",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85233",
      "recalling_firm": "Preventice Services, LLC",
      "address_1": "1717 N Sam Houston Pkwy W Ste 100",
      "address_2": "N/A",
      "postal_code": "77038-1324",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Telephone",
      "distribution_pattern": "US Nationwide Distribution to AK, AL, GA, NJ, OH, and TX.  There was no foreign/military/government distribution.",
      "recall_number": "Z-2479-2020",
      "product_description": "BodyGuardian Heart Remote Monitoring Kit, consisting of two BG Heart monitors, either Models BC-1001-A or BC1001-D, and a cell phone, either Models BP1008 or BP10014, that is paired to both of the monitors.",
      "product_quantity": "8 kits",
      "reason_for_recall": "The device data being collected and transferred to the monitoring center may not be accurate due to nonvalidated association between the phone software and the heart monitors, therefore, the patient's report should not be used to evaluate their condition.",
      "recall_initiation_date": "20190806",
      "center_classification_date": "20200701",
      "termination_date": "20221118",
      "report_date": "20200708",
      "code_info": "Version 2.3 phone software, kit IDs BG2570727, BG2565845, BG2567817, BG2566832, BG2566830, BG2570699,  BG2566530, and BG2570700."
    }
  ]
}